Agreement of keratometric readings measured using rotating Scheimpflug imaging, auto-refractokeratometer, and biograph in eyes with keratoconus.

Meltem Guzin Altınel1, Hasim Uslu2.   

Abstract

AIM: To determine agreement in keratometric readings obtained using rotating Scheimpflug imaging with Pentacam, biograph with Lenstar LS900, and Topcon KR-8100P auto-keratorefractometer in eyes with different stages of keratoconus.
METHODS: A total of 89 eyes of 58 patients with keratoconus were examined in this study, retrospectively. The eyes were divided into two groups: mild group (group 1: 42 eyes) (Amsler-Krumeich stage 1) and moderate-to-severe group (group 2: 47 eyes) (Amsler-Krumeich stage 2, 3, 4). The keratometric readings measured using the Pentacam Scheimpflug system, Lenstar LS900, and Topcon KR-8100P auto-keratorefractometer were compared between the groups. The effects of the measurements of anterior chamber depth, Q value, axial length, central corneal thickness (CCT), and maximum value of keratometry (Kmax) on the differences of devices for keratometric readings were investigated.
RESULTS: The mean values of the keratometric readings obtained using the Lenstar were steeper than with the Pentacam and Topcon, especially in group 2. In group 1, the mean K2 values measured using the Lenstar were significantly steeper than with the Topcon (p < 0.05); however, the devices were accordant for the other keratometric readings. In group 2, there was an agreement between the Pentacam and Topcon for the mean K1 and Km values; however, there were significant differences between the devices for the other values. The Q value and CCT had a negative correlation, and Kmax had a positive correlation with the differences of Lenstar-Pentacam and Lenstar-Topcon (p < 0.01).
CONCLUSION: According to our results, Pentacam-Topcon and Pentacam-Lenstar can be used interchangeably for keratometry in mild stages of keratoconus. The keratometric readings of Lenstar were found steeper than the other devices with increasing grades of keratoconus. None of these devices can be used interchangeably in moderate-to-severe stages of keratoconus.

Entities:  

Keywords:  Keratoconus; Keratometry; Scheimpflug system

Mesh:

Year:  2021        PMID: 33544351     DOI: 10.1007/s10792-021-01720-5

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  14 in total

1.  Scheimpflug-Placido topographer and optical low-coherence reflectometry biometer: repeatability and agreement.

Authors:  Wuhe Chen; Colm McAlinden; Konrad Pesudovs; Qinmei Wang; Fan Lu; Yifan Feng; Jie Chen; Jinhai Huang
Journal:  J Cataract Refract Surg       Date:  2012-07-03       Impact factor: 3.351

2.  A new optical low coherence reflectometry device for ocular biometry in cataract patients.

Authors:  P J Buckhurst; J S Wolffsohn; S Shah; S A Naroo; L N Davies; E J Berrow
Journal:  Br J Ophthalmol       Date:  2009-04-19       Impact factor: 4.638

3.  Accuracy of a new partial coherence interferometry analyser for biometric measurements.

Authors:  M P Holzer; M Mamusa; G U Auffarth
Journal:  Br J Ophthalmol       Date:  2009-03-15       Impact factor: 4.638

4.  Comparison of anterior segment measurements with rotating Scheimpflug photography and partial coherence reflectometry.

Authors:  Jinhai Huang; Konrad Pesudovs; Daizong Wen; Shihao Chen; Thomas Wright; Xiaoyu Wang; Yini Li; Qinmei Wang
Journal:  J Cataract Refract Surg       Date:  2011-02       Impact factor: 3.351

5.  Effect of keratoconus grades on repeatability of keratometry readings: Comparison of 5 devices.

Authors:  Hassan Hashemi; Abbasali Yekta; Mehdi Khabazkhoob
Journal:  J Cataract Refract Surg       Date:  2015-05       Impact factor: 3.351

6.  Comparison of keratometry and white-to-white measurements obtained by Lenstar with those obtained by autokeratometry and corneal topography.

Authors:  Jinhai Huang; Giacomo Savini; Binbin Su; Rui Zhu; Yifan Feng; Shishi Lin; Hao Chen; Qinmei Wang
Journal:  Cont Lens Anterior Eye       Date:  2015-05-05       Impact factor: 3.077

7.  Comparison of 2 laser instruments for measuring axial length.

Authors:  Kenneth J Hoffer; H John Shammas; Giacomo Savini
Journal:  J Cataract Refract Surg       Date:  2010-04       Impact factor: 3.351

8.  Cataract surgery in keratoconus.

Authors:  Nuthida Thebpatiphat; Kristin M Hammersmith; Christopher J Rapuano; Brandon D Ayres; Elisabeth J Cohen
Journal:  Eye Contact Lens       Date:  2007-09       Impact factor: 2.018

9.  Evaluation of the Lenstar LS 900 non-contact biometer.

Authors:  L P J Cruysberg; M Doors; F Verbakel; T T J M Berendschot; J De Brabander; R M M A Nuijts
Journal:  Br J Ophthalmol       Date:  2009-08-18       Impact factor: 4.638

10.  Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace.

Authors:  Yanjun Hua; Zequan Xu; Wei Qiu; Qiang Wu
Journal:  PLoS One       Date:  2016-01-11       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.