A comparison of lithium dilution cardiac output measurements made using central and antecubital venous injection of lithium chloride.

M M Jonas1, F E Kelly, R A Linton, D M Band, T K O'Brien, N W Linton.   

Abstract

OBJECTIVE: We have previously described an indicator dilution technique of measuring cardiac output in which lithium chloride is injected as a bolus via a central venous catheter and cardiac output derived from the arterial lithium dilution curve recorded from a lithium-selective electrode, which we have developed for this purpose. It would be an advantage if the lithium could be injected via the basilic vein (in the antecubital fossa) in those patients who do not need central venous catheterisation for other reasons. We have therefore compared cardiac output measurements made using these two routes of lithium chloride administration.
METHODS: Lithium dilution cardiac output was measured 10 times in each of 10 patients, injecting the lithium chloride alternately via the basilic or central venous catheter.
RESULTS: The mean difference was 0.8 +/- 5.2% (SD) (range -8.5 to +7.0%) over a range of cardiac output of 4.5-13 l/min.
CONCLUSIONS: Injection of lithium chloride via the basilic vein in the antecubital fossa allows accurate lithium dilution cardiac output measurements to be made in patients who do not have central venous catheters in place.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 12578051     DOI: 10.1023/a:1009914714769

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  9 in total

1.  Limitations of indicator-dilution methods in experimental shock.

Authors:  A Oriol; P Sekelj; M McGregor
Journal:  J Appl Physiol       Date:  1967-10       Impact factor: 3.531

2.  Errors in measurement of cardiac output with dye-dilution curves in shock.

Authors:  L D Homer; G S Moss; C M Herman
Journal:  J Appl Physiol       Date:  1969-07       Impact factor: 3.531

3.  A new method of analysing indicator dilution curves.

Authors:  R A Linton; N W Linton; D M Band
Journal:  Cardiovasc Res       Date:  1995-12       Impact factor: 10.787

4.  Comparison of the accuracy of the lithium dilution technique with the thermodilution technique for measurement of cardiac output.

Authors:  T Kurita; K Morita; S Kato; M Kikura; M Horie; K Ikeda
Journal:  Br J Anaesth       Date:  1997-12       Impact factor: 9.166

5.  The shape of indicator dilution curves used for cardiac output measurement in man.

Authors:  D M Band; R A Linton; T K O'Brien; M M Jonas; N W Linton
Journal:  J Physiol       Date:  1997-01-01       Impact factor: 5.182

6.  Lithium dilution cardiac output measurement: a comparison with thermodilution.

Authors:  R Linton; D Band; T O'Brien; M Jonas; R Leach
Journal:  Crit Care Med       Date:  1997-11       Impact factor: 7.598

7.  The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators.

Authors:  A F Connors; T Speroff; N V Dawson; C Thomas; F E Harrell; D Wagner; N Desbiens; L Goldman; A W Wu; R M Califf; W J Fulkerson; H Vidaillet; S Broste; P Bellamy; J Lynn; W A Knaus
Journal:  JAMA       Date:  1996-09-18       Impact factor: 56.272

8.  A new method of measuring cardiac output in man using lithium dilution.

Authors:  R A Linton; D M Band; K M Haire
Journal:  Br J Anaesth       Date:  1993-08       Impact factor: 9.166

Review 9.  Errors in the measurement of cardiac output by thermodilution.

Authors:  T Nishikawa; S Dohi
Journal:  Can J Anaesth       Date:  1993-02       Impact factor: 5.063

  9 in total
  4 in total

1.  Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique.

Authors:  M Cecconi; D Dawson; R M Grounds; A Rhodes
Journal:  Intensive Care Med       Date:  2008-09-18       Impact factor: 17.440

Review 2.  Pulse waveform hemodynamic monitoring devices: recent advances and the place in goal-directed therapy in cardiac surgical patients.

Authors:  Adham Hendy; Şerban Bubenek
Journal:  Rom J Anaesth Intensive Care       Date:  2016-04

3.  Useful techniques for ultrasonography-guided free-hand puncture of a deep peripheral vein and catheter placement to obtain CT angiography.

Authors:  Toshiyuki Irie
Journal:  Jpn J Radiol       Date:  2013-09-15       Impact factor: 2.374

Review 4.  Equipment review: an appraisal of the LiDCO plus method of measuring cardiac output.

Authors:  Rupert M Pearse; Kashif Ikram; John Barry
Journal:  Crit Care       Date:  2004-05-05       Impact factor: 9.097

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.