| Literature DB >> 22089023 |
A Kyle Jones1, Alexander S Pasciak.
Abstract
While direct measurement of the peak skin dose resulting from a fluoroscopically-guided procedure is possible, the decision must be made a priori at additional cost and time. It is most often the case that the need for accurate knowledge of the peak skin dose is realized only after a procedure has been completed, or after a suspected reaction has been discovered. Part I of this review article discusses methods for calculating the peak skin dose across a range of clinical scenarios. In some cases, a wealth of data are available, while in other cases few data are available and additional data must be measured in order to estimate the peak skin dose. Data may be gathered from a dose report, the DICOM headers of images, or from staff and physician interviews. After data are gathered, specific steps must be followed to convert dose metrics, such as the reference point air kerma (K(a,r)) or the kerma area product (KAP), into peak skin dose. These steps require knowledge of other related factors, such as the f-factor and the backscatter factor, tables of which are provided in this manuscript. Sources of error and the impact of these errors on the accuracy of the final estimate of the peak skin dose are discussed.Entities:
Mesh:
Year: 2011 PMID: 22089023 PMCID: PMC5718743 DOI: 10.1120/jacmp.v12i4.3670
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Abbreviations used in this manuscript.
|
|
|
|
|---|---|---|
| Peak skin dose | PSD | The highest dose to a single area of the skin, units of Gy. |
| Fluoroscopically‐guided intervention | FGI | Any medical intervention using fluoroscopy for guidance. |
| Reference point air kerma |
| Cumulative air kerma at the interventional reference point, units of Gy. |
| Kerma area product | KAP | Product of air kerma and X‐ray field size, units of Gy‐cm2. |
| Digital Imaging and Communications in Medicine | DICOM | Standard for communication used by medical imaging equipment. |
| Interventional reference point | IRP | Point located 15 cm back towards the focal spot from the isocenter of a C‐arm fluoroscope(
|
| Source‐to‐image distance | SID | Distance from the focal spot to the center of the image receptor, units of mm. |
| Field of view | FOV | Size of the X‐ray field at the image receptor, units of cm. |
| Entrance skin air kerma | ESAK | Air kerma measured at the entrance surface of the patient, units of Gy. |
| Source‐to‐patient distance | SPD | Distance from the focal spot to the entrance surface of the patient, units of mm. |
| Backscatter factor | BSF | Factor that is applied to calculate entrance surface dose from the ESAK, accounts for the fact that many X‐rays of diagnostic energy are backscattered in tissue, unit‐less. |
| f‐factor | f | Factor used to convert exposure to dose in tissue, units of Gy/R. A similar, unit‐less quantity can be used to convert from air kerma to dose in tissue. |
| Digital acquisition series | DAS | A series of images generated using digital acquisition imaging. |
| Fluoroscopic reference point air kerma |
| Total reference point air kerma resulting from fluoroscopic imaging, units of Gy. |
| Digital acquisition reference point air kerma |
| Total reference point air kerma resulting from digital acquisition imaging, units of Gy. |
| Projected X‐ray field size on the skin |
| The area of the projected X‐ray field at the entrance surface of the patient, units of cm2. |
Data that can be mined from the DICOM header. Data available will vary by manufacturer and model.
|
|
|
|
|
|---|---|---|---|
| kVp | (0018,0060) | X‐RAY ACQUISITION MODULE, X‐RAY ACQUISITION DOSE MODULE, XA/XRF ACQUISITION MODULE | Peak kilovoltage output of X‐ray generator. |
| Intensifier size | (0018,1162) | X‐RAY ACQUISITION, X‐RAY IMAGE INTENSIFIER MODULE | Diameter of X‐ray intensifier in mm. |
| Field of view dimension(s) | (0018,1149) | X‐RAY ACQUISITION | Dimensions of image intensifier field of view in mm. If rectangle, row followed by column; if round, diameter. |
| Imager pixel spacing | (0018,1164) | X‐RAY ACQUISITION | Physical distance between pixels measured at front plane of image receptor housing: row spacing(delimiter)column spacing. |
| Collimator shape | (0018,1700) | X‐RAY COLLIMATOR MODULE | Shape of collimator: RECTANGULAR, CIRCULAR, POLYGONAL |
| Collimator left vertical edge | (0018,1702) | X‐RAY COLLIMATOR DIMENSIONS MACRO | Required if |
| Collimator right vertical edge | (0018,1704) | X‐RAY COLLIMATOR DIMENSIONS MACRO | Required if |
| Collimator upper horizontal edge | (0018,1706) | X‐RAY COLLIMATOR DIMENSIONS MACRO | Required if |
| Collimator lower horizontal edge | (0018,1708) | X‐RAY COLLIMATOR DIMENSIONS MACRO | Required if |
| Center of circular collimator | (0018,1710) | X‐RAY COLLIMATOR DIMENSIONS MACRO | Required if |
| Radius of circular collimator | (0018,1712) | X‐RAY COLLIMATOR DIMENSIONS MACRO | Required if |
| Vertices of polygonal collimator | (0018,1720) | X‐RAY COLLIMATOR DIMENSIONS MACRO | Required if |
| Distance Source to Patient | (0018,1111) | XA POSITIONER MODULE, XRF POSITIONER MODULE, X‐RAY ACQUISITION DOSE MODULE | Distance in mm from source to isocenter. NOTE: May be distance from source to table for some manufacturers. |
| Distance Source to Detector | (0018,1110) | XA POSITIONER MODULE, XRF POSITIONER MODULE, X‐RAY ACQUISITION DOSE MODULE | Distance in mm from source to detector center. |
| Estimated Radiographic Magnification Factor | (0018,1114) | XA POSITIONER MODULE and XRF POSITIONER MODULE | Ratio of Source Image Receptor Distance (SID) over Source Object Distance |
| Positioner Primary Angle | (0018,1510) | XA POSITIONER MODULE | Position of the X‐ray Image Intensifier about the patient from the RAO to LAO direction where movement from RAO to vertical is positive. |
| Positioner Secondary Angle | (0018,1511) | XA POSITIONER MODULE | Position of the X‐ray Image Intensifier about the patient from the CAU to CRA direction where movement from CAU to vertical is positive. |
| Image and Fluoroscopy Area Dose Product | (0018,115E) | X‐RAY ACQUISITION DOSE MODULE | Dose area product in dGy |
| Acquired Image Area Dose Product | (0018,9473) | XA/XRF ACQUISITION MODULE | Dose area product to which the patient was exposed in dGy |
| Number of frames | (0028,0008) | MULTI‐FRAME MODULE ATTRIBUTES | Number of frames in a multi‐frame image. |
Other information may be available in Modality Performed Procedure Step, including the Radiation Dose Module
Figure 1Methods for accessing data in the DICOM header: (a) Section of DICOM header as displayed using PACS, (b) section of DICOM header as displayed using ImageJ, (c) section of DICOM header as displayed using MATLAB.
Figure 2Format of a proprietary dose report.
Figure 3Backscatter factor as a function of patient size for a range of beam qualities. Red shading indicates typical values for fluoroscopy, green shading indicates typical values for digital acquisition imaging. To find the BSF for a given patient size and imaging mode: 1. Follow the value for the patient size across to the appropriate shaded region; 2. Follow the shaded region to the Half Value Layer axis; 3. Draw a line from the intersection of the shaded region with the Half Value Layer axis to the X‐ray field size used; 4. Draw a line from the X‐ray field size curve to the BSF axis to find the BSF. Data used to generate figure were taken from ICRU.( )
Values of the f‐factor for ICRU soft tissue for beam qualities commonly encountered in fluoroscopy.
|
|
|
|---|---|
| 3.0 – 3.5 | 1.058 |
| 3.5 – 4.0 | 1.059 |
| 4.0 – 4.5 | 1.061 |
| 4.5 – 5.0 | 1.062 |
| 5.0 – 5.5 | 1.062 |
| 5.5 – 6.0 | 1.063 |
| 6.0 – 6.5 | 1.066 |
| 6.5 – 7.0 | 1.068 |
Values of the f‐factor for ICRU soft tissue as a function of kVp, for cases where the HVL is not known.
|
| ||
|---|---|---|
|
|
|
|
| 60 | 1.061 | 1.056 |
| 65 | 1.063 | 1.058 |
| 70 | 1.065 | 1.059 |
| 75 | 1.066 | 1.061 |
| 85 | 1.068 | 1.063 |
| 95 | 1.069 | 1.066 |
Figure 4Flowchart outlining the basic steps in calculating the PSD from . A similar workflow would be used to calculate the PSD from the KAP, with the added step of converting from KAP to by dividing by the X‐ray field size.