Comparison of laser interferometry and ultrasound A-scan in the measurement of axial length.

R Goyal1, R V North, J E Morgan.   

Abstract

AIM: Laser interferometry is a new, non-contact technique for the measurement of axial length. In this study we compared measurements of axial length obtained with this technique with those obtained with ultrasound (A-scan). The reproducibility and examiner-dependency of the two methods were also analysed.
METHODS: Patients presenting at the cataract assessment clinic were invited to participate in the study. Axial length measurements were obtained both by contact ultrasound (A-scan) and by non-contact laser interferometry (IOLMasterTM V1.1, Carl Zeiss, Jena, Germany). Intraocular lens powers were calculated using both sets of measurements. The coefficient of variation served as a measure of reproducibility.
RESULTS: A total of 100 eyes in 100 patients were evaluated after informed consent had been obtained. Although estimates of axial length obtained with the two techniques were highly correlated, axial lengths obtained with the contact method (mean 23.35 mm, SD 1.81 mm) were consistently lower than those obtained with the non-contact method (mean 23.55 mm, SD 1.76 mm) and the difference was statistically significant (p < 0.001). The coefficient of variation was lower with laser interferometry (0.1%) than with the ultrasound technique (0.49%).
CONCLUSIONS: Different estimates of axial length are obtained using contact and non-contact techniques, with the latter producing consistently higher measurements than the former. Laser interferometry provides more reproducible results that should improve the accuracy of measurements of axial length in the clinical setting.

Entities:  

Mesh:

Year:  2003        PMID: 12859258     DOI: 10.1034/j.1600-0420.2003.00092.x

Source DB:  PubMed          Journal:  Acta Ophthalmol Scand        ISSN: 1395-3907


  7 in total

1.  Relationship between the retinal thickness of the macula and the difference in axial length.

Authors:  Tetsuo Ueda; Yoshiaki Nawa; Yoshiaki Hara
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-08-13       Impact factor: 3.117

Review 2.  Optimising biometry for best outcomes in cataract surgery.

Authors:  R Sheard
Journal:  Eye (Lond)       Date:  2013-12-06       Impact factor: 3.775

3.  Which keratometer is most reliable for correcting astigmatism with toric intraocular lenses?

Authors:  Minwook Chang; Su-Yeon Kang; Hyo Myung Kim
Journal:  Korean J Ophthalmol       Date:  2012-01-14

4.  Prediction accuracy of intraocular lens power calculation methods after laser refractive surgery.

Authors:  Yubo Wu; Songyu Liu; Rongfeng Liao
Journal:  BMC Ophthalmol       Date:  2017-04-08       Impact factor: 2.209

5.  Refractive predictability of partial coherence interferometry and factors that can affect it.

Authors:  Seung Mo Kim; Joohyun Choi; Sangkyung Choi
Journal:  Korean J Ophthalmol       Date:  2009-03-09

6.  Distribution of axial length, anterior chamber depth, and corneal curvature in an aged population in South China.

Authors:  Hui Chen; Haotian Lin; Zhuoling Lin; Jingjing Chen; Weirong Chen
Journal:  BMC Ophthalmol       Date:  2016-05-01       Impact factor: 2.209

7.  Minimum effective volume of local anesthetic in peribulbar block: does it differ with the eyeball axial length?

Authors:  Sanaa M El Fawal; Walid H Nofal; Eman A S Sabek; Wail Ahmed Abdelaal
Journal:  Braz J Anesthesiol       Date:  2021-09-22
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.