Gijs K A van Wermeskerken1, Adriaan F van Olphen, Kees Graamans.
Abstract
This study assessed the electrode position in cochlear implant patients and evaluated the extent to which the electrode position is determinative in the electrophysiological functioning of the cochlear implant system. Five consecutively implanted adult patients received a multichannel cochlear implant. In all patients, the electrical impedance and the electrically evoked compound action potentials were recorded immediately after implantation. Multislice computer tomography was performed 6 weeks postoperatively before switch-on of the cochlear implant. The electrode position relative to the modiolus was assessed and correlated to the electrophysiological measurements. All electrodes were fully inserted; this was confirmed by computer tomography. The individual electrode distance toward the modiolus could be most precisely analyzed for the basal part of the electrode array. It was thus decided to study the data of electrodes one, four, and seven. No correlation was found between electrical impedance and electrode distance. A significant correlation was found between electrode distance and the electrically evoked compound action potentials, with a 96% probability using Kendall's rank correlation. We conclude that the electrode-modiolus distance is of importance to the stimulation of auditory nerve fibers. Future developments in imaging will further improve and refine our insight in the relation between electrode positioning.Entities:
Mesh:
Year: 2009 PMID: 19308437 PMCID: PMC2734256 DOI: 10.1007/s00405-009-0939-2
Source DB: PubMed Journal: Eur Arch Otorhinolaryngol ISSN: 0937-4477 Impact factor: 2.503
Patients’ characteristics
| Male/Female | 2/3 |
| Mean age at implantation (range) | 58.8 (45–75) |
| Mean age of deafness (range) | 28.3 (4–45) |
ECAP test parameters
| Stimulation rate | 80 Hz |
| Pulse width | 25 μs |
| Masker level | +5 CL above probe level |
| Masker probe interval | 500 μs |
| Gain/number of sweeps | 60 dB/100 or 40 dB/200 |
| Delay | 100 μs |
Fig. 1HRCT image with an MPR in the plane of the electrode array
Fig. 2Distance electrode–modiolus
Fig. 3Electrode–modiolus distance versus ECAP