| Literature DB >> 21388521 |
Marijn P Rolf1, Mark B M Hofman, Peter D Gatehouse, Karin Markenroth-Bloch, Martijn W Heymans, Tino Ebbers, Martin J Graves, John J Totman, Beat Werner, Albert C van Rossum, Philip J Kilner, Rob M Heethaar.
Abstract
PURPOSE: Eddy current induced velocity offsets are of concern for accuracy in cardiovascular magnetic resonance (CMR) volume flow quantification. However, currently known theoretical aspects of eddy current behavior have not led to effective guidelines for the optimization of flow quantification sequences. This study is aimed at identifying correlations between protocol parameters and the resulting velocity error in clinical CMR flow measurements in a multi-vendor study.Entities:
Mesh:
Year: 2011 PMID: 21388521 PMCID: PMC3065419 DOI: 10.1186/1532-429X-13-18
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Figure 1Slice orientations. The pulmonary slice is rotated 45° from transverse to coronal. The oblique aortic slice is rotated 45° from transverse to sagittal. The transverse aortic slice is used for flow measurements through the aorta at the level of pulmonary artery bifurcation.
Parameters used for protocol testing.
| Parameter | GE Signa Excite | Philips Achieva | Siemens Avanto |
|---|---|---|---|
| Gradient speed | |||
| Partial Echo | on (75%)/ | ||
| Bandwidth (Hz/pix) | 200, 250, 300, | ||
| Venc (cm/s) | 120, | ||
| Slice Thickness (mm) | |||
| Delay Td (ms) | - | ||
| Velocity encoding | - | symmetric/ | |
| Slew rate (%) & amplitude (%) | - | - | 20, 40, 60, 80, |
Protocol parameters used for protocol testing listed per scanner type. In bold the basic protocol setting.
Figure 2Velocity offset as a function of several protocol parameters. Every data point represents an average of three samples of the same scanner-type. Standard deviation of the basic protocol was 0.4 cm/s (average for the three scanner-types). From these graphs it is clear that there are no general guidelines across different types for velocity offset optimization. Slice orientation, however, had a strong influence on the velocity offset; a transverse aortic slice gave generally the lowest offset. Results from statistical analysis of this data is shown in Table 2.
Results of regular protocol variation by multi-level analysis.
| systematic difference | |||||||
|---|---|---|---|---|---|---|---|
| oblique aortic - transverse aortic | pulmonary - transverse aortic | ||||||
| slope | P | (cm/s) | P | (cm/s) | P | ||
| Gradient Speed | -0.54 | cm/s | 0.01 | 0.38 | 0.25 | 1.37 | < 0.01 |
| Partial Echo | -0.93 | cm/s | 0.02 | 0.52 | 0.47 | 2.34 | < 0.01 |
| Bandwidth | 0.002 | cm/s/ Hz/pix | < 0.01 | 0.63 | 0.09 | 1.69 | < 0.01 |
| Venc | 0.0003 | cm/s/ cm/s | 0.59 | 0.77 | 0.09 | 1.79 | < 0.01 |
| Slice Thickness | 0.03 | cm/s/ mm | 0.07 | 0.76 | 0.03 | 1.85 | < 0.01 |
| average | 0.61 | 1.8 | |||||
Graphs corresponding to the data are shown in Figure 2. Some slopes (increase of velocity offset per change of protocol parameter) were significant, but in every case this effect was specific to just one of the three scanner-types (gradient speed, partial echo) or not large enough to be of use (bandwidth). Oblique aortic slices showed systematic 0.61 cm/s higher offset than transverse aortic slices but this difference was not always significant. Pulmonary slices showed systematic 1.8 cm/s higher offset than transverse aortic slices.
Figure 3Velocity offset as a function of time delay. Velocity offset as a function of a delay between the bipolar flow encoding gradient and the read-out gradient, T. Colors indicate different scanner samples. On the left, results of three samples of one scanner type, on the right, the results of three samples of another scanner type.
Figure 4Velocity offset with asymmetric and symmetric velocity encoding. Colors indicate different scanner samples. On the left, results of three samples of one type, on the right, the results of three samples of another type. Symmetric encoding was lower, but only significant in type 1 (type 1: -0.4 cm/s, P < 0.01, type 2: -0.6 cm/s, P = 0.40).
Figure 5Velocity offset as a function of bipolar gradient amplitude with varying slew rate. Lines represent linear fits to the data per slice orientation. Example from Siemens sample 2, complete multiple regression results are shown in Table 3.
Statistical results from exploratory measurements beyond the protocol.
| Siemens | sample 1 | sample 2 | sample 3 | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| transv. aortic | obl. aortic | pulm | transv. aortic | obl. aortic | pulm | transv. aortic | obl. aortic | pulm | ||
| gradient amplitude | cm/s per mT/m | 0.02 | 0.05 | 0.04 | 0.01 | 0.03 | 0.03 | 0.04 | 0.11 | 0.07 |
| slew rate | cm/s per mT/m/s | 2.28 | 0.85 | -0.86 | 2.23 | 1.23 | 0.14 | 0.85 | -0.34 | 0.3 |
| r2 | 0.57 | 0.49 | 0.47 | 0.66 | 0.73 | 0.61 | 0.81 | 0.89 | 0.81 | |
Results from beyond protocol variation by multiple regression of gradient amplitude and slew rate on velocity offset. All correlations found were highly significant (P < < 0.01) but not very strong and varied considerably among scanner samples of the same type. As an example, the data points of sample 2 are shown in Figure 5.