Which keratometer is most reliable for correcting astigmatism with toric intraocular lenses?

Minwook Chang1, Su-Yeon Kang, Hyo Myung Kim.   

Abstract

PURPOSE: To evaluate the accuracy of preoperative keratometers used in cataract surgery with toric intraocular lens (IOL).
METHODS: Twenty-five eyes received an AcrySof toric IOL implantation. Four different keratometric methods, a manual keratometer, an IOL master, a Pentacam and an auto keratometer, were performed preoperatively in order to evaluate preexisting corneal astigmatism. Differences between the true residual astigmatism and the anticipated residual astigmatism (keratometric error) were compared at one and three months after surgery by using a separate vector analysis to identify the keratometric method that provided the highest accuracy for astigmatism control.
RESULTS: The mean keratomeric error was 0.52 diopters (0.17-1.17) for the manual keratometer, 0.62 (0-1.31) for the IOL master, 0.69 (0.08-1.92) for the Pentacam, and 0.59 (0.08-0.94) for the auto keratometer. The manual keratometer was the most accurate, although there was no significant difference between the keratometers (p > 0.05). All of the keratometers achieved an average keratometric error of less than one diopter.
CONCLUSIONS: Manual keratometry was the most accurate of the four methods evaluated, although the other techniques were equally satisfactory in determining corneal astigmatism.

Entities:  

Keywords:  Astigmatism; Cataract; Keratometer; Keratometric error; Toric intraocular lenses

Mesh:

Year:  2012        PMID: 22323879      PMCID: PMC3268162          DOI: 10.3341/kjo.2012.26.1.10

Source DB:  PubMed          Journal:  Korean J Ophthalmol        ISSN: 1011-8942


  21 in total

1.  Accuracy and reproducibility of biometry using partial coherence interferometry.

Authors:  Rafe Connors; Peter Boseman; Randall J Olson
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

2.  Clinical evaluation of the Shin-Nippon NVision-K 5001/Grand Seiko WR-5100K autorefractor.

Authors:  Leon Nicholas Davies; Edward Arthur Harry Mallen; James Stuart Wolffsohn; Bernard Gilmartin
Journal:  Optom Vis Sci       Date:  2003-04       Impact factor: 1.973

3.  A comparison of the ARK-700A autokeratometer and Medmont E300 corneal topographer when measuring peripheral corneal curvature.

Authors:  J M González-Méijome; J Jorge; A Queiros; J B Almeida; M A Parafita
Journal:  Ophthalmic Physiol Opt       Date:  2004-09       Impact factor: 3.117

4.  Clinical evaluation of the Grand Seiko Auto Ref/Keratometer WAM-5500.

Authors:  Amy L Sheppard; Leon N Davies
Journal:  Ophthalmic Physiol Opt       Date:  2009-12-09       Impact factor: 3.117

5.  Computerized videokeratography and keratometry in determining intraocular lens calculations.

Authors:  S E Husain; T Kohnen; R Maturi; H Er; D D Koch
Journal:  J Cataract Refract Surg       Date:  1996-04       Impact factor: 3.351

6.  Biometry of 7,500 cataractous eyes.

Authors:  K J Hoffer
Journal:  Am J Ophthalmol       Date:  1980-09       Impact factor: 5.258

Review 7.  Status of toric intraocular lenses.

Authors:  Jeffrey D Horn
Journal:  Curr Opin Ophthalmol       Date:  2007-02       Impact factor: 3.761

8.  Comparison of laser interferometry and ultrasound A-scan in the measurement of axial length.

Authors:  R Goyal; R V North; J E Morgan
Journal:  Acta Ophthalmol Scand       Date:  2003-08

9.  Cataract patients in a defined Swedish population 1986-1990. II. Preoperative observations.

Authors:  K Ninn-Pedersen; U Stenevi; B Ehinger
Journal:  Acta Ophthalmol (Copenh)       Date:  1994-02

10.  Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis.

Authors:  W Haigis; B Lege; N Miller; B Schneider
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2000-09       Impact factor: 3.117

View more
  13 in total

1.  Agreement of keratometric readings measured using rotating Scheimpflug imaging, auto-refractokeratometer, and biograph in eyes with keratoconus.

Authors:  Meltem Guzin Altınel; Hasim Uslu
Journal:  Int Ophthalmol       Date:  2021-02-05       Impact factor: 2.031

2.  Keratometry with five different techniques: a study of device repeatability and inter-device agreement.

Authors:  Shiva Mehravaran; Soheila Asgari; Sara Bigdeli; Ashkan Shahnazi; Hassan Hashemi
Journal:  Int Ophthalmol       Date:  2014-02-23       Impact factor: 2.031

3.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

4.  Subjective and objective depth of field measures in pseudophakic eyes: comparison between extended depth of focus, trifocal and bifocal intraocular lenses.

Authors:  Carlos Palomino-Bautista; Rubén Sánchez-Jean; David Carmona-González; David P Piñero; Ainhoa Molina-Martín
Journal:  Int Ophthalmol       Date:  2019-10-03       Impact factor: 2.031

5.  Effects of posterior corneal astigmatism on the accuracy of AcrySof toric intraocular lens astigmatism correction.

Authors:  Bin Zhang; Jing-Xue Ma; Dan-Yan Liu; Cong-Rong Guo; Ying-Hua Du; Xiu-Jin Guo; Yue-Xian Cui
Journal:  Int J Ophthalmol       Date:  2016-09-18       Impact factor: 1.779

6.  Prospective study of toric IOL outcomes based on the Lenstar LS 900® dual zone automated keratometer.

Authors:  Kjell Gunnar Gundersen; Richard Potvin
Journal:  BMC Ophthalmol       Date:  2012-07-16       Impact factor: 2.209

7.  Torsional and flattening effect on corneal astigmatism after cataract surgery: a retrospective analysis.

Authors:  Yuli Park; Hyun Seung Kim
Journal:  BMC Ophthalmol       Date:  2017-02-08       Impact factor: 2.209

8.  Investigation of a real-time location system of corneal astigmatic axis.

Authors:  Jian-Guo Zhao; An-Peng Pan; Ke Zheng; A-Yong Yu
Journal:  Eye Vis (Lond)       Date:  2017-09-14

9.  Use of the toric intraocular lens for keratoconus treatment.

Authors:  Jasmin Zvornicanin; Emir Cabric; Vahid Jusufovic; Zlatko Musanovic; Edita Zvornicanin
Journal:  Acta Inform Med       Date:  2014-04

10.  Preoperative corneal astigmatism among adult patients with cataract in Northern Nigeria.

Authors:  Mohammed Isyaku; Syed A Ali; Sadiq Hassan
Journal:  Indian J Ophthalmol       Date:  2014-11       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.