Peter D Gatehouse1, Marijn P Rolf, Martin J Graves, Mark Bm Hofman, John Totman, Beat Werner, Rebecca A Quest, Yingmin Liu, Jochen von Spiczak, Matthias Dieringer, David N Firmin, Albert van Rossum, Massimo Lombardi, Juerg Schwitter, Jeanette Schulz-Menger, Philip J Kilner.
Abstract
AIMS: Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error. We studied the severity of uncorrected velocity offset errors across sites and CMR systems. METHODS ANDEntities:
Mesh:
Substances:
Year: 2010 PMID: 20074359 PMCID: PMC2818657 DOI: 10.1186/1532-429X-12-5
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Figure 1A systolic frame of an aortic flow acquisition. (170 ms after R-wave, at Venc = 150 cm/s). (a) Signal magnitude image, (b) Phase contrast velocity image shown at normal greyscale settings (black = -150 cm/s, white = +150 cm/s) where there apparently uniformly grey chest wall fails to reveal the background offset error. The same image is therefore reprinted in (c) with more extreme greyscale contrast to show up the background offset errors (black ≤ -15 cm/s, white ≥ +15 cm/s) (d) Phase contrast image using identical sequence protocol, but of static gelatin phantom, displayed with same greyscale as (c), demonstrating the phase offset.
Figure 2Aortic flow 64 ml/beat measured from Figure 1. The background in the aortic region was measured in the phantom, as in Figure 1d. The aortic flow curve includes 8.4 ml/beat due to the background offset of 1.6 cm/s in the aortic region. The true aortic flow is 56 ml/beat. The relative error in the calculated flow measurement is therefore 15%. Although the example in Figure 1 may be relatively easy to correct by correcting phase offset errors of signal across the relatively large regions of static chest wall and liver, correction is not always as straightforward in clinical acquisitions. Without such an independent correction of the background offset, it would be difficult to correct the aortic flow curve by using physiological assumptions such as negligible flow in diastole.
Figure 3Coronal (upper) and sagittal (lower) illustrations of the oblique planes of velocity acquisition (thick black lines). The Aortic plane, for which an antero-posterior phase-encode direction was always chosen, was at 45° between transverse and sagittal. The main pulmonary artery (MPA) plane, which was acquired twice, either with left-right or head-foot phase-encode directions, was at 45° between transverse and coronal. Each plane passed through the isocentre. The dark grey region represents the uniform gelatin phantom, with dotted lines indicating a corresponding patient position with typical orientations of the aorta and MPA in which flows are typically measured through slices similar to those of this study.
Figure 4The flow offset analysis method. The mean phase offset over the smaller 30 mm diameter circle, which represents a typical "great vessel" ROI, was calculated for all positions of such ROIs with their centres located up to 50 mm from the isocentre of the magnet (up to 70 mm for MPA planes). The largest ROI value found was taken as the 'worst case' result for that plane of acquisition. All image planes passed through the isocentre. In this example, the largest error was found in the mean ROI positioned at the lower edge of the search region.
Largest mean uncorrected velocity offset (cm/s) in regions centred within 50 mm of isocentre for the aortic plane, and within 70 mm for the MPA plane (head-foot or left-right phase-encoding).
| cm/s | Scanner | Aorta | MPA | MPA |
|---|---|---|---|---|
| Scanner type 1 | 1 | -2.2 cm/s | -3.8 cm/s | -5.6 cm/s |
| 2 | -2.8 | -3.1 | -5.5 | |
| 3 | -0.7 | -3.2 | -5.3 | |
| 4 | -1.3 | -3.5 | -4.7 | |
| Scanner type 2 | 5 | -2.2 | -3.4 | -5.4 |
| 6 | -1.1 | -3.9 | -3.1 | |
| 7 | 1.6 | -4.0 | -3.6 | |
| 8 | -2.7 | -4.9 | -4.9 | |
| Scanner type 3 | 9 | 1.6 | 1.2 | 1.8 |
| 10 | 1.2 | -0.7 | 0.4 | |
| 11 | 0.9 | 1.6 | 2.1 | |
| 12 | 1.1 | -1.7 | 2.0 | |
| Mean of all 12 absolute values | 1.6 cm/s | 2.9 cm/s | 3.7 cm/s | |
| Range of all 12 absolute values | 0.7 to 2.8 cm/s | 0.7 to 4.9 cm/s | 0.4 to 5.6 cm/s | |
The twelve rows are from the four sites using each scanner type. The scanner types are not identified.
Figure 5Uncorrected velocity offset results from all 12 sites. The largest mean value of the ROI found anywhere up to 5 cm inplane from isocenter for the aortic slice (up to 7 cm for the two MPA slices). The green zone represents the 0.6 cm/s offset described in the Discussion section.