K Strach1, C Meyer, D Thomas, C P Naehle, C Schmitz, H Litt, A Bernstein, B Cheng, H Schild, T Sommer.
Abstract
The purpose of this study was to evaluate high-resolution (HR) myocardial first-pass perfusion in healthy volunteers at 3 T compared to a typical clinical imaging protocol at 1.5 T, with respect to overall image quality and the presence of subendocardial dark rim artifacts. Myocardial first-pass rest perfusion studies were performed at both field strengths using a T1-weighted saturation-recovery segmented k-space gradient-echo sequence combined with parallel imaging (Gd-DTPA 0.05 mmol/kg). Twenty-six healthy volunteers underwent (1) a HR perfusion scan at 3 T(pixel size 3.78 mm(2)) and (2) a standard perfusion approach at 1.5 T(pixel size 9.86 mm(2)). The contrast enhancement ratio (CER) and overall image quality (4-point grading scale: 4: excellent; 1: non-diagnostic) were assessed, and a semiquantitative analysis of dark rim artifacts was performed for all studies. CER was slightly higher (1.31 +/- 0.32 vs. 1.14 +/- 0.34; p<0.01), overall image quality was significantly improved (3.03 +/- 0.43 vs. 2.37 +/- 0.39; p<0.01), and the number of dark rim artifacts (139 +/- 2.09 vs. 243 +/- 2.33; p<0.01) was significantly reduced for HR perfusion imaging at 3 T compared to the standard approach at 1.5 T. HR myocardial rest perfusion at 3 T is superior to the typical clinical perfusion protocol performed at 1.5 T with respect to the overall image quality and presence of subendocardial dark rim artifacts.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17429650 DOI: 10.1007/s00330-006-0560-3
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315