| Literature DB >> 25679159 |
Maria D Falco1, Salvatore Masala, Matteo Stefanini, Roberto Fiori, Roberto Gandini, Paolo Bagalà, Daniele Morosetti, Eros Calabria, Alessia Tonnetti, Gianluca Verona-Rinati, Riccardo Santoni, Giovanni Simonetti.
Abstract
The purpose of this work has been to dosimetrically investigate four fluoroscopically guided interventions: the percutaneous vertebroplasty (PVP), the percutaneous disc decompression (PDD), the radiofrequency medial branch neurolysis (RF) (hereafter named spine procedures), and the endovascular treatment for the critical limb ischemia (CLI). The X-ray equipment used was a Philips Integris Allura Xper FD20 imaging system provided with a dose-area product (DAP) meter. The parameters investigated were: maximum skin dose (MSD), air kerma (Ka,r), DAP, and fluoroscopy time (FT). In order to measure the maximum skin dose, we employed a system based on MOSFET detectors. Before using the system on patients, a calibration factor Fc and correction factors for energy (CkV) and field size (CFD) dependence were determined. Ka,r, DAP, and FT were extrapolated from the X-ray equipment. The analysis was carried out on 40 patients, 10 for each procedure. The average fluoroscopy time and DAP values were compared with the reference levels (RLs) proposed in literature. Finally, the correlations between MSD, FT, Ka,r, and DAP values, as well as between DAP and FT values, were studied in terms of Pearson's product-moment coefficients for spine procedures only. An Fc value of 0.20 and a very low dependence of CFD on field size were found. A third-order polynomial function was chosen for CkV. The mean values of MSD ranged from 2.3 to 10.8cGy for CLI and PVP, respectively. For these procedures, the DAP and FT values were within the proposed RL values. The statistical analysis showed little correlation between the investigated parameters. The interventional procedures investigated were found to be both safe with regard to deterministic effects and optimized for stochastic ones. In the spine procedures, the observed correlations indicated that the estimation of MSD from Ka,r or DAP was not accurate and a direct measure of MSD is therefore recommended.Entities:
Mesh:
Year: 2015 PMID: 25679159 PMCID: PMC5689972 DOI: 10.1120/jacmp.v16i1.5020
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1X‐ray equipment in the angiographic room for the in phantom calibration in the reference setting: tube rotation angle at 180°, SSD on the couch of 50 cm (couch thickness 9 cm), and SID 120 cm.
Figure 2Energy correction factors as a function of the tube voltage in reference conditions. A third‐grade polynomial fit was superimposed. The error propagation formula was used to calculate the error bars, taking into account the uncertainty of the device (4%) for the UNFORS, and the reproducibility at the corresponding dose value (4%) for the MOSFET reading.
Figure 3MOSFET reading in reference conditions vs. the dose measured by the reference detector. We considered the uncertainty of the device (4%) for the UNFORS, and the reproducibility at the corresponding dose value (4%) for the MOSFET reading. In this figure, only the error bars relative to MOSFET reading were shown.
Measured data for spine interventional procedures on 30 patients (10 for procedure)
|
| |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
|
|
|
| PVP | 70 | AP | 90 (71‐120) | 14 (13‐16) | 103 (71‐120) | 16 (13‐20) | 17.2 (6.8‐25.2) | 0 |
| 1106.4 (402‐2610) | 114.6 (42.2‐240.5) |
| LL | 107 (85‐120) | 17 (14‐20) | 2 |
| |||||||
| PDD | 56 | AP | 95 (85‐120) | 15 (14‐16) | 110 (85‐120) | 16 (14‐19) | 4.6 (1.9‐13.2) | 0 |
| 442.5 (96.3‐1695.5) | 25.4 (11.8‐49.5) |
| LL | 115 (90‐120) | 17 (14‐19) | 0 |
| |||||||
| RF | 74 | AP | 108 (80‐120) | 16 (15‐18) | 5.9 (2‐9.1) | 0 |
| 306.3 (125‐554.3) | 30.5 (14.1‐53.3) | ||
; ; ; ; ; .
Measured data for critical limb ischemia treatments on 10 patients
|
|
| |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
| |
| FEMUR | 67 (60‐81) | 11 (7‐21) | 72 (65‐80) | 10 (7‐18) |
|
| ||||
| KNEE | 65 (58‐75) | 10 (6‐15) | 72 (65‐80) | 9 (5‐12) | 41.1 (14.5‐139.6) | 125 |
|
| 485.8 (112.9‐2160) | 74.6 (29.9‐245.0) |
|
|
| |||||||||
| ANKLE | 64 (52‐70) | 7,5 (5‐12) | 72 (65‐80) | 5 (2‐9) |
|
| ||||
; ; ; ; ; ; .
coefficients (95% CI) and p‐values for the Pearson's test
|
|
| |
|---|---|---|
|
| 0.76 |
|
|
| 0.29 | 0.16 |
|
| 0.40 |
|
|
| 0.46 |
|
; ; ; ; .
Figure 4Measured DAP for all the investigated spine interventional procedures as a function of the fluoroscopy time. The continuous line represents the linear best fit.
Figure 5Measured DAP for all the investigated spine interventional procedures as a function of the MSD. The continuous line represents the linear best fit.
Figure 6Measured Ka,r for all the investigated spine interventional procedures as a function of the MSD. The continuous line represents the linear best fit.
Comparison between our data and those reported in the literature for the investigated procedures
|
|
| ||||||||
|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
|
| PVP | 34 | 10 | (90‐100) | (6‐15) | 10.4 (3‐28) |
| 41.0 (10.5‐117.0) |
|
|
| 35 | 11 | (80‐110) | (2.5‐4.5) | 27.7 (18.8‐43.1) |
| ‐ |
| ||
| 36 | 10 | ‐ | ‐ | 16.5 (10‐34) |
| 54 (14‐133) |
| ‐ | |
| 32 | 61 | ‐ | ‐ | 17.4 (4.2‐54) | 82 | 77.6 (6.6–335.3) | 68.4 | ||
| 37 | 16 | ‐ | ‐ | ‐ | ‐ | ‐ |
|
| |
| This work | 10 | 103 | 16 | 17.2 (6.8‐25.2) | 2 | 114.6 (42.2‐240.5) |
|
| |
| RF | 38 | 3 | 63.3 (54‐93) | ‐ | 0.192 (1.4s‐30.0s) | ‐ | ‐ | 1.4 (0.6‐2.7) | |
| 39 | 16 | ‐ | ‐ | 0.212 (6s‐23s) | ‐ | ‐ | ‐ | ||
| This work | 10 | 101 | 16 | 5.9 (2‐9.1) | 0 | 30.5 (14.1‐53.3) |
| ||
| CLI | 41 | 8 | 69 (54–78) | ‐ | 2.2 (0.1–6.3) | 36 | 18 (0.2–43) | 3.6 (0.1–15.9) | |
| 40 | 96 | ‐ | ‐ | 16 (2.8–51.1) | ‐ | 140.5 (5.9‐1506.4) | ‐ | ||
| This work | 10 | 65 | 9.5 | 41.1 (14.5‐139.6) | 125 | 74.6 (29.9‐245.0) |
| ||
Values averaged, for each part of the body under investigation, over 10 patients.
b Peak skin dose (PSD) defined as the highest air kerma at any portion of a patient's skin during a procedure.
c The values refer to therapeutic procedure.
; ; , ; .