3.0 Tesla magnetic resonance angiography of endovascular aortic stent grafts: phantom measurements in comparison with 1.5 Tesla.

Stefan C Krämer1, Alexander Wall, David Maintz, Rainald Bachmann, Harald Kugel, Walter Heindel.   

Abstract

RATIONALE AND
OBJECTIVES: This study evaluated different stent grafts by 3 T magnetic resonance angiography (MRA) with respect to lumen visibility, susceptibility-induced signal loss, and type of stent artifacts compared with 1.5 T MRA in a phantom model.
METHODS: Six different stent-grafts (tube: n = 3, bifurcated: n = 3) were evaluated by 3 T and 1.5 T MRA using a tube phantom. MRA was performed using T1-weighted sequences at both systems with comparable parameters (3T: TR 5.4/TE 2.0/FA 30 degrees, 1.5 T: TR 6.2/TE 2.2/FA 30 degrees). A blind study of the image quality, including artifacts, was performed by 3 radiologists. Furthermore, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) values were calculated. Statistical analysis was performed with Student's t test (P < 0.05).
RESULTS: One Elgiloy stent graft showed almost a complete intraluminal signal loss at 1.5 and 3 T. All other models could be evaluated by both systems by MRA, resulting in a favorable lumen visibility (score: 1) for prostheses made of nitinol. Scores for overall image quality and artifacts were the same for both MR systems. SNR and CNR values of the stented part of the vessel phantom increased from 320 +/- 33 to 618 +/- 40 and from 306 +/- 34 to 596 +/- 40 at 3 T when compared with 1.5 T, resulting in a significant signal gain of 93% at the higher field strength.
CONCLUSIONS: 3 Tesla MRA of aortic stent grafts in a phantom model demonstrates an increase in SNR and CNR when compared with 1.5 T. However, the magnitude of imaging artifacts as well as coherent intraluminal signal loss within the stent does not increase equally in both MR systems.

Entities:  

Mesh:

Year:  2004        PMID: 15194912     DOI: 10.1097/01.rli.0000129470.63758.c1

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  5 in total

1.  MR liver imaging and cholangiography in the presence of surgical metallic clips at 1.5 and 3 Tesla.

Authors:  Elmar M Merkle; Brian M Dale; John Thomas; Erik K Paulson
Journal:  Eur Radiol       Date:  2006-05-16       Impact factor: 5.315

2.  Carotid artery stents: in vitro comparison of different stent designs and sizes using CT angiography and contrast-enhanced MR angiography at 1.5T and 3T.

Authors:  M Lettau; A Sauer; S Heiland; S Rohde; M Bendszus; S Hähnel
Journal:  AJNR Am J Neuroradiol       Date:  2009-09-12       Impact factor: 3.825

3.  MRI assessment of thoracic stent grafts after emergency implantation in multi trauma patients: a feasibility study.

Authors:  Volker Rasche; Alexander Oberhuber; Stephan Trumpp; Axel Bornstedt; Karl-Heinz Orend; Nico Merkle; Wolfgang Rottbauer; Martin Hoffmann
Journal:  Eur Radiol       Date:  2011-02-18       Impact factor: 5.315

4.  In vitro comparison of different carotid artery stents: a pixel-by-pixel analysis using CT angiography and contrast-enhanced MR angiography at 1.5 and 3 T.

Authors:  Michael Lettau; Annett Sauer; Sabine Heiland; Stefan Rohde; Julia Reinhardt; Martin Bendszus; Stefan Hähnel
Journal:  Neuroradiology       Date:  2009-11-19       Impact factor: 2.804

5.  MR Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types.

Authors:  Matthias C Burg; Alexander C Bunck; Harald Seifarth; Boris Buerke; Harald Kugel; Volker Hesselmann; Michael Köhler; Walter Heindel; David Maintz
Journal:  Radiol Res Pract       Date:  2010-07-27
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.