PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery.

Timothy T Hamilton1, Lynne M Huber, Michael E Jessen.   

Abstract

BACKGROUND: Cardiac output is often monitored after cardiac operations with a pulmonary artery catheter. A new method has been introduced that measures cardiac output by lithium dilution (LiDCO) and uses these data to calibrate a system (PulseCO) that calculates cardiac output continuously from the energy of the arterial pressure waveform. It is unknown whether PulseCO measurements are valid early after cardiac surgery when changes in temperature and vascular tone or intermittent use of the arterial line for blood sampling may occur. This study assessed the reliability of cardiac output determinations by PulseCO in the first 8 hours after cardiac surgery.
METHODS: After a one-time PulseCO calibration, cardiac output was measured in 20 patients who had undergone coronary artery bypass grafting at 0, 2, 4, 6, and 8 hours after arrival in the intensive care unit using (1) thermodilution through a pulmonary artery catheter (Thermo); (2) lithium dilution (LiDCO); and (3) PulseCO. Concordance correlations were calculated between methods, and differences were compared by Wilcoxon paired rank test and Bland-Altman analysis.
RESULTS: Cardiac output ranged from 3.4 to 8.5 L/min. No significant differences were noted between measurements obtained by each technique at any time point. Concordance correlations and Bland-Altman analysis confirmed good agreement between PulseCO and Thermo determinations of cardiac output during the study interval.
CONCLUSIONS: PulseCO measurements remain reliable without recalibration for at least 8 hours after cardiac surgery and may offer a less-invasive approach for early postoperative cardiac output monitoring.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12400827     DOI: 10.1016/s0003-4975(02)04059-6

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  26 in total

1.  Real-time stroke volume measurements for the optimization of cardiac resynchronization therapy parameters.

Authors:  José M Dizon; T Alexander Quinn; Santos E Cabreriza; Daniel Wang; Henry M Spotnitz; Kathleen Hickey; Hasan Garan
Journal:  Europace       Date:  2010-06-04       Impact factor: 5.214

2.  Using a Calculated Pulse Rate with an Artificial Neural Network to Detect Irregular Interbeats.

Authors:  Bih-Chyun Yeh; Wen-Piao Lin
Journal:  J Med Syst       Date:  2015-12-07       Impact factor: 4.460

3.  Estimation of cardiac output in a pharmacological trial using a simple method based on arterial blood pressure signal waveform: a comparison with pulmonary thermodilution and echocardiographic methods.

Authors:  Jani Penttilä; Amir Snapir; Erkki Kentala; Juha Koskenvuo; Jussi Posti; Mika Scheinin; Harry Scheinin; Tom Kuusela
Journal:  Eur J Clin Pharmacol       Date:  2006-03-29       Impact factor: 2.953

4.  Subject-specific model estimation of cardiac output and blood volume during hemorrhage.

Authors:  Maxwell Lewis Neal; James B Bassingthwaighte
Journal:  Cardiovasc Eng       Date:  2007-09

5.  Pulse pressure: more than 100 years of changes in stroke volume.

Authors:  Maurizio Cecconi; Andrew Rhodes
Journal:  Intensive Care Med       Date:  2011-03-05       Impact factor: 17.440

Review 6.  Methods in pharmacology: measurement of cardiac output.

Authors:  Bart F Geerts; Leon P Aarts; Jos R Jansen
Journal:  Br J Clin Pharmacol       Date:  2011-03       Impact factor: 4.335

7.  Cross-comparisons of trending accuracies of continuous cardiac-output measurements: pulse contour analysis, bioreactance, and pulmonary-artery catheter.

Authors:  Bouchra Lamia; Hyung Kook Kim; Donald A Severyn; Michael R Pinsky
Journal:  J Clin Monit Comput       Date:  2017-02-10       Impact factor: 2.502

8.  Cardiac output measurement by arterial pressure waveform analysis during optimization of biventricular pacing after cardiac surgery.

Authors:  Justin H Booth; T Alexander Quinn; Marc E Richmond; Santos E Cabreriza; Alan D Weinberg; Taylor Johnston; Henry M Spotnitz
Journal:  ASAIO J       Date:  2009 Nov-Dec       Impact factor: 2.872

9.  Importance of re-calibration time on pulse contour analysis agreement with thermodilution measurements of cardiac output: a retrospective analysis of intensive care unit patients.

Authors:  Christopher G Scully; Shanti Gomatam; Shawn Forrest; David G Strauss
Journal:  J Clin Monit Comput       Date:  2015-08-19       Impact factor: 2.502

10.  Continuous and intermittent cardiac output measurement in hyperdynamic conditions: pulmonary artery catheter vs. lithium dilution technique.

Authors:  Maria Gabriella Costa; Giorgio Della Rocca; Paolo Chiarandini; Silvia Mattelig; Livia Pompei; Mauricio Sainz Barriga; Toby Reynolds; Maurizio Cecconi; Paolo Pietropaoli
Journal:  Intensive Care Med       Date:  2007-10-06       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.