The effect of amorphous selenium detector thickness on dual-energy digital breast imaging.

Yue-Houng Hu1, Wei Zhao1.   

Abstract

PURPOSE: Contrast enhanced (CE) imaging techniques for both planar digital mammography (DM) and three-dimensional (3D) digital breast tomosynthesis (DBT) applications requires x-ray photon energies higher than the k-edge of iodine (33.2 keV). As a result, x-ray tube potentials much higher (>40 kVp) than those typical for screening mammography must be utilized. Amorphous selenium (a-Se) based direct conversion flat-panel imagers (FPI) have been widely used in DM and DBT imaging systems. The a-Se layer is typically 200 μm thick with quantum detective efficiency (QDE) >87% for x-ray energies below 26 keV. However, QDE decreases substantially above this energy. To improve the object detectability of either CE-DM or CE-DBT, it may be advantageous to increase the thickness (dSe) of the a-Se layer. Increasing the dSe will improve the detective quantum efficiency (DQE) at the higher energies used in CE imaging. However, because most DBT systems are designed with partially isocentric geometries, where the gantry moves about a stationary detector, the oblique entry of x-rays will introduce additional blur to the system. The present investigation quantifies the effect of a-Se thickness on imaging performance for both CE-DM and CE-DBT, discussing the effects of improving photon absorption and blurring from oblique entry of x-rays.
METHODS: In this paper, a cascaded linear system model (CLSM) was used to investigate the effect of dSe on the imaging performance (i.e., MTF, NPS, and DQE) of FPI in CE-DM and CE-DBT. The results from the model are used to calculate the ideal observer signal-to-noise ratio, d', which is used as a figure-of-merit to determine the total effect of increasing dSe for CE-DM and CE-DBT.
RESULTS: The results of the CLSM show that increasing dSe causes a substantial increase in QDE at the high energies used in CE-DM. However, at the oblique projection angles used in DBT, the increased length of penetration through a-Se introduces additional image blur. The reduced MTF and DQE at high spatial frequencies lead to reduced two-dimensional d'. These losses in projection image resolution may subsequently result in a decrease in the 3D d', but the degree of which is largely dependent on the DBT reconstruction algorithm. For a filtered backprojection (FBP) algorithm with spectral apodization and slice-thickness filters, which dominate the blur for reconstructed images at oblique angles, the effect of oblique entry of x-rays on 3D d' is minimal. Thus, increasing dSe results in an improvement in d' for both CE-DM and CE-DBT with typical FBP reconstruction parameters.
CONCLUSIONS: Increased dSe improves CE breast imaging performance by increasing QDE of detectors at higher energies, e.g., 49 kVp. Although there is additional blur in the oblique angled projections of a DBT scan, the overall 3D d' for DBT is not degraded because the dominant source blur at these angles results from the reconstruction filters of the employed FBP algorithm.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25370637      PMCID: PMC4218697          DOI: 10.1118/1.4897244

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

1.  Generalized DQE analysis of radiographic and dual-energy imaging using flat-panel detectors.

Authors:  S Richard; J H Siewerdsen; D A Jaffray; D J Moseley; B Bakhtiar
Journal:  Med Phys       Date:  2005-05       Impact factor: 4.071

2.  Optimization of dual-energy imaging systems using generalized NEQ and imaging task.

Authors:  S Richard; J H Siewerdsen
Journal:  Med Phys       Date:  2007-01       Impact factor: 4.071

3.  Resolution at oblique incidence angles of a flat panel imager for breast tomosynthesis.

Authors:  James G Mainprize; Aili K Bloomquist; Michael P Kempston; Martin J Yaffe
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

4.  An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

5.  Digital radiology using active matrix readout of amorphous selenium: theoretical analysis of detective quantum efficiency.

Authors:  W Zhao; J A Rowlands
Journal:  Med Phys       Date:  1997-12       Impact factor: 4.071

6.  Digital radiology using active matrix readout of amorphous selenium: detectors with high voltage protection.

Authors:  W Zhao; J Law; D Waechter; Z Huang; J A Rowlands
Journal:  Med Phys       Date:  1998-04       Impact factor: 4.071

7.  Evaluation of tumor angiogenesis of breast carcinoma using contrast-enhanced digital mammography.

Authors:  Clarisse Dromain; Corrine Balleyguier; Serge Muller; Marie-Christine Mathieu; France Rochard; Paule Opolon; Robert Sigal
Journal:  AJR Am J Roentgenol       Date:  2006-11       Impact factor: 3.959

8.  Effects of characteristic x rays on the noise power spectra and detective quantum efficiency of photoconductive x-ray detectors.

Authors:  W Zhao; W G Ji; J A Rowlands
Journal:  Med Phys       Date:  2001-10       Impact factor: 4.071

9.  Digital mammography using iodine-based contrast media: initial clinical experience with dynamic contrast medium enhancement.

Authors:  Felix Diekmann; Susanne Diekmann; Fanny Jeunehomme; Serge Muller; Bernd Hamm; Ulrich Bick
Journal:  Invest Radiol       Date:  2005-07       Impact factor: 6.016

10.  Dual-energy contrast-enhanced digital subtraction mammography: feasibility.

Authors:  John M Lewin; Pamela K Isaacs; Virginia Vance; Fred J Larke
Journal:  Radiology       Date:  2003-07-29       Impact factor: 11.105

View more
  3 in total

1.  Physics considerations in MV-CBCT multi-layer imager design.

Authors:  Yue-Houng Hu; Rony Fueglistaller; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Shedlock; Daniel Morf; Paul Baturin; Pascal Huber; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-06-20       Impact factor: 3.609

2.  Evaluation of a hybrid direct-indirect active matrix flat-panel imager using Monte Carlo simulation.

Authors:  Scott Dow; Adrian Howansky; Anthony R Lubinsky; Wei Zhao
Journal:  J Med Imaging (Bellingham)       Date:  2020-05-12

3.  A novel method for quantification of beam's-eye-view tumor tracking performance.

Authors:  Yue-Houng Hu; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Morf; Daniel Shedlock; Paul Baturin; Josh Star-Lack; Ross Berbeco
Journal:  Med Phys       Date:  2017-10-13       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.