IOLMaster biometry: refractive results of 100 consecutive cases.

H Eleftheriadis1.   

Abstract

AIMS: To study the refractive outcome of cataract surgery employing IOLMaster biometry data and to compare it with that of applanation ultrasonography in a prospective study of 100 eyes that underwent phacoemulsification with intraocular lens implantation.
METHODS: The Holladay formula using IOLMaster data was employed for the prediction of implanted intraocular lenses (IOLs). One month after cataract surgery the refractive outcome was determined. Preoperative applanation ultrasonography data were used retrospectively to calculate the IOL prediction error. The two different biometry methods are compared.
RESULTS: 100 patients, 75.42 (SD 7.58) years of age, underwent phacoemulsification with IOL implantation. The optical axial length obtained by the IOLMaster was significantly longer (p<0.001, Student's t test) than the axial length by applanation ultrasound, 23.36 (SD 0.85) mm v 22.89 (0.83) mm. The mean postoperative spherical equivalent was 0.00 (0.40) D and the mean prediction error -0.15 (0.38) D. The mean absolute prediction error was 0.29 (0.27) D. 96% of the eyes were within 1 D from the intended refraction and 93% achieved unaided visual acuity of 6/9 or better. The Holladay formula performed better than the SRK/T, SRK II, and Hoffer Q formulas. Applanation ultrasonography after optimisation of the surgeon factor yielded a greater absolute prediction error than the optimised IOLMaster biometry, 0.41 (0.38) D v 0.25 (0.27) D, with 93% of the eyes within 1 D from the predicted refraction.
CONCLUSION: IOLMaster optical biometry improves the refractive results of selected cataract surgery patients and is more accurate than applanation ultrasound biometry.

Entities:  

Mesh:

Year:  2003        PMID: 12881334      PMCID: PMC1771781          DOI: 10.1136/bjo.87.8.960

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  32 in total

1.  Accuracy and reproducibility of biometry using partial coherence interferometry.

Authors:  Rafe Connors; Peter Boseman; Randall J Olson
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

2.  Sources of error in intraocular lens power calculation.

Authors:  T Olsen
Journal:  J Cataract Refract Surg       Date:  1992-03       Impact factor: 3.351

3.  Eye-length measurement by interferometry with partially coherent light.

Authors:  A F Fercher; K Mengedoht; W Werner
Journal:  Opt Lett       Date:  1988-03-01       Impact factor: 3.776

4.  Biometry of cataractous eyes using partial coherence interferometry: clinical feasibility study of a commercial prototype I.

Authors:  Barbara Kiss; Oliver Findl; Rupert Menapace; Matthias Wirtitsch; Wolfgang Drexler; Christoph K Hitzenberger; Adolf F Fercher
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

5.  Immersion versus contact technique in the measurement of axial length by ultrasound.

Authors:  T Olsen; P J Nielsen
Journal:  Acta Ophthalmol (Copenh)       Date:  1989-02

6.  Stop and chop phacoemulsification.

Authors:  P S Koch; L E Katzen
Journal:  J Cataract Refract Surg       Date:  1994-09       Impact factor: 3.351

7.  Astigmatic stabilization of 3.0 mm temporal clear corneal cataract incisions.

Authors:  S Masket; D G Tennen
Journal:  J Cataract Refract Surg       Date:  1996-12       Impact factor: 3.351

8.  Partial coherence interferometry: a novel approach to biometry in cataract surgery.

Authors:  W Drexler; O Findl; R Menapace; G Rainer; C Vass; C K Hitzenberger; A F Fercher
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

9.  Comparison of A-scan device accuracy.

Authors:  U Giers; C Epple
Journal:  J Cataract Refract Surg       Date:  1990-03       Impact factor: 3.351

10.  Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations.

Authors:  M S Rajan; I Keilhorn; J A Bell
Journal:  Eye (Lond)       Date:  2002-09       Impact factor: 3.775

View more
  29 in total

1.  Relationship between the retinal thickness of the macula and the difference in axial length.

Authors:  Tetsuo Ueda; Yoshiaki Nawa; Yoshiaki Hara
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-08-13       Impact factor: 3.117

2.  Corrective Techniques and Future Directions for Treatment of Residual Refractive Error Following Cataract Surgery.

Authors:  Majid Moshirfar; Michael V McCaughey; Luis Santiago-Caban
Journal:  Expert Rev Ophthalmol       Date:  2014-12

3.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

4.  Prediction of refractive outcome after cataract surgery using partial coherence interferometry: comparison of SRK/T and Haigis formulae.

Authors:  Rohit Sharma; P Maharajan; S Kotta; S Maharajan
Journal:  Int Ophthalmol       Date:  2014-03-30       Impact factor: 2.031

5.  Refractive results of phacoemulsification in vitrectomized patients.

Authors:  Jia-Kang Wang; Shu-Wen Chang
Journal:  Int Ophthalmol       Date:  2016-08-25       Impact factor: 2.031

6.  Accuracy of eight intraocular lens power calculation formulas for segmented multifocal intraocular lens.

Authors:  Jing Zhao; Liang-Ping Liu; Huan-Huan Cheng; Jian-Bing Li; Xiao-Tong Han; Yu Liu; Ming-Xing Wu
Journal:  Int J Ophthalmol       Date:  2020-09-18       Impact factor: 1.779

7.  Axial biometry of the entire eye using ultra-long scan depth optical coherence tomography.

Authors:  Jianguang Zhong; Yilei Shao; Aizhu Tao; Hong Jiang; Che Liu; Huicheng Zhang; Jianhua Wang
Journal:  Am J Ophthalmol       Date:  2013-10-07       Impact factor: 5.258

8.  Long-term Outcomes of Undercorrection Versus Full Correction After Unilateral Intraocular Lens Implantation in Children.

Authors:  Scott R Lambert; Steven M Archer; M Edward Wilson; Rupal H Trivedi; Monte A del Monte; Michael Lynn
Journal:  Am J Ophthalmol       Date:  2011-12-28       Impact factor: 5.258

9.  Agreement of IOL power and axial length obtained by IOLMaster 500 vs IOLMaster 500 with Sonolink connection.

Authors:  Sabong Srivannaboon; Chareenun Chirapapaisan; Patchara Nantasri; Mathinee Chongchareon; Pratuangsri Chonpimai
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-12       Impact factor: 3.117

10.  Comparison of preoperative and postoperative ocular biometry in eyes with phakic intraocular lens implantations.

Authors:  Joo Youn Shin; Jae Bum Lee; Kyoung Yul Seo; Eung Kweon Kim; Tae-im Kim
Journal:  Yonsei Med J       Date:  2013-09       Impact factor: 2.759

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.