Literature DB >> 11012441

Stent appearance at contrast-enhanced MR angiography: in vitro examination with 14 stents.

M Lenhart1, M Völk, C Manke, W R Nitz, M Strotzer, S Feuerbach, J Link.   

Abstract

PURPOSE: To evaluate signal intensity changes influencing assessment of stent patency at contrast material-enhanced magnetic resonance (MR) angiography.
MATERIALS AND METHODS: By using an in vitro model, 14 stents-nine nitinol, one tantalum, two stainless steel, and two cobalt alloy-were investigated regarding their appearance at MR imaging. A vascular phantom consisting of tubes filled with 2.00 mmol/L gadopentetate dimeglumine in saline solution was studied in different orientations within the magnetic field. Imaging was performed with a fast three-dimensional gradient-echo sequence (4. 70/1.89 [repetition time msec/echo time msec]). Relative signal intensity reduction within the stents and the degree of artificial narrowing of the stent lumen were calculated.
RESULTS: The stent lumen was visible within 13 stents. A total signal void inside the stent lumen appeared in only one cobalt alloy stent. Artificial narrowing of the diameter was less than 33% in 10 of 14 stents. The tantalum stent and four nitinol stents seemed best suited for contrast-enhanced MR angiography. A bandlike artifact occurred at the ends of the stents when positioned along the readout direction.
CONCLUSION: To differentiate between artifacts and stenoses, knowledge of the degree of signal intensity reduction and artificial lumen narrowing within vascular stents is essential. Stent geometry, relative orientation to the magnetic field, and alloy composition influence signal intensity alteration within the stent lumen.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11012441     DOI: 10.1148/radiology.217.1.r00se28173

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  21 in total

1.  Stent-assisted coil placement in a wide-necked persistent trigeminal artery aneurysm with jailing of the trigeminal artery: a case report.

Authors:  Mohammed I Mohammed; Johnny S Sandhu; Ajay K Wakhloo
Journal:  AJNR Am J Neuroradiol       Date:  2002-03       Impact factor: 3.825

2.  MR-velocity mapping in vascular stents to assess peak systolic velocity. In vitro comparison of various stent designs made of Stainless Steel and Nitinol.

Authors:  Jacqueline van Holten; Patrik Kunz; Paul G H Mulder; Peter M T Pattynama; Hildo J Lamb; Lukas C van Dijk
Journal:  MAGMA       Date:  2002-11       Impact factor: 2.310

3.  Thoracic aortic stent graft: comparison of contrast-enhanced MR angiography and CT angiography in the follow-up: initial results.

Authors:  Stefanie Weigel; Bernd Tombach; David Maintz; Stefan Klotz; Thomas Vestring; Walter Heindel; Roman Fischbach
Journal:  Eur Radiol       Date:  2003-02-19       Impact factor: 5.315

Review 4.  Modern cross-sectional imaging in the diagnosis and follow-up of intracranial aneurysms.

Authors:  Karsten Papke; Friedhelm Brassel
Journal:  Eur Radiol       Date:  2006-01-14       Impact factor: 5.315

5.  Carotid artery stents: in vitro comparison of different stent designs and sizes using CT angiography and contrast-enhanced MR angiography at 1.5T and 3T.

Authors:  M Lettau; A Sauer; S Heiland; S Rohde; M Bendszus; S Hähnel
Journal:  AJNR Am J Neuroradiol       Date:  2009-09-12       Impact factor: 3.825

6.  Improved detection of in-stent restenosis by blood pool agent-enhanced, high-resolution, steady-state magnetic resonance angiography.

Authors:  Christina M Plank; Florian Wolf; Herbert Langenberger; Michael Weber; Dietrich Beitzke; Alfred Stadler; Martin Schillinger; Johannes Lammer; Christian Loewe
Journal:  Eur Radiol       Date:  2011-05-10       Impact factor: 5.315

7.  Contrast-Enhanced and Time-of-Flight MRA at 3T Compared with DSA for the Follow-Up of Intracranial Aneurysms Treated with the WEB Device.

Authors:  C Timsit; S Soize; A Benaissa; C Portefaix; J-Y Gauvrit; L Pierot
Journal:  AJNR Am J Neuroradiol       Date:  2016-04-21       Impact factor: 3.825

8.  Evaluation of various image reconstruction parameters in lower extremity stents using multidetector-row CT angiography: initial findings.

Authors:  Martin Heuschmid; Benjamin Wiesinger; Gunnar Tepe; Oliver Luz; Andreas F Kopp; Claus D Claussen; Stephan H Duda
Journal:  Eur Radiol       Date:  2006-05-30       Impact factor: 5.315

9.  Angiographic CT: in vitro comparison of different carotid artery stents-does stent orientation matter?

Authors:  Michael Lettau; Martin Bendszus; Stefan Hähnel
Journal:  Neuroradiology       Date:  2013-02-15       Impact factor: 2.804

10.  In vitro comparison of different carotid artery stents: a pixel-by-pixel analysis using CT angiography and contrast-enhanced MR angiography at 1.5 and 3 T.

Authors:  Michael Lettau; Annett Sauer; Sabine Heiland; Stefan Rohde; Julia Reinhardt; Martin Bendszus; Stefan Hähnel
Journal:  Neuroradiology       Date:  2009-11-19       Impact factor: 2.804