| Literature DB >> 11108638 |
R Noeske1, F Seifert, K H Rhein, H Rinneberg.
Abstract
Using a two-element phased array receiver coil, single breath-hold, ECG gated cardiac images of signal-to-noise ratios up to 130 and contrast-to-noise ratios exceeding 35 between myocardium and blood were recorded at 3 T. At several locations within the myocardium, T*(2) and B(0) inhomogeneity were determined. Because of shorter T*(2) times and larger B(0) inhomogeneities attributable to enhanced susceptibility effects, real-time cardiac imaging, the use of spiral scans, and echo planar imaging are expected to be considerably more difficult at 3 T.Entities:
Mesh:
Year: 2000 PMID: 11108638 DOI: 10.1002/1522-2594(200012)44:6<978::aid-mrm22>3.0.co;2-9
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668