| Literature DB >> 22091380 |
Matthias C Burg1, Alexander C Bunck, Harald Seifarth, Boris Buerke, Harald Kugel, Volker Hesselmann, Michael Köhler, Walter Heindel, David Maintz.
Abstract
Purpose. To evaluate stent lumen visibility of a large sample of different peripheral arterial (iliac, renal, carotid) stents using magnetic resonance angiography in vitro. Materials and Methods. 21 different stents and one stentgraft (10 nitinol, 7 316L, 2 tantalum, 1 cobalt superalloy, 1 PET + cobalt superalloy, and 1 platinum alloy) were examined in a vessel phantom (vessel diameters ranging from 5 to 13 mm) filled with a solution of Gd-DTPA. Stents were imaged at 1.5 Tesla using a T1-weighted 3D spoiled gradient-echo sequence. Image analysis was performed measuring three categories: Signal intensity in the stent lumen, lumen visibility of the stented lumen, and homogeneity of the stented lumen. The results were classified using a 3-point scale (good, intermediate, and poor results). Results. 7 stents showed good MR lumen visibility (4x nitinol, 2x tantalum, and 1x cobalt superalloy). 9 stents showed intermediate results (5x nitinol, 2x 316L, 1x PET + cobalt superalloy, and 1x platinum alloy) and 6 stents showed poor results (1x nitinol, and 5x 316L). Conclusion. Stent lumen visibility varies depending on the stent material and type. Some products show good lumen visibility which may allow the detection of stenoses inside the lumen, while other products cause artifacts which prevent reliable evaluation of the stent lumen with this technique.Entities:
Year: 2010 PMID: 22091380 PMCID: PMC3197262 DOI: 10.1155/2011/478175
Source DB: PubMed Journal: Radiol Res Pract ISSN: 2090-195X
Name, manufacturer, material, primary area of use, diameter, and length of the examined stents.
| No. | Name | Manufacturer | Material | Primary area of use | Diameter (mm) | Length (mm) |
|---|---|---|---|---|---|---|
| 1 | SAXX Small | Devon | 316 L | Arterial vessels | 5 | 17 |
| 2 | SelfX Xpert | Abbott | nitinol | Peripheral vasculature/biliary duct | 5 | 20 |
| 3 | Palmaz Genesis | Cordis | 316 L | Peripheral arteries below aortic arch/biliary tree | 6 | 39 |
| 4 | Absolute | Guidant | nitinol | Biliary/peripheral vessels | 7 | 60 |
| 5 | AccuLink Carotid | Guidant | nitinol | Internal carotid and common carotid arteries | 7 | 40 |
| 6 | Express Vascular LD | Boston Scientific | 316 L | Peripheral vessels | 7 | 37 |
| 7 | OmniLink 0,018 | Guidant | 316 L | Biliary/peripheral arteries | 7 | 18 |
| 8 | OmniLink 0,035 | Guidant | 316 L | Biliary/peripheral arteries | 7 | 35 |
| 9 | Palmaz Corinthian IQ | Cordis | 316 L | Peripheral vessels | 7 | 40 |
| 10 | Renal109 | Abbott | tantalum | Renal artery | 7 | 18 |
| 11 | Renal137 | Abbott | tantalum | Renal artery | 7 | 18 |
| 12 | Sentinol | Boston Scientific | nitinol | Peripheral vessels | 7 | 59 |
| 13 | Symphony | Boston Scientific | nitinol | Iliac artery | 7 | 40 |
| 14 | Vascuflex SE | B. Braun | nitinol | Peripheral vessels | 7 | 20 |
| 15 | Wallstent Uni | Boston Scientific | cobalt-superalloy | Iliac artery, sfa (superficial femoral artery), tracheal | 7 | 60 |
| 16 | Zilver | Cook | nitinol | Carotid artery | 7 | 40 |
| 17 | Evo | pfm | nitinol | Pelvic arteries and peripheral vessels/biliary | 8 | 50 |
| 18 | RxCarotid | Abbott | nitinol | Carotid artery | 8 | 30 |
| 19 | SAXX Large | Devon | 316 L | Arterial vessels | 8 | 35 |
| 20 | Evo Target | pfm | nitinol | Intravascular/biliary | 10 | 80 |
| 21 | Wallgraft Endoprothesis | Boston Scientific | braided polyester graft bonded to the outside of a Wallstent (cobalt-superalloy) | Trachea/bronchus (off label: peripheral arteries) | 10 | 70 |
| 22 | CP Stent | pfm | 90% platinum 10% iridium | Aorta | 13 | 28 |
Figure 1Comparison of 22 different peripheral arterial stents. Longitudinal MIP reformations are shown.
Results of signal intensity measurements, visible lumen measurements, and lumen homogeneity measurements.
| No. | Name | % of signal intensity inside the stented lumen | Score signal intensity | Visible lumen diameter (%) | Score lumen narrowing | SD Lumen/Mean Lumen | Score homogeneity |
|---|---|---|---|---|---|---|---|
| 1 | SAXX Small | 7 | 1 | 27 | 1 | 0.43 | 1 |
| 2 | SelfX Xpert | 62 | 3 | 58 | 2 | 0.11 | 2 |
| 3 | Palmaz Genesis | 6 | 1 | 0 | 1 | 0.44 | 1 |
| 4 | Absolute | 80 | 3 | 69 | 2 | 0.06 | 3 |
| 5 | AccuLink Carotid | 57 | 2 | 69 | 2 | 0.10 | 2 |
| 6 | Express Vascular LD | 21 | 1 | 0 | 1 | 0.48 | 1 |
| 7 | OmniLink 0,018 | 46 | 2 | 38 | 2 | 0.33 | 2 |
| 8 | OmniLink 0,035 | 45 | 2 | 51 | 2 | 0.21 | 2 |
| 9 | Palmaz Corinthian IQ | 12 | 1 | 0 | 1 | 0.42 | 1 |
| 10 | Renal109 | 66 | 3 | 72 | 3 | 0.16 | 2 |
| 11 | Renal137 | 69 | 3 | 58 | 2 | 0.07 | 3 |
| 12 | Sentinol | 51 | 2 | 89 | 3 | 0.04 | 3 |
| 13 | Symphony | 15 | 1 | 60 | 2 | 0.35 | 2 |
| 14 | Vascuflex SE | 42 | 2 | 80 | 3 | 0.09 | 3 |
| 15 | Wallstent Uni | 117 | 3 | 60 | 2 | 0.08 | 3 |
| 16 | Zilver | 66 | 3 | 66 | 2 | 0.07 | 3 |
| 17 | Evo | 45 | 2 | 80 | 3 | 0.19 | 2 |
| 18 | RxCarotid | 10 | 1 | 65 | 2 | 0.81 | 1 |
| 19 | SAXX Large | 7 | 1 | 0 | 1 | 0.40 | 1 |
| 20 | Evo Target | 43 | 2 | 92 | 3 | 0.16 | 2 |
| 21 | Wallgraft Endoprothesis | 52 | 2 | 75 | 3 | 0.16 | 2 |
| 22 | CP Stent | 21 | 1 | 86 | 3 | 0.14 | 2 |
Overall results.
| No. | Name | Score signal intensity | Score lumen narrowing | Score lumen homogeneity | Overall Score | Overall Evaluation |
|---|---|---|---|---|---|---|
| 1 | SAXX Small | 1 | 1 | 1 | 3 | Poor |
| 2 | SelfX Xpert | 3 | 2 | 2 | 7 | Intermediate |
| 3 | Palmaz Genesis | 1 | 1 | 1 | 3 | Poor |
| 4 | Absolute | 3 | 2 | 3 | 8 | Good |
| 5 | AccuLink Carotid | 2 | 2 | 2 | 6 | Intermediate |
| 6 | Express Vascular LD | 1 | 1 | 1 | 3 | Poor |
| 7 | OmniLink 0.018 | 2 | 2 | 2 | 6 | Intermediate |
| 8 | OmniLink 0.035 | 2 | 2 | 2 | 6 | Intermediate |
| 9 | Palmaz Corinthian IQ | 1 | 1 | 1 | 3 | Poor |
| 10 | Renal109 | 3 | 3 | 2 | 7 | Good |
| 11 | Renal137 | 3 | 2 | 3 | 8 | Good |
| 12 | Sentinol | 2 | 3 | 3 | 8 | Good |
| 13 | Symphony | 1 | 2 | 2 | 5 | Intermediate |
| 14 | Vascuflex SE | 2 | 3 | 3 | 8 | Good |
| 15 | Wallstent Uni | 3 | 2 | 3 | 8 | Good |
| 16 | Zilver | 3 | 2 | 3 | 8 | Good |
| 17 | Evo | 2 | 3 | 2 | 7 | Intermediate |
| 18 | RxCarotid | 1 | 2 | 1 | 4 | Poor |
| 19 | SAXX Large | 1 | 1 | 1 | 3 | Poor |
| 20 | Evo Target | 2 | 3 | 2 | 7 | Intermediate |
| 21 | Wallgraft Endoprothesis | 2 | 3 | 2 | 7 | Intermediate |
| 22 | CP Stent | 1 | 3 | 2 | 6 | Intermediate |