| Literature DB >> 15153237 |
Rupert M Pearse1, Kashif Ikram, John Barry.
Abstract
The LiDCO plus system is a minimally/non-invasive technique of continuous cardiac output measurement. In common with all cardiac output monitors this technology has both strengths and weaknesses. This review discusses the technological basis of the device and its clinical application.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15153237 PMCID: PMC468899 DOI: 10.1186/cc2852
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Strengths and weaknesses of the LiDCO™plus method of measuring cardiac output
| Strengths | Weaknesses |
|---|---|
| May be used in conscious and unconscious patients | Arterial waveform artefact may significantly affect data accuracy |
| May be calibrated by nursing or medical staff in 10 min | Irregular pulse rate may affect data accuracy |
| Provides dynamic markers of fluid responsiveness | Nondepolarizing muscle relaxants interfere with calibration |