In vivo evaluation of patency and in-stent stenoses after implantation of nitinol stents in iliac arteries using MR angiography.

Okka W Hamer1, Thomas Finkenzeller, Ingitha Borisch, Christian Paetzel, Niels Zorger, Stefan Feuerbach, Wolfgang Nitz.   

Abstract

OBJECTIVE: Our study was a prospective in vivo study to evaluate whether MR angiography is suitable for assessing stent patency and grading in-stent stenoses and to examine whether the accuracy of MR angiography changes with time after stent implantation. SUBJECTS AND METHODS: In a prospective study, 34 iliac stenoses in 27 patients were treated by implantation of 35 nitinol stents. MR angiography was performed immediately after stent placement for 32 stents, and both digital subtraction angiography (DSA) and MR angiography were repeated at the 6-month follow-up for 23 stents. Three blinded observers assessed stent patency and the degree of in-stent stenoses on MR angiography and DSA (the standard of reference) images. The difference between the observers' grading of stenoses on DSA and on MR angiography was determined. Statistical analysis was performed using the Student's t test for paired samples.
RESULTS: Stent patency was assessed correctly for all stents and both sets of MR angiography images. Evaluation of DSA 1 images (obtained at end of implantation procedure) revealed that 96.9% of in-stent stenoses were less than 50%. On DSA 2 images (obtained at follow-up), 95.7% of in-stent stenoses were graded as less than 50%. The difference between grading of stenoses on DSA and MR angiography images was 15.0% +/- 16.0% (minimum, 0.0%; maximum, 63.3%) for DSA 1 versus MR angiography 1 (statistically significant, p = 0.037) and 9.8% +/- 13.5% (minimum, 0.0%; maximum, 63.3%) for MR angiography 2 versus DSA 2 (not statistically significant, p = 0.355).
CONCLUSION: Patency was correctly assessed for all stents on MR angiography. The quality of MR angiography regarding characterization of in-stent stenoses improved with time after stent placement. However, discrepancies of more than 60% between grading of lumen narrowing on DSA and MR angiography images occurred even at the 6-month follow-up. Thus, MR angiography is not yet a reliable technique for characterization of in-stent stenoses.

Entities:  

Mesh:

Year:  2005        PMID: 16247150     DOI: 10.2214/AJR.04.1389

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  5 in total

1.  Accuracy of multislice CT angiography for the assessment of in-stent restenoses in the iliac arteries at reduced dose: a phantom study.

Authors:  K Perisinakis; E Manousaki; K Zourari; D Tsetis; A Tzedakis; A Papadakis; A Karantanas; J Damilakis
Journal:  Br J Radiol       Date:  2011-03       Impact factor: 3.039

Review 2.  An overview on the advances in cardiovascular interventional MR imaging.

Authors:  Olaf Saborowski; Maythem Saeed
Journal:  MAGMA       Date:  2007-05-09       Impact factor: 2.310

3.  Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types.

Authors:  Michael Köhler; Matthias C Burg; Alexander C Bunck; Walter Heindel; Harald Seifarth; David Maintz
Journal:  Radiol Res Pract       Date:  2011-07-18

4.  MR Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types.

Authors:  Matthias C Burg; Alexander C Bunck; Harald Seifarth; Boris Buerke; Harald Kugel; Volker Hesselmann; Michael Köhler; Walter Heindel; David Maintz
Journal:  Radiol Res Pract       Date:  2010-07-27

5.  Optimization of MR Parameters of 3D TOF-MRA for Various Intracranial Stents at 3.0T MRI.

Authors:  Jin Woo Choi; Hong Gee Roh; Won-Jin Moon; Young Il Chun; Chung Hwan Kang
Journal:  Neurointervention       Date:  2011-08-31