Task-based modeling and optimization of a cone-beam CT scanner for musculoskeletal imaging.

P Prakash1, W Zbijewski, G J Gang, Y Ding, J W Stayman, J Yorkston, J A Carrino, J H Siewerdsen.   

Abstract

PURPOSE: This work applies a cascaded systems model for cone-beam CT imaging performance to the design and optimization of a system for musculoskeletal extremity imaging. The model provides a quantitative guide to the selection of system geometry, source and detector components, acquisition techniques, and reconstruction parameters.
METHODS: The model is based on cascaded systems analysis of the 3D noise-power spectrum (NPS) and noise-equivalent quanta (NEQ) combined with factors of system geometry (magnification, focal spot size, and scatter-to-primary ratio) and anatomical background clutter. The model was extended to task-based analysis of detectability index (d') for tasks ranging in contrast and frequency content, and d' was computed as a function of system magnification, detector pixel size, focal spot size, kVp, dose, electronic noise, voxel size, and reconstruction filter to examine trade-offs and optima among such factors in multivariate analysis. The model was tested quantitatively versus the measured NPS and qualitatively in cadaver images as a function of kVp, dose, pixel size, and reconstruction filter under conditions corresponding to the proposed scanner.
RESULTS: The analysis quantified trade-offs among factors of spatial resolution, noise, and dose. System magnification (M) was a critical design parameter with strong effect on spatial resolution, dose, and x-ray scatter, and a fairly robust optimum was identified at M ∼ 1.3 for the imaging tasks considered. The results suggested kVp selection in the range of ∼65-90 kVp, the lower end (65 kVp) maximizing subject contrast and the upper end maximizing NEQ (90 kVp). The analysis quantified fairly intuitive results-e.g., ∼0.1-0.2 mm pixel size (and a sharp reconstruction filter) optimal for high-frequency tasks (bone detail) compared to ∼0.4 mm pixel size (and a smooth reconstruction filter) for low-frequency (soft-tissue) tasks. This result suggests a specific protocol for 1 × 1 (full-resolution) projection data acquisition followed by full-resolution reconstruction with a sharp filter for high-frequency tasks along with 2 × 2 binning reconstruction with a smooth filter for low-frequency tasks. The analysis guided selection of specific source and detector components implemented on the proposed scanner. The analysis also quantified the potential benefits and points of diminishing return in focal spot size, reduced electronic noise, finer detector pixels, and low-dose limits of detectability. Theoretical results agreed quantitatively with the measured NPS and qualitatively with evaluation of cadaver images by a musculoskeletal radiologist.
CONCLUSIONS: A fairly comprehensive model for 3D imaging performance in cone-beam CT combines factors of quantum noise, system geometry, anatomical background, and imaging task. The analysis provided a valuable, quantitative guide to design, optimization, and technique selection for a musculoskeletal extremities imaging system under development.

Entities:  

Mesh:

Year:  2011        PMID: 21992379      PMCID: PMC3208412          DOI: 10.1118/1.3633937

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  47 in total

1.  Visual signal detection in structured backgrounds. III. Calculation of figures of merit for model observers in statistically nonstationary backgrounds.

Authors:  F O Bochud; C K Abbey; M P Eckstein
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2000-02       Impact factor: 2.129

2.  Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2001-02       Impact factor: 4.071

3.  Micro-angiography for neuro-vascular imaging. II. Cascade model analysis.

Authors:  Arundhuti Ganguly; Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Med Phys       Date:  2003-11       Impact factor: 4.071

4.  Task-based assessment of breast tomosynthesis: effect of acquisition parameters and quantum noise.

Authors:  I Reiser; R M Nishikawa
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

5.  Optimization of dual-energy imaging systems using generalized NEQ and imaging task.

Authors:  S Richard; J H Siewerdsen
Journal:  Med Phys       Date:  2007-01       Impact factor: 4.071

6.  Optimization of image acquisition techniques for dual-energy imaging of the chest.

Authors:  N A Shkumat; J H Siewerdsen; A C Dhanantwari; D B Williams; S Richard; N S Paul; J Yorkston; R Van Metter
Journal:  Med Phys       Date:  2007-10       Impact factor: 4.071

7.  Experimental validation of a three-dimensional linear system model for breast tomosynthesis.

Authors:  Bo Zhao; Jun Zhou; Yue-Houng Hu; Thomas Mertelmeier; Jasmina Ludwig; Wei Zhao
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

8.  A spatial-frequency dependent quantum accounting diagram and detective quantum efficiency model of signal and noise propagation in cascaded imaging systems.

Authors:  I A Cunningham; M S Westmore; A Fenster
Journal:  Med Phys       Date:  1994-03       Impact factor: 4.071

9.  Human observer detection experiments with mammograms and power-law noise.

Authors:  A E Burgess; F L Jacobson; P F Judy
Journal:  Med Phys       Date:  2001-04       Impact factor: 4.071

10.  Effect of noise correlation on detectability of disk signals in medical imaging.

Authors:  K J Myers; H H Barrett; M C Borgstrom; D D Patton; G W Seeley
Journal:  J Opt Soc Am A       Date:  1985-10       Impact factor: 2.129

View more
  22 in total

1.  Cascaded systems analysis of noise and detectability in dual-energy cone-beam CT.

Authors:  Grace J Gang; Wojciech Zbijewski; J Webster Stayman; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

2.  Quantitative Assessment Of Bone And Joint Health On A Dedicated Extremities Cone-Beam CT System.

Authors:  Wojciech Zbijewski; Qian Cao; Steven Tilley; Alejandro Sisniega; J Webster Stayman; John Yorkston; Jeffrey H Siewerden
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-06       Impact factor: 2.924

3.  Effects of Image Quality on the Fundamental Limits of Image Registration Accuracy.

Authors:  Michael D Ketcha; Tharindu De Silva; Runze Han; Ali Uneri; Joseph Goerres; Matthew W Jacobson; Sebastian Vogt; Gerhard Kleinszig; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2017-07-11       Impact factor: 10.048

4.  Task-driven source-detector trajectories in cone-beam computed tomography: I. Theory and methods.

Authors:  J Webster Stayman; Sarah Capostagno; Grace J Gang; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-02

5.  Fundamental relationship between the noise properties of grating-based differential phase contrast CT and absorption CT: theoretical framework using a cascaded system model and experimental validation.

Authors:  Ke Li; Nicholas Bevins; Joseph Zambelli; Guang-Hong Chen
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

6.  Task-based detectability in CT image reconstruction by filtered backprojection and penalized likelihood estimation.

Authors:  Grace J Gang; J Webster Stayman; Wojciech Zbijewski; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

7.  Statistical model based iterative reconstruction (MBIR) in clinical CT systems. Part II. Experimental assessment of spatial resolution performance.

Authors:  Ke Li; John Garrett; Yongshuai Ge; Guang-Hong Chen
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

8.  Dedicated cone-beam CT system for extremity imaging.

Authors:  John A Carrino; Abdullah Al Muhit; Wojciech Zbijewski; Gaurav K Thawait; J Webster Stayman; Nathan Packard; Robert Senn; Dong Yang; David H Foos; John Yorkston; Jeffrey H Siewerdsen
Journal:  Radiology       Date:  2013-11-18       Impact factor: 11.105

9.  Task-driven optimization of CT tube current modulation and regularization in model-based iterative reconstruction.

Authors:  Grace J Gang; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Phys Med Biol       Date:  2017-03-31       Impact factor: 3.609

10.  A dedicated cone-beam CT system for musculoskeletal extremities imaging: design, optimization, and initial performance characterization.

Authors:  W Zbijewski; P De Jean; P Prakash; Y Ding; J W Stayman; N Packard; R Senn; D Yang; J Yorkston; A Machado; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.