Lithium dilution cardiac output measurement: a comparison with thermodilution.

R Linton1, D Band, T O'Brien, M Jonas, R Leach.   

Abstract

OBJECTIVE: To compare the results of cardiac output measurements obtained by lithium dilution and thermodilution.
DESIGN: Case series, observational study.
SETTING: High-dependency postoperative unit and intensive care unit of a teaching hospital. PATIENTS: Forty patients were studied. Thirty-four patients had undergoing heart surgery requiring cardiopulmonary bypass within the previous 2 days; the diagnoses in the other patients were myocardial infarct (n = 2), septicemia (n = 2), adult respiratory distress syndrome, and pericardectomy.
INTERVENTIONS: Cardiac output was measured five times in each patient, using lithium dilution (single measurement) and bolus thermodilution (series of three to six measurements according to standard clinical practice, taking the average of the closest three). In a subgroup of 14 patients, cardiac output was also measured using "continuous thermodilution."
MEASUREMENTS AND MAIN RESULTS: Comparing lithium dilution with bolus thermodilution, the mean of the differences (lithium dilution-thermodilution) was -0.25 +/- 0.46 [SD] L/min. Linear regression analysis gave y = 0.31 + 0.89x (r2 = .94) for lithium dilution vs. thermodilution.
CONCLUSIONS: The overall agreement between the two methods was good. The variability of the thermodilution measurements was greater than that of the lithium dilution measurements. The lithium dilution method is at least as accurate as bolus thermodilution and, since pulmonary artery catheterization is not needed, it has the advantages of being safe and quick to perform.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9366760     DOI: 10.1097/00003246-199711000-00015

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  17 in total

1.  Real-time stroke volume measurements for the optimization of cardiac resynchronization therapy parameters.

Authors:  José M Dizon; T Alexander Quinn; Santos E Cabreriza; Daniel Wang; Henry M Spotnitz; Kathleen Hickey; Hasan Garan
Journal:  Europace       Date:  2010-06-04       Impact factor: 5.214

Review 2.  [Measurement of cardiac output].

Authors:  D A Reuter; A E Goetz
Journal:  Anaesthesist       Date:  2005-11       Impact factor: 1.041

Review 3.  Cardiac output in 1998.

Authors:  M Singer
Journal:  Heart       Date:  1998-05       Impact factor: 5.994

Review 4.  Methods in pharmacology: measurement of cardiac output.

Authors:  Bart F Geerts; Leon P Aarts; Jos R Jansen
Journal:  Br J Clin Pharmacol       Date:  2011-03       Impact factor: 4.335

Review 5.  Minimally invasive or noninvasive cardiac output measurement: an update.

Authors:  Lisa Sangkum; Geoffrey L Liu; Ling Yu; Hong Yan; Alan D Kaye; Henry Liu
Journal:  J Anesth       Date:  2016-03-09       Impact factor: 2.078

Review 6.  Pulse waveform hemodynamic monitoring devices: recent advances and the place in goal-directed therapy in cardiac surgical patients.

Authors:  Adham Hendy; Şerban Bubenek
Journal:  Rom J Anaesth Intensive Care       Date:  2016-04

7.  A comparison of lithium dilution cardiac output measurements made using central and antecubital venous injection of lithium chloride.

Authors:  M M Jonas; F E Kelly; R A Linton; D M Band; T K O'Brien; N W Linton
Journal:  J Clin Monit Comput       Date:  1999-12       Impact factor: 2.502

8.  Ability of pulse power, esophageal Doppler, and arterial pulse pressure to estimate rapid changes in stroke volume in humans.

Authors:  José Marquez; Kenneth McCurry; Donald A Severyn; Michael R Pinsky
Journal:  Crit Care Med       Date:  2008-11       Impact factor: 7.598

9.  Effects of vasodilation on cardiac output measured by PulseCO.

Authors:  Koichi Yamashita; Tomoki Nishiyama; Takeshi Yokoyama; Hidehiro Abe; Masanobu Manabe
Journal:  J Clin Monit Comput       Date:  2007-09-25       Impact factor: 2.502

10.  Continuous and intermittent cardiac output measurement in hyperdynamic conditions: pulmonary artery catheter vs. lithium dilution technique.

Authors:  Maria Gabriella Costa; Giorgio Della Rocca; Paolo Chiarandini; Silvia Mattelig; Livia Pompei; Mauricio Sainz Barriga; Toby Reynolds; Maurizio Cecconi; Paolo Pietropaoli
Journal:  Intensive Care Med       Date:  2007-10-06       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.