A tracking system to calculate patient skin dose in real-time during neurointerventional procedures using a biplane x-ray imaging system.

V K Rana1, S Rudin2, D R Bednarek2.   

Abstract

PURPOSE: Neurovascular interventional procedures using biplane fluoroscopic imaging systems can lead to increased risk of radiation-induced skin injuries. The authors developed a biplane dose tracking system (Biplane-DTS) to calculate the cumulative skin dose distribution from the frontal and lateral x-ray tubes and display it in real-time as a color-coded map on a 3D graphic of the patient for immediate feedback to the physician. The agreement of the calculated values with the dose measured on phantoms was evaluated.
METHODS: The Biplane-DTS consists of multiple components including 3D graphic models of the imaging system and patient, an interactive graphical user interface, a data acquisition module to collect geometry and exposure parameters, the computer graphics processing unit, and functions for determining which parts of the patient graphic skin surface are within the beam and for calculating dose. The dose is calculated to individual points on the patient graphic using premeasured calibration files of entrance skin dose per mAs including backscatter; corrections are applied for field area, distance from the focal spot and patient table and pad attenuation when appropriate. The agreement of the calculated patient skin dose and its spatial distribution with measured values was evaluated in 2D and 3D for simulated procedure conditions using a PMMA block phantom and an SK-150 head phantom, respectively. Dose values calculated by the Biplane-DTS were compared to the measurements made on the phantom surface with radiochromic film and a calibrated ionization chamber, which was also used to calibrate the DTS. The agreement with measurements was specifically evaluated with variation in kVp, gantry angle, and field size.
RESULTS: The dose tracking system that was developed is able to acquire data from the two x-ray gantries on a biplane imaging system and calculate the skin dose for each exposure pulse to those vertices of a patient graphic that are determined to be in the beam. The calculations are done in real-time with a typical graphic update time of 30 ms and an average vertex separation of 3 mm. With appropriate corrections applied, the Biplane-DTS was able to determine the entrance dose within 6% and the spatial distribution of the dose within 4% compared to the measurements with the ionization chamber and film for the SK150 head phantom. The cumulative dose for overlapping fields from both gantries showed similar agreement.
CONCLUSIONS: The Biplane-DTS can provide a good estimate of the peak skin dose and cumulative skin dose distribution during biplane neurointerventional procedures. Real-time display of this information should help the physician manage patient dose to reduce the risk of radiation-induced skin injuries.

Entities:  

Mesh:

Year:  2016        PMID: 27587043      PMCID: PMC4991993          DOI: 10.1118/1.4960368

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  Temporary hair loss simulating alopecia areata after endovascular surgery of cerebral arteriovenous malformations: a report of 3 cases.

Authors:  A Tosti; B M Piraccini; G Alagna
Journal:  Arch Dermatol       Date:  1999-12

2.  Skin dose mapping for fluoroscopically guided interventions.

Authors:  Perry B Johnson; David Borrego; Stephen Balter; Kevin Johnson; Daniel Siragusa; Wesley E Bolch
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

3.  Characteristics of Gafchromic XR-RV2 radiochromic film.

Authors:  Andrew Blair; Juergen Meyer
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

4.  Reference air kerma and kerma-area product as estimators of peak skin dose for fluoroscopically guided interventions.

Authors:  Deukwoo Kwon; Mark P Little; Donald L Miller
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

5.  Verification of the performance accuracy of a real-time skin-dose tracking system for interventional fluoroscopic procedures.

Authors:  Daniel R Bednarek; Jeffery Barbarits; Vijay K Rana; Srikanta P Nagaraja; Madhur S Josan; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2011-02-13

6.  Validation and initial clinical use of automatic peak skin dose localization with fluoroscopic and interventional procedures.

Authors:  Yasaman Khodadadegan; Muhong Zhang; William Pavlicek; Robert G Paden; Brian Chong; Eric A Huettl; Beth A Schueler; Kenneth A Fetterly; Steve G Langer; Teresa Wu
Journal:  Radiology       Date:  2012-11-09       Impact factor: 11.105

7.  Measuring skin dose with radiochromic dosimetry film in the cardiac catheterization laboratory.

Authors:  Edwin R Giles; Paul H Murphy
Journal:  Health Phys       Date:  2002-06       Impact factor: 1.316

Review 8.  Fluoroscopically guided interventional procedures: a review of radiation effects on patients' skin and hair.

Authors:  Stephen Balter; John W Hopewell; Donald L Miller; Louis K Wagner; Michael J Zelefsky
Journal:  Radiology       Date:  2010-02       Impact factor: 11.105

Review 9.  Methods for measuring fluoroscopic skin dose.

Authors:  Stephen Balter
Journal:  Pediatr Radiol       Date:  2006-09

Review 10.  Calculating the peak skin dose resulting from fluoroscopically guided interventions. Part I: Methods.

Authors:  A Kyle Jones; Alexander S Pasciak
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

View more
  17 in total

1.  Monte Carlo investigation of backscatter point spread function for X-ray imaging examinations.

Authors:  Zhenyu Xiong; Sarath Vijayan; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-09

2.  Calculation of the entrance skin dose distribution for fluoroscopically guided interventions using a pencil beam backscatter model.

Authors:  Sarath Vijayan; Zhenyu Xiong; Stephen Rudin; Daniel R Bednarek
Journal:  J Med Imaging (Bellingham)       Date:  2017-06-14

3.  Comparison of skin dose calculated by the dose tracking system (DTS) with a beam angular correction factor and that calculated by Monte-Carlo.

Authors:  Sheng-Hsuan Sun; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-02-15

4.  The effect of underlying bone on the beam angular correction in calculating the skin dose of the head in neuro-interventional imaging.

Authors:  Sheng-Hsuan Sun; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-02-15

5.  A Patient Dose-Reduction Technique for Neuroendovascular Image-Guided Interventions: Image-Quality Comparison Study.

Authors:  A Sonig; S V Setlur Nagesh; V S Fennell; S Gandhi; L Rangel-Castilla; C N Ionita; K V Snyder; L N Hopkins; D R Bednarek; S Rudin; A H Siddiqui; E I Levy
Journal:  AJNR Am J Neuroradiol       Date:  2018-02-15       Impact factor: 3.825

6.  A parametric fitting technique for rapid determination of a skin-dose correction factor for angle of beam incidence during image-guided endovascular procedures.

Authors:  Sheng-Hsuan Sun; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-02-15

7.  High-Definition Zoom Mode: A High Resolution X-ray Microscope for Neurointerventional Treatment Procedures.

Authors:  Swetadri Vasan Setlur Nagesh; Kunal Vakharia; Muhammad Waqas; Vernard S Fennell; Gursant S Atwal; Hussain Shallwani; Daniel R Bednarek; Jason M Davies; Kenneth V Snyder; Maxim Mokin; Stephen Rudin; Elad I Levy; Adnan H Siddiqui
Journal:  J Neuroimaging       Date:  2019-07-24       Impact factor: 2.486

8.  Estimation of Patient Eye-Lens Dose During Neuro-Interventional Procedures using a Dense Neural Network (DNN).

Authors:  J Collins; S Sun; C Guo; A Podgorsak; S Rudin; D R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-02-15

9.  Organ doses in pediatric patients undergoing cardiac-centered fluoroscopically guided interventions: Comparison of three methods for computational phantom alignment.

Authors:  Emily L Marshall; David Borrego; James C Fudge; Dhanashree Rajderkar; Wesley E Bolch
Journal:  Med Phys       Date:  2018-06-13       Impact factor: 4.071

10.  Developing a database of 3-D scattered radiation distributions for a c-arm fluoroscope as a function of exposure parameters and phantom.

Authors:  Chao Guo; Zhenyu Xiong; Sarath Vijayan; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.