Analysis of Fourier-domain task-based detectability index in tomosynthesis and cone-beam CT in relation to human observer performance.

Grace J Gang1, Junghoon Lee, J Webster Stayman, Daniel J Tward, W Zbijewski, Jerry L Prince, Jeffrey H Siewerdsen.   

Abstract

PURPOSE: Design and optimization of medical imaging systems benefit from accurate theoretical modeling that identifies the physical factors governing image quality, particularly in the early stages of system development. This work extends Fourier metrics of imaging performance and detectability index (d') to tomosynthesis and cone-beam CT (CBCT) and investigates the extent to which d' is a valid descriptor of task-based imaging performance as assessed by human observers,
METHODS: The detectability index for tasks presented in 2D slices (d'(slice)) was derived from 3D cascaded systems analysis of tomosynthesis and CBCT. Anatomical background noise measured in a physical phantom presenting power-law spectral density was incorporated in the "generalized" noise-equivalent quanta. Theoretical calculations of d'(slice) were performed as a function of total angular extent (theta(tot)) of source-detector orbit ranging 10 degrees - 360 degrees under two acquisition schemes: (i) Constant angular separation between projections (constant-delta theta), giving variable number of projections (N(proj)) and dose vs theta(tot) and (ii) constant number of projections (constant-N(proj)), giving constant dose (but variable angular sampling) with theta(tot). Five simple observer models were investigated: Prewhitening (PW), prewhitening with eye filter and internal noise (PWEi), nonprewhitening (NPW), nonprewhitening with eye filter (NPWE), and nonprewhitening with eye filter and internal noise (NPWEi). Human observer performance was measured in 9AFC tests for five simple imaging tasks presented within uniform and power-law clutter backgrounds. Measurements (from 9AFC tests) and theoretical calculations (from cascaded systems analysis of d'(slice)) were compared in terms of area under the ROC curve (A(z))
RESULTS: Reasonable correspondence between theoretical calculations and human observer performance was achieved for all imaging tasks over the broad range of experimental conditions and acquisition schemes. The PW and PWEi observer models tended to overestimate detectability, while the various NPW models predicted observer performance fairly well, with NPWEi giving the best overall agreement. Detectability was shown to increase with theta(tot) due to the reduction of out-of-plane clutter, reaching a plateau after a particular theta(tot) that depended on the imaging task. Depending on the acquisition scheme, however (i.e., constant-N(proj) or delta theta), detectability was seen in some cases to decline at higher theta(tot) due to tradeoffs among quantum noise, background clutter, and view sampling.
CONCLUSIONS: Generalized detectability index derived from a 3D cascaded systems model shows reasonable correspondence with human observer performance over a fairly broad range of imaging tasks and conditions, although discrepancies were observed in cases relating to orbits intermediate to 180 degrees and 360 degrees. The basic correspondence of theoretical and measured performance supports the application of such a theoretical framework for system design and optimization of tomosynthesis and CBCT.

Entities:  

Mesh:

Year:  2011        PMID: 21626910      PMCID: PMC3069989          DOI: 10.1118/1.3560428

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  50 in total

1.  Characterizing human perceptual inefficiencies with equivalent internal noise.

Authors:  Z L Lu; B A Dosher
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  1999-03       Impact factor: 2.129

2.  Effect of random background inhomogeneity on observer detection performance.

Authors:  J P Rolland; H H Barrett
Journal:  J Opt Soc Am A       Date:  1992-05       Impact factor: 2.129

3.  Generalized DQE analysis of radiographic and dual-energy imaging using flat-panel detectors.

Authors:  S Richard; J H Siewerdsen; D A Jaffray; D J Moseley; B Bakhtiar
Journal:  Med Phys       Date:  2005-05       Impact factor: 4.071

4.  Objective assessment of image quality: effects of quantum noise and object variability.

Authors:  H H Barrett
Journal:  J Opt Soc Am A       Date:  1990-07       Impact factor: 2.129

5.  Objective assessment of image quality. III. ROC metrics, ideal observers, and likelihood-generating functions.

Authors:  H H Barrett; C K Abbey; E Clarkson
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  1998-06       Impact factor: 2.129

6.  Digital radiology using active matrix readout of amorphous selenium: theoretical analysis of detective quantum efficiency.

Authors:  W Zhao; J A Rowlands
Journal:  Med Phys       Date:  1997-12       Impact factor: 4.071

7.  Signal, noise power spectrum, and detective quantum efficiency of indirect-detection flat-panel imagers for diagnostic radiology.

Authors:  J H Siewerdsen; L E Antonuk; Y el-Mohri; J Yorkston; W Huang; I A Cunningham
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

8.  Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology.

Authors:  J H Siewerdsen; L E Antonuk; Y el-Mohri; J Yorkston; W Huang; J M Boudry; I A Cunningham
Journal:  Med Phys       Date:  1997-01       Impact factor: 4.071

9.  Visual signal detectability with two noise components: anomalous masking effects.

Authors:  A E Burgess; X Li; C K Abbey
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  1997-09       Impact factor: 2.129

10.  Objective assessment of image quality. II. Fisher information, Fourier crosstalk, and figures of merit for task performance.

Authors:  H H Barrett; J L Denny; R F Wagner; K J Myers
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  1995-05       Impact factor: 2.129

View more
  48 in total

1.  Task-based modeling and optimization of a cone-beam CT scanner for musculoskeletal imaging.

Authors:  P Prakash; W Zbijewski; G J Gang; Y Ding; J W Stayman; J Yorkston; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

2.  Cascaded systems analysis of noise and detectability in dual-energy cone-beam CT.

Authors:  Grace J Gang; Wojciech Zbijewski; J Webster Stayman; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

3.  Effects of Image Quality on the Fundamental Limits of Image Registration Accuracy.

Authors:  Michael D Ketcha; Tharindu De Silva; Runze Han; Ali Uneri; Joseph Goerres; Matthew W Jacobson; Sebastian Vogt; Gerhard Kleinszig; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2017-07-11       Impact factor: 10.048

4.  Effects of automatic tube potential selection on radiation dose index, image quality, and lesion detectability in pediatric abdominopelvic CT and CTA: a phantom study.

Authors:  Michael F Brinkley; Juan C Ramirez-Giraldo; Ehsan Samei; Daniel J Frush; Kingshuk Roy Choudhury; Joshua M Wilson; Olav I Christianson; Donald P Frush
Journal:  Eur Radiol       Date:  2015-05-20       Impact factor: 5.315

5.  The effect of amorphous selenium detector thickness on dual-energy digital breast imaging.

Authors:  Yue-Houng Hu; Wei Zhao
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

6.  Task-driven source-detector trajectories in cone-beam computed tomography: I. Theory and methods.

Authors:  J Webster Stayman; Sarah Capostagno; Grace J Gang; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-02

7.  Fundamental limits of image registration performance: Effects of image noise and resolution in CT-guided interventions.

Authors:  M D Ketcha; T de Silva; R Han; A Uneri; J Goerres; M Jacobson; S Vogt; G Kleinszig; J H Siewerdsen
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-03

8.  Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography.

Authors:  Justin Solomon; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2016-09-22

9.  Physics considerations in MV-CBCT multi-layer imager design.

Authors:  Yue-Houng Hu; Rony Fueglistaller; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Shedlock; Daniel Morf; Paul Baturin; Pascal Huber; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-06-20       Impact factor: 3.609

10.  Investigating simulation-based metrics for characterizing linear iterative reconstruction in digital breast tomosynthesis.

Authors:  Sean D Rose; Adrian A Sanchez; Emil Y Sidky; Xiaochuan Pan
Journal:  Med Phys       Date:  2017-09       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.