Andrew Milewski1, Guang Li1. 1. Department of Medical Physics, 5803Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Abstract
To assess the stability of patient-specific phase shifts between external- and internal-respiratory motion waveforms, the reliability of enhanced external-internal correlation with phase-shift correction, and the feasibility of guiding respiratory-gated radiotherapy (RGRT) over 30 min. In this clinical feasibility investigation, external bellows and internal-navigator waveforms were simultaneously and prospectively acquired along with two four-dimensional magnetic resonance imaging (4DMRI) scans (6-15 m each) with 15-20 m intervals in 10 volunteers. A bellows was placed 5 cm inferior to the xiphoid to monitor abdominal motion, and an MR navigator was used to track the diaphragmatic motion. The mean phase-domain (MPD) method was applied, which combines three individual phase-calculating methods: phase-space oval fitting, principal component analysis, and analytic signal analysis, weighted by the reciprocal of their residual errors (RE) excluding outliers (RE >2σ). The time-domain cross-correlation (TCC) analysis was applied for comparison. Dynamic phase-shift correction was performed based on the phase shift detected on the fly within two 10 s moving datasets. Simulating bellows-triggered gating, the median and 95% confidence interval for the navigator's position at beam-on/beam-off and %harm (percentage of beam-on time outside the safety margin) were calculated. Averaged across all subjects, the mean phase shifts are found indistinguishable (p > .05) between scan 1 (55˚ ± 9˚) and scan 2 (59˚ ± 11˚). Using the MPD method the averaged correlation increases from 0.56 ± 0.22 to 0.85 ± 0.11 for scan 1 and from 0.47 ± 0.30 to 0.84 ± 0.08 for scan 2. The TCC correction results in similar results. After phase-shift correction, the number of cases that were suitable for amplitude gating (with <10%harm) increased from 2 to 17 out of 20 cases. A patient-specific, stable phase-shift between the external and internal motions was observed and corrected using the MPD and TCC methods, producing long-lasting enhanced motion correlation over 30m. Phase-shift correction offers a feasible strategy for improving the accuracy of tumor-motion prediction during RGRT.
To assess the stability of patient-specific phase shifts between external- and internal-respiratory motion waveforms, the reliability of enhanced external-internal correlation with phase-shift correction, and the feasibility of guiding respiratory-gated radiotherapy (RGRT) over 30 min. In this clinical feasibility investigation, external bellows and internal-navigator waveforms were simultaneously and prospectively acquired along with two four-dimensional magnetic resonance imaging (4DMRI) scans (6-15 m each) with 15-20 m intervals in 10 volunteers. A bellows was placed 5 cm inferior to the xiphoid to monitor abdominal motion, and an MR navigator was used to track the diaphragmatic motion. The mean phase-domain (MPD) method was applied, which combines three individual phase-calculating methods: phase-space oval fitting, principal component analysis, and analytic signal analysis, weighted by the reciprocal of their residual errors (RE) excluding outliers (RE >2σ). The time-domain cross-correlation (TCC) analysis was applied for comparison. Dynamic phase-shift correction was performed based on the phase shift detected on the fly within two 10 s moving datasets. Simulating bellows-triggered gating, the median and 95% confidence interval for the navigator's position at beam-on/beam-off and %harm (percentage of beam-on time outside the safety margin) were calculated. Averaged across all subjects, the mean phase shifts are found indistinguishable (p > .05) between scan 1 (55˚ ± 9˚) and scan 2 (59˚ ± 11˚). Using the MPD method the averaged correlation increases from 0.56 ± 0.22 to 0.85 ± 0.11 for scan 1 and from 0.47 ± 0.30 to 0.84 ± 0.08 for scan 2. The TCC correction results in similar results. After phase-shift correction, the number of cases that were suitable for amplitude gating (with <10%harm) increased from 2 to 17 out of 20 cases. A patient-specific, stable phase-shift between the external and internal motions was observed and corrected using the MPD and TCC methods, producing long-lasting enhanced motion correlation over 30m. Phase-shift correction offers a feasible strategy for improving the accuracy of tumor-motion prediction during RGRT.
Respiratory-induced tumor motion is one of the largest sources of uncertainties in
radiation therapy of a mobile tumor, such as lung cancer, and a large motion margin
must be applied to overcome the motion uncertainty. The motion safety margin
enlarges the treatment target volume multifold from the clinical tumor volume (CTV)
to the internal tumor volume (ITV), defined as the motion envelope of the CTV (ICRU, 2010).
Because the ITV encompasses a large volume of healthy tissue, a lowered
radiation dose is often prescribed and delivered to the tumor to avoid exceeding the
tolerable toxicity of nearby organs at risk (OARs), including the central lung
region with serial functionality, the so-called “no-fly-zone.”
To reduce the motion margin, various methods have been proposed, developed,
and applied in the clinic, including breath-holding, respiratory gating, and tumor
tracking.[3-7] Several approaches, including
the CyberKnife LINAC (linear accelerator) system, rely on fluoroscopy or frequent
radiographic imaging coupled with a real-time tumor-motion prediction model and an
external surrogate.[8-10] The Calypso
system using electromagnetic transponders,[7,11,12] and, more recently, the
MR-integrated LINAC (MRL) system using 2D cine MR[13-15] have been applied clinically,
but implanting the Calypso transponder(s) precludes the possibility of follow-up MR
imaging at the disease site and the MRL may yet require substantial investment and
resources. Therefore, applying real-time optical imaging (or external surrogates)
together with an external-to-internal motion prediction model and periodic
radiographic verification remains a viable clinical option in the majority of
conventional iso-centric LINACs for respiratory-gated radiotherapy (RGRT).Although various external–internal motion prediction models have been reported, such
as a 5D model,
deformation models,[17,18] a biomechanical model,
a semi-physical sprint-dashpot model,
a physical motion perturbation model,
and machine-learning models,[22,23] only correlation-based models
have been applied clinically for real-time, image-guided radiotherapy (IGRT). To
apply a correlation-based prediction model, a patient should demonstrate a high
external–internal motion correlation at the simulation using fiducial markers that
are implanted close to the tumor.
Marker-less approaches that simplify the clinical procedure and avoid any
implant complications have also been studied,[8,25,26] and some of these strategies
have been implemented in radiotherapy clinics.
Even if the motion correlation is high, the correlation model must be
periodically verified and updated with radiographic imaging to accommodate changes
in the patient's breathing behavior, which often necessitates rebuilding the
predictive model and prolonging treatment.The quality of the external–internal motion correlation is significantly affected by
the phase shift between the external- and internal-motion waveforms.[29-31] Large patient-specific phase
shifts may deteriorate the correlation between the two waveforms. We previously
developed a phase-domain approach to detect the phase shifts between the
external-bellows and internal-navigator waveforms acquired simultaneously during a
4DMRI scan, and we evaluated whether the correlation can be enhanced by correcting
the phase shift.
However, the duration of the 4DMRI scans was shorter than a radiotherapy
treatment, necessitating an assessment of the stability and reliability of phase
shifts and enhanced correlations over timeframes that are relevant for radiotherapy
treatments. That different external surrogates may disagree in their predictions of
a tumor's position raises additional concerns about the accuracy of tumor-position
prediction by the correlative strategies employed in RGRT.
Due to the uncertainty in correlative prediction, x-ray imaging is often
required to verify the tumor's position when the radiation beam is turned on or
off.[31,34]In this study, we employed two sets of 4DMRI scans (6–15 m) separated by a 15–20 m
gap in 10 healthy volunteers under an IRB-approved protocol. The external motion was
detected using a bellows system and the internal motion was measured concurrently
from the diaphragm's motion detected by an MR navigator during 4DMRI scans. The
dynamically detected phase-shift and enhanced correlation were evaluated within each
scan and compared between the two scans. We also examined whether correcting the
phase shift improves the quality of amplitude-triggered respiratory gating. We
evaluated a theoretical tumor's position at beam-on/beam-off times and we simulated
the percentage of time (%harm) that radiation would be delivered outside of a safety
margin.
Methods
In this clinical feasibility investigation study, volunteer subjects’ data were
acquired under clinical conditions and employed to assess the stability of
patient-specific phase shift, reliability of the enhancement of external–internal
motion correlation, and feasibility of applying the phase-shift correction technique
to improve RGRT.
Concurrent External and Internal Waveforms Acquired During Multiple 4DMRI
Scans
Motion waveforms from the external bellows (at 496 Hz) and the internal navigator
(at 20 Hz) were acquired concurrently and prospectively during two
respiratory-correlated 4DMRI scans using a 3T MR scanner (Ingenia, Philips
Healthcare, Amsterdam, Netherland) in 10 healthy volunteers under an
IRB-approved protocol. The bellows was placed 5 cm inferior to the sternum's
xiphoid process and the navigator (3 × 3 × 6 cm3) was configured at
the dome of the right diaphragm. The first 4DMRI scan was in the coronal
direction and the second scan was in the sagittal direction. Depending on
breathing irregularities, a 4DMRI scan lasted 6–15 min, and the second scan
started 15–20 min after the first scan ended. The navigator signal was
continuous except at times when 2D coronal-or sagittal-slice images were
acquired for the reconstruction of 4DMRI.
The bellows signal was continuous, not interrupted by MR acquisition.
Because both waveforms were acquired concurrently on the fly prospectively
during the 4DMRI scans, they represented the motion of the respiratory organ in
real time.The initial timestamps (in milliseconds) in the log files of the scanner and
bellows systems were used to synchronize the navigator and bellows waveforms. To
assess the phase shift, an automated algorithm identified navigator-waveform
segments containing holes no longer than 0.3 s, and the small gaps (up to 0.3 s)
were filled in by linear interpolation. Corresponding time segments from the
bellows signal were extracted and downsampled to match the 20 Hz frequency of
the navigator.To assess the phase shift, all available pairs of navigator and bellows time
segments throughout a scan were divided into 10-s sampling windows. When the
length of a time segment permitted multiple sampling windows, each 10-s sampling
window overlapped with 5 s of the subsequent window.In phase space, the occurrence of ellipsoidal trajectories (navigator vs.
bellows) confirmed the existence of a phase shift between the two
waveforms.[29,32] Estimates of the phase shift were obtained via
phase-space oval fitting (POF), principal component analysis (PCA), and analytic
signal analysis (ASA) methods. These results were combined as the mean
phase-domain (MPD) method (described below). After correcting the calculated
phase shift, the resulting correlation between the two waveforms was compared
with the original correlation and with the maximum value of the time-domain
cross-correlation (TCC) (Figure 1).
Figure 1.
Methods for estimating the phase shift. (A) Example waveforms acquired by
the external bellows (orange) and internal navigator (purple) for a
healthy volunteer. The two waveforms reach their respective maximum
values at different times, indicating the presence of a phase shift. (B)
Ovoid trajectories (red) seen in a phase-space plot reiterate the
existence of a phase shift between the external-bellows and
internal-navigator waveforms. As was previously described, the phase
shift can be efficiently estimated by a weighted average of three
phase-domain techniques: best-fit oval, principal components analysis,
and analytic signal analysis.
The trajectories in phase space become nearly linear (blue) after
correction of the phase shift, reflecting an enhanced correlation
between the external-bellows and internal-navigator waveforms. (C) The
phase shift can also be estimated in the time domain by calculating the
time shift that maximizes the cross-correlation between the two
waveforms. (D) Correction of the phase shift—calculated
by either method—translates the bellows waveform in time (gray arrow),
thereby achieving a greater similarity and, consequently, a stronger
correlation between the two waveforms.
Methods for estimating the phase shift. (A) Example waveforms acquired by
the external bellows (orange) and internal navigator (purple) for a
healthy volunteer. The two waveforms reach their respective maximum
values at different times, indicating the presence of a phase shift. (B)
Ovoid trajectories (red) seen in a phase-space plot reiterate the
existence of a phase shift between the external-bellows and
internal-navigator waveforms. As was previously described, the phase
shift can be efficiently estimated by a weighted average of three
phase-domain techniques: best-fit oval, principal components analysis,
and analytic signal analysis.
The trajectories in phase space become nearly linear (blue) after
correction of the phase shift, reflecting an enhanced correlation
between the external-bellows and internal-navigator waveforms. (C) The
phase shift can also be estimated in the time domain by calculating the
time shift that maximizes the cross-correlation between the two
waveforms. (D) Correction of the phase shift—calculated
by either method—translates the bellows waveform in time (gray arrow),
thereby achieving a greater similarity and, consequently, a stronger
correlation between the two waveforms.
MPD Method
A combination of three phase-domain methods (POF, PCA, and ASA) was developed and
published previously as the MPD method
and re-implemented in MATLAB (Version 2020a) to allow automatic data
sampling for dynamically estimating the phase shift. Automating the
data-sampling strategy significantly reduced data-processing times; the updated
algorithm can readily be applied for real-time, online data processing in future
clinical applications. When a phase shift was calculated using the three
methods, their residual errors (RE) were also estimated and
used to calculate the weighting factor (reciprocal of RE) in
the MPD method, as shown in Equation 1.
where
is the weighting factor for the corresponding phase-shift
calculation method (i = POF, PCA, and ASA). At each successive
time point, the phase shift was estimated by halving the weighted average of the
phase shifts calculated by the individual methods at the current time and adding
the result to half of the previous moving-average value.A moving standard deviation was calculated as the average standard deviation of
the individual phase shifts found at all the previous time points. When the
phase shift calculated by an individual method was an outlier (
> 2σ) compared to the other methods or to the historical
values of the phase shift, the weighting factor
for this term was set to 0. Namely, the contribution from this
method was removed, and the phase shift for that sampling window was then
calculated with the remaining methods. This weighting strategy minimizes the
idiosyncratic responses of the individual methods, yielding a more robust
estimation of the phase shift. When all three
are outliers, the algorithm assumes that an abrupt change has
occurred in the phase shift and restarts the running average. If such a rare
event occurred in a future application, the radiation beam should be temporarily
held. The detailed mathematical equations and evaluations were reported earlier.
Phase-Shift Correction for Improved External–Internal Correlation
The phase shift that was calculated for each sampling window was then corrected
to dynamically boost the correlation between the external and internal
waveforms. The enhanced correlation was compared with the original correlation.
The MPD results were also validated against the TCC method, which maximizes the
cross-correlation between the two signals. The time-domain and phase-domain
methods are illustrated in Figure 1.To correct the phase shift estimated by the TCC method, the navigator signal was
shifted in time by an amount equal to the calculated phase shift divided by the
waveform's frequency. The frequency of the waveform was estimated in four ways:
the frequency at which the power spectrum of the bellows or navigator signals
achieved their maximum value, and the average slope of the bellows’ or
navigator's instantaneous phase. At each time point, the frequency estimate was
chosen to maximize the correlation between the waveforms following phase-shift
correction. By comparing the correlation coefficient for the original waveforms
to that of the phase-shift corrected waveforms, we assessed whether correcting
the phase shift enhances the correlation between the external and internal
signals.
Applying Phase-Shift Correction in Respiratory-Gated Radiotherapy
To evaluate the usefulness of enhanced correlation after phase-shift correction,
we examined a plausible clinical scenario wherein a 50% reduction in a tumor's
motion margin is sought through RGRT. A theoretical threshold for triggering the
radiation beam was set at the lower 30% of the bellows’ amplitude (near the full
exhalation) during the initial 10 s of each 4DMRI scan. In this hypothetical
scenario, the radiation beam would turn on anytime the position of the bellows
falls below the triggering threshold and the beam would turn off when the
bellows rises above the threshold. This strategy aims to deliver radiation only
when the navigator's position is within the lower 50% of its amplitude, and
accordingly, a 50% motion margin should be applied in the RGRT planning. A
somewhat conservative triggering threshold was selected to account for
differences in the shapes of the external and internal waveforms. Because the
phase shift does not account for baseline drift (although the standard 5.0 mm
setup margin may account for some drift), periodic radiographic checking may be
necessary to correct the potential baseline drift.The gating accuracy was assessed by calculating the percentage of beam-on time
(%harm) that radiation would have theoretically been delivered outside the
motion margin before and after correcting the phase shifts estimated by the MPD
and TCC methods. The median position of the navigator and 95% confidence
interval (CI) at the instant the radiation beam was triggered to turn on were
also calculated before and after phase-shift correction.
Results
Dynamic Detection of Subject-Specific Phase Shifts in 2 Separate
Scans
Table 1 tabulates
the durations of the 2 4DMRI scans and the time sampling for phase-shift
analysis in 10 volunteer subjects. Every scan lasted 5–16 min and a 15–20 m gap
separated the pair of scans that each volunteer received. Because fewer image
slices are needed to span a volunteer's anterior-to-posterior dimension than
that needed for the lateral direction, the coronal scans (scan 1) were shorter
on average (8.7 ± 2.7 m) than the sagittal scans (scan 2) (11.8 ± 2.2 m). Across
all of the scans, a total of 1213 time windows encompassing 114.9 m of data were
analyzed.
Table 1.
Scan Time, Inter-Scan Gap, and Sampling Time for the Concurrently
Acquired Motion Waveforms From 2 4dmri Scans Used in the Dynamic
Phase-Shift Analyses of 10 Volunteers.
Volunteer
Gender
Scan Length (m)
Scan
No. of Time
Windowsa
Time Analyzed (m)
Scan 1
Scan 2
Gap (m)
Scan 1
Scan 2
Scan 1
Scan 2
1
♀
6.5
12.1
20.4
6
50
0.8
4.6
2
♂
6.9
9.7
15.8
8
102
0.9
8.7
3
♂
9.4
9.6
16.9
16
44
2.0
4.3
4
♂
14.7
14.1
18.7
28
67
4.2
6.3
5
♀
8.8
16.1
29.7
68
153
6.3
14.2
6
♀
5.3
10.2
27.7
53
70
4.6
6.8
7
♂
8.2
9.6
16.5
63
83
5.5
7.8
8
♀
6.5
11.5
16.6
66
121
5.8
11.0
9
♂
9.7
11.6
18.6
12
103
1.6
9.3
10
♀
11.1
13.9
19.4
7
93
0.8
9.4
Mean
8.7
11.8
20.0
32.7
88.6
3.3
8.2
St Dev
2.7
2.2
4.8
26.7
33.2
2.2
3.0
There are 1213 time windows in total and each contains 10 s.
Scan Time, Inter-Scan Gap, and Sampling Time for the Concurrently
Acquired Motion Waveforms From 2 4dmri Scans Used in the Dynamic
Phase-Shift Analyses of 10 Volunteers.There are 1213 time windows in total and each contains 10 s.Table 2 shows the
estimated phase shifts for both scans using the MPD and TCC methods. The average
phase shifts determined by the MPD method are approximately 1.00 radians (57°
out of 360° or 16% of a breathing cycle) and are similar in both scans with a
mean ratio of 1.07 (p = .67). The standard deviations are
relatively small (<18%), except for two subjects in scan 1 (#4 and #10) and
four volunteers in scan 2 (#5, #6, #7, and #9) who experienced higher breathing
irregularities (23%–68%). The phase-shift ratio between scan 1 and scan 2 is
0.95 ± 0.19, excluding subjects 5 and 7 as 2 outliers. The TCC method calculates
a smaller mean phase shift (0.85 radians), but usually yields a larger variation
(ratio = 1.01, p = .51).
Table 2.
Phase shifts (in radians) determined by Both the Mean Phase-Domain (MPD)
Method and the Time-domain Cross-correlation (TCC) method in 4DMRI scan
1 and scan 2.
Subject
Phase Shifts (Radian) by MPD
method
Phase Shifts (Radian) by TCC
method
Scan 1
Scan 2
Ratio of 2 means
Scan 1
Scan 2
Ratio of 2 means
Mean
SD
Mean
SD
Mean
SD
Mean
SD
1
0.67
0.01
0.74
0.05
0.91
0.58
0.04
0.63
0.10
0.92
2
1.00
0.02
1.22
0.08
0.82
0.84
0.04
0.98
0.20
0.86
3
0.59
0.04
0.55
0.03
1.07
0.55
0.06
0.51
0.05
1.08
4
1.17
0.32
1.27
0.18
0.92
1.16
0.35
1.23
0.42
0.94
5
0.65
0.10
0.28
0.48
2.32
0.43
0.09
0.21
0.31
2.05
6
0.99
0.16
1.02
0.23
0.97
0.79
0.22
0.84
0.47
0.94
7
0.97
0.07
1.60
0.24
0.61
0.86
0.15
1.57
0.27
0.55
8
1.57
0.03
1.55
0.08
1.01
1.33
0.26
1.34
0.29
0.99
9
0.96
0.14
1.24
0.35
0.77
0.85
0.22
1.14
0.38
0.75
10
1.07
0.73
0.84
0.13
1.27
0.57
0.37
0.59
0.20
0.97
Mean
0.96
0.16
1.03
0.19
1.07
0.80
0.18
0.90
0.27
1.01
SD
0.29
0.22
0.41
0.14
0.45
0.28
0.12
0.40
0.13
0.38
p-value
0.67
0.51
Phase shifts (in radians) determined by Both the Mean Phase-Domain (MPD)
Method and the Time-domain Cross-correlation (TCC) method in 4DMRI scan
1 and scan 2.Figure 2 illustrates the
phase-shift results based on both the MPD and TCC methods. For more than half of
the volunteers, the phase shift was relatively stable throughout both scans
using the MPD (Figure
2A) and TCC (Figure
2C) methods. The box-and-whisker diagrams reveal that the phase-shift
distributions are usually broader for scan 2 (Figures 2B and 2D), an effect that may arise from
having subjects participate in several breath-holding scans before the second
4DMRI scans. Notwithstanding differences in the distributions’ spread, the
median phase shifts in each pair of scans are similar, except for subjects 7 and
9. Some scattered outliers in the phase-shift distributions were caused by
extreme breathing irregularities.
Figure 2.
Phase-shift distributions across two scans. (A) Phase shifts were
calculated by the mean phase-domain (MPD) method for the first (top) and
second (bottom) scans of volunteer 3. Each listed time indicates the
beginning of a time segment, and each point on the curve marks the phase
shift for a 10-s window. The dashed, red line indicates the median phase
shift for each scan. (B) Box-and-whisker diagrams portray the
distributions of the phase shifts estimated by the MPD method during the
first (F) and second (S) 4D-MRI scans for each of the 10 volunteers. The
red, horizontal lines indicate the median of each distribution, the
boxes encompass the middle two quartiles, the whiskers span each
distribution's range, and the xs mark outliers that are three or more
interquartile lengths below the lower quartile or above the upper
quartile. The distributions appear similar for half of the volunteers
(1, 3, 6, 8, and 10). (C, D) The phase-shift results for the time-domain
cross-correlation (TCC) method are portrayed as for panels A and B.
Phase-shift distributions across two scans. (A) Phase shifts were
calculated by the mean phase-domain (MPD) method for the first (top) and
second (bottom) scans of volunteer 3. Each listed time indicates the
beginning of a time segment, and each point on the curve marks the phase
shift for a 10-s window. The dashed, red line indicates the median phase
shift for each scan. (B) Box-and-whisker diagrams portray the
distributions of the phase shifts estimated by the MPD method during the
first (F) and second (S) 4D-MRI scans for each of the 10 volunteers. The
red, horizontal lines indicate the median of each distribution, the
boxes encompass the middle two quartiles, the whiskers span each
distribution's range, and the xs mark outliers that are three or more
interquartile lengths below the lower quartile or above the upper
quartile. The distributions appear similar for half of the volunteers
(1, 3, 6, 8, and 10). (C, D) The phase-shift results for the time-domain
cross-correlation (TCC) method are portrayed as for panels A and B.
Enhanced Motion Correlation with Dynamic Phase-Shift Correction
Plotting the phase shifts and the initial correlations on the same time axis
reveals that the two variables nearly mirror each other, suggesting an inverse
relationship: low initial correlations are associated with large phase shifts
(Figures 3A and
3B). For most of
the volunteers, correcting the phase shift yields a correlation curve close to
1.0 in almost every time window. Examining the relationship between the phase
shift and the initial correlation for the 1213 sampling windows analyzed for all
10 volunteers provides further evidence of the inverse relationship (Figures 3C and 3D). Indeed, the
Pearson's correlation coefficient between the MPD phase shifts and initial
correlations is −0.93, and a zero-parameter fit to the line, Initial
Correlation = 1 − (2/π)*abs(ϕMPD), yields an average residual error
of 0.16. Better agreement is found for the curve, Initial
Correlation = cos(ϕMPD), which yields an average residual error
of 0.07 (Figure 3C,
top). Correcting the MPD phase shift enhances the correlation between the
navigator and bellows waveforms in 92.25% ( = 1109/1213) of the analyzed
sampling windows. On average, the correlation rose by 0.37 after correcting the
phase shift (Figure 3C,
bottom).
Figure 3.
An inverse relationship exists between the phase shift and initial
correlation. (A) Phase shifts estimated by the MPD method (dashed,
purple curves), the initial correlation between the bellows and
navigator waveforms (solid, black curves), and the correlation between
the waveforms after phase-shift correction (solid, blue curves)
(mean = 0.87 ± 0.07) for volunteer 7. (B) Phase shifts estimated by the
TCC method (dashed, gold), the initial correlation between the bellows
and navigator waveforms (solid, black), and the correlation between the
waveforms after phase-shift correction (solid, green). (C) Each black
point in the scatter plot (top panel) indicates the calculated MPD phase
shift and initial correlation between the bellows and navigator
waveforms for a single time window. This relation is reasonably
described by Initial Correlation = 1 – 2*abs(ϕ)/π (red, dashed line) and
even better described by Initial Correlation = cos(ϕ) (orange, dashed
curve). Each black point is paired with a blue point that indicates the
correlation between the waveforms following phase-shift correction;
vertical gray lines indicate improved correlations and vertical red
lines indicate correlations that were worse. Sorting the change in
correlation from least to greatest for the 1213 time-segment windows
from all 10 volunteers (bottom panel) illustrates that enhancement of
the correlation was achieved for 1119 windows (92.25%). (D) The TCC
results are shown in the same fashion, but no correction reductions (red
lines) are observed, resulting in 100% improvement.
An inverse relationship exists between the phase shift and initial
correlation. (A) Phase shifts estimated by the MPD method (dashed,
purple curves), the initial correlation between the bellows and
navigator waveforms (solid, black curves), and the correlation between
the waveforms after phase-shift correction (solid, blue curves)
(mean = 0.87 ± 0.07) for volunteer 7. (B) Phase shifts estimated by the
TCC method (dashed, gold), the initial correlation between the bellows
and navigator waveforms (solid, black), and the correlation between the
waveforms after phase-shift correction (solid, green). (C) Each black
point in the scatter plot (top panel) indicates the calculated MPD phase
shift and initial correlation between the bellows and navigator
waveforms for a single time window. This relation is reasonably
described by Initial Correlation = 1 – 2*abs(ϕ)/π (red, dashed line) and
even better described by Initial Correlation = cos(ϕ) (orange, dashed
curve). Each black point is paired with a blue point that indicates the
correlation between the waveforms following phase-shift correction;
vertical gray lines indicate improved correlations and vertical red
lines indicate correlations that were worse. Sorting the change in
correlation from least to greatest for the 1213 time-segment windows
from all 10 volunteers (bottom panel) illustrates that enhancement of
the correlation was achieved for 1119 windows (92.25%). (D) The TCC
results are shown in the same fashion, but no correction reductions (red
lines) are observed, resulting in 100% improvement.In the 94 sampling windows (or 7.75%) for which phase-shift correction worsened
the correlation, the correlation decrease exceeded 0.1 in only 20 windows
(21.3%). Interestingly, 76 of the 94 windows (80.9%) with worse correlation
belong to volunteer 5, who had the smallest average phase shift and,
consequently, one of the highest uncorrected correlations (0.80 in scan 1 and
0.83 in scan 2), leaving little room for further improvement. Additionally, the
shapes of this volunteer's waveforms are much broader in the bellows than in the
navigator, making the optimal phase shift ambiguous and leading to a slightly
worse correlation between the two waveforms at some time points using the MPD
method. In all sampling windows for subject 5, however, the overall mean
correlation is still enhanced to a correlation of 0.86 in scan 1 and unchanged
in scan 2.Pearson's correlation coefficient between the TCC phase shifts and initial
correlations is −0.89, and a zero-parameter fit to the line, Initial
Correlation = 1 − (2/π)*abs(ϕTCC), yields an average residual error
of 0.12, and the average residual error was 0.15 for the curve, Initial
Correlation = cos(ϕTCC) (Figure 3D, top). Because the TCC method
maximizes the cross-correlation between the bellows and navigator waveforms,
correcting the TCC phase shift never worsened the correlation. Across all
sampling windows, the correlation rose by 0.40 on average and was uncharged in
only 10 windows (0.82%) (Figure 3D, bottom).Table 3 provides the
quantitative correlation enhancements for both scans using the MPD and TCC
methods. Averaged across all 10 volunteers, the correlation increases
substantially from 0.56 ± 0.22 for the original navigator and bellows waveforms
to 0.85 ± 0.11 (p = .0003) after correcting the MPD phase shift
and to 0.89 ± 0.06 (p = .0002) after correcting the TCC phase
shift in scan 1, and from 0.47 ± 0.30 to 0.84 ± 0.08 (p = .002)
after MPD correction and to 0.87 ± 0.06 (p = .001) after TCC
correction in scan 2. Both methods yield similar enhancements in the correlation
between the bellows and navigator waveforms. The set of mean enhanced
correlations for the 10 volunteers after correcting the MPD phase shift is
indistinguishable from that for the TCC method (p = .33 for
scan 1, p = .31 for scan 2), and the %Diff between the two
methods averaged across all 10 volunteers is 0.06 ± 0.11 in scan 1 and
0.04 ± 0.03 in scan 2.
Table 3.
Correlation Enhancement Following Phase-Shift Corrections in Both Scans
Using the MPD Method or the TCC Methoda.
Subject
Correlation enhancement in Scan
1
Correlation enhancement in Scan
2
Original
By MPD
By TCC
%Diffb
Original
By MPD
By TCC
%Diff*
Mean
SD
Mean
SD
Mean
SD
Mean
SD
Mean
SD
Mean
SD
1
0.79
0.01
0.96
0.01
0.96
0.01
0.00
0.72
0.10
0.90
0.05
0.91
0.05
0.01
2
0.59
0.04
0.93
0.01
0.94
0.01
0.01
0.40
0.11
0.79
0.12
0.83
0.10
0.05
3
0.81
0.05
0.96
0.03
0.96
0.03
0.00
0.85
0.03
0.98
0.01
0.99
0.01
0.01
4
0.33
0.25
0.86
0.15
0.87
0.12
0.01
0.28
0.25
0.85
0.15
0.88
0.09
0.04
5
0.80
0.09
0.88
0.09
0.91
0.07
0.03
0.83
0.11
0.83
0.13
0.88
0.08
0.06
6
0.59
0.14
0.87
0.09
0.90
0.07
0.03
0.52
0.27
0.82
0.14
0.86
0.12
0.05
7
0.58
0.10
0.91
0.04
0.92
0.04
0.01
0.05
0.23
0.87
0.07
0.88
0.06
0.01
8
0.10
0.09
0.70
0.09
0.76
0.06
0.09
0.03
0.10
0.69
0.10
0.75
0.08
0.09
9
0.53
0.15
0.82
0.09
0.83
0.09
0.01
0.36
0.30
0.87
0.10
0.88
0.08
0.01
10
0.47
0.56
0.64
0.31
0.87
0.09
0.36
0.66
0.13
0.82
0.11
0.88
0.08
0.07
Mean
0.56
0.15
0.85
0.09
0.89
0.06
0.06
0.47
0.16
0.84
0.10
0.87
0.08
0.04
SD
0.22
0.16
0.11
0.09
0.06
0.04
0.11c
0.30
0.09
0.08
0.04
0.06
0.03
0.03
p-value
.0003
.0002
.0012
.0005
Both methods provide similar correlation enhancement. Overall, the
correlation enhancements are all significant, except for subject 10
in scan 1 and subject 5 in scan 2 using the MPD method.
%Diff = (r2−r1)/r2 × 100% is between
MPD-corrected and TCC-corrected correlations.
This SD is distorted by the outlier from patient 10 and does not
accurately represent the actual distribution.
Correlation Enhancement Following Phase-Shift Corrections in Both Scans
Using the MPD Method or the TCC Methoda.Both methods provide similar correlation enhancement. Overall, the
correlation enhancements are all significant, except for subject 10
in scan 1 and subject 5 in scan 2 using the MPD method.%Diff = (r2−r1)/r2 × 100% is between
MPD-corrected and TCC-corrected correlations.This SD is distorted by the outlier from patient 10 and does not
accurately represent the actual distribution.A statistically significant enhancement in the correlation was found for every
volunteer in both scans after correcting the TCC phase shift and for all, except
for volunteer 5 in scan 2 and for volunteer 10 in scan 1, after correcting the
MPD phase shift (Figure
4). For volunteer 5 in scan 2, despite yielding a worse correlation
in 72 of 153 windows (47.1%), correction of the MPD phase shift produced an
average correlation that was no different than the average uncorrected
correlation. For volunteer 10 in scan 1, only 7 sampling windows were available
for analysis likely contributed to the correlation enhancement not reaching
statistical significance after correcting the MPD phase shift.
Figure 4.
Phase-shift correction enhances the correlation between bellows and
navigator waveforms. A bar chart portrays the mean correlation and
standard error before (I, gray bars) and after correcting the MPD (M,
blue bars) or TCC (T, green bars) phase shift for every volunteer in
both the first scan (F, darker shades) and second scan (S, lighter
shades). An asterisk indicates that the underlying bar represents a
statistically significant enhancement compared to the initial
correlation, p < .05. Correcting the TCC phase shift
succeeded in enhancing the correlation in both scans for every
volunteer. A statistically significant correlation enhancement was not
achieved by correcting the MPD phase shift throughout the second scan
for volunteer 5 or the first scan for volunteer 10 (red “x”). The
dashed, gray line marks a correlation of 0.8 for reference.
Phase-shift correction enhances the correlation between bellows and
navigator waveforms. A bar chart portrays the mean correlation and
standard error before (I, gray bars) and after correcting the MPD (M,
blue bars) or TCC (T, green bars) phase shift for every volunteer in
both the first scan (F, darker shades) and second scan (S, lighter
shades). An asterisk indicates that the underlying bar represents a
statistically significant enhancement compared to the initial
correlation, p < .05. Correcting the TCC phase shift
succeeded in enhancing the correlation in both scans for every
volunteer. A statistically significant correlation enhancement was not
achieved by correcting the MPD phase shift throughout the second scan
for volunteer 5 or the first scan for volunteer 10 (red “x”). The
dashed, gray line marks a correlation of 0.8 for reference.
Improvement in Respiratory-Gated Radiotherapy via Correlation
Enhancement
Figure 5 illustrates how
correcting the phase shift improves the accuracy of amplitude-triggered
respiratory gating around full exhalation. By translating the bellows signal in
time, phase-shift correction shifts the beam-on window to within the navigator's
acceptable motion margin: the bellows triggers the radiation beam to turn on at
more appropriate times and the %harm (percentage of beam-on time that radiation
is delivered outside the motion margin) to the patient is reduced. Table 4 tabulates
each volunteer's motion margin, the median, and 95% CI for the navigator's
position at the instant the beam turns on, and the %harm. Before correcting the
phase shift, only 4 cases achieve acceptable gating, defined as <10% harm.
The number of cases achieving acceptable gating increases to 9 cases after
correcting the MPD phase shift and to 17 cases after the TCC correction.
Baseline drifts—with respect to the navigator's position during the initial 10 s
of a scan—occur in some cases and result in negative values for the lower bound
of the 95% CI. The baseline drifts cannot be rectified by phase-shift
correction, but maybe addressed by including a 5.0 mm setup margin.
Figure 5.
Phase-shift correction improves the accuracy of amplitude-triggered
respiratory gating. A theoretical threshold for triggering the radiation
beam to turn on is set at 30% of the bellows’ amplitude from the initial
10s of a 4DMRI scan (dashed, horizontal line in the top graphs) and an
acceptable margin for internal motion is set at 50% of the navigator's
amplitude during the same timeframe (dashed, horizontal line in the
bottom graphs). Based on the gating strategy, the colored areas indicate
the times that the radiation beam would theoretically be on. (A) Before
correcting the phase shift, the beam-on is triggered before the
navigator reaches the acceptable motion margin and causes theoretical
harm; the red areas encompass the times that the beam is aberrantly
turned on. The %harm is calculated as the sum of the widths of the red
areas divided by that of the red and green areas. The red arrows
indicate the navigator's position at the instant beam-on time; the
median and 95% CI for these positions are tabulated in Table 4. (B)
After correcting the phase shift, the bellows triggers the beam-on
entirely within the navigator's motion margin, yielding 100% gating
accuracy in this example (green areas). Because the bellows most often
leads the navigator, the radiation beam usually turns off (orange
arrows) well within the navigator's motion margin and does not
contribute to the theoretical harm. Because the shapes of the bellows
and navigator waveforms differ significantly, the bellows-triggered
gating is inefficient and remains an area of active investigation.
Table 4.
The Motion Margin, the Median, and 95% Confidence Interval (CI) for the
Navigator's Position at the Instant the Beam Turns On, and the
Percentage of Harm (Percentage of Beam-On Time That Radiation Is
Delivered Outside The Motion Margin) Before and After Correcting the MPD
and TCC Phase Shifts
.
Imaging session
Vol
Motion margin (mm)
Original
After MPD correction
After TCC correction
median (mm)
95% CI (mm)
% harm
median (mm)
95% CI (mm)
% harm
median (mm)
95% CI (mm)
% harm
Scan 1
1
5.0
6.6
4.8–6.9
15
5.9
4.5–6.5
9
4.7
2.7–4.9
0
2
6.8
8.2
0.7–11.0
16
6.2
0.5–8.9
4
3.1
0.5–4.5
0
3
7.7
19.2
8.9–43.7
41
16.6
6.7–37.6
39
13.7
3.8–28.5
37
4
2.8
0.8
−1.1–4.2
6
0.6
−1.0–3.7
4
0.4
−1.3–2.9
1
5
8.8
19.3
9.6–41.2
60
17.7
8.3–39.4
59
16.6
7.1–36.5
58
6
7.7
10.4
−4.0–38.5
19
7.4
−3.3–30.5
11
3.6
−3.7–14.7
5
7
7.7
5.7
0.3–27.0
33
4.2
0.0–23.9
25
2.7
0.1–15.3
16
8
11.6
20.8
1.1–30.3
45
15.1
0.5–24.9
18
3.1
0.0–10.6
0
9
3.6
0.2
−1.1–6.1
8
0.2
−1.1–5.2
3
0.2
−0.9–3.2
0
10
12.0
6.2
0.0–19.3
5
4.5
−0.2–14.5
2
3.1
−0.2–10.6
0
Avg
25
17
12
SD
19
19
20
Scan 2
1
6.9
7.0
0.3–11.6
13
6.1
0.1–10.5
7
4.8
0.4–8.2
1
2
6.9
10.2
2.2–18.6
32
6.4
1.5–12.9
11
2.2
0.6–5.7
0
3
9.1
11.3
6.4–21.7
10
9.6
5.4–18.4
5
6.6
3.0–12.2
1
4
35.0
65.9
1.7–117.6
40
57.9
0.2–39.6
25
19.9
1.0–40.2
1
5
11.0
−1.3
−9.7–18.0
3
−1.9
−9.7–18.7
3
−2.3
−9.8–17.4
3
6
7.7
7.1
1.4–23.7
23
4.8
0.7–16.2
9
2.7
−0.4–5.5
6
7
9.1
19.3
1.5–34.0
58
16.8
0.2–26.0
40
7.5
0.2–18.5
7
8
6.6
16.1
1.7–24.3
53
10.8
1.4–19.5
21
2.5
0.7–5.5
0
9
4.0
1.4
−0.6–15.7
35
1.3
−0.6–11.1
22
1.0
−0.6–6.5
3
10
14.0
5.2
−7.6–53.2
17
3.6
−7.6–39.1
10
2.8
−7.8–29.7
5
Avg
28
15
3
SD
18
11
3
Among 20 cases of 10 volunteers (vol), only 4 cases produce
<10%harm before phase-shift correction, but the number increases
to 9 cases after the MPD correction and to 17 cases after the TCC
correction. A negative value in the lower bound of the 95% CI
indicates a baseline drift in the navigator from its position during
the initial 10 s of the 4DMRI scan. Because the baseline drift is
independent of the phase shift, neither method for correcting the
phase shift can rectify the drift. Five cases contain drifts that
exceed 5.0 mm and cannot be addressed by a 5.0 mm setup margin.
Phase-shift correction improves the accuracy of amplitude-triggered
respiratory gating. A theoretical threshold for triggering the radiation
beam to turn on is set at 30% of the bellows’ amplitude from the initial
10s of a 4DMRI scan (dashed, horizontal line in the top graphs) and an
acceptable margin for internal motion is set at 50% of the navigator's
amplitude during the same timeframe (dashed, horizontal line in the
bottom graphs). Based on the gating strategy, the colored areas indicate
the times that the radiation beam would theoretically be on. (A) Before
correcting the phase shift, the beam-on is triggered before the
navigator reaches the acceptable motion margin and causes theoretical
harm; the red areas encompass the times that the beam is aberrantly
turned on. The %harm is calculated as the sum of the widths of the red
areas divided by that of the red and green areas. The red arrows
indicate the navigator's position at the instant beam-on time; the
median and 95% CI for these positions are tabulated in Table 4. (B)
After correcting the phase shift, the bellows triggers the beam-on
entirely within the navigator's motion margin, yielding 100% gating
accuracy in this example (green areas). Because the bellows most often
leads the navigator, the radiation beam usually turns off (orange
arrows) well within the navigator's motion margin and does not
contribute to the theoretical harm. Because the shapes of the bellows
and navigator waveforms differ significantly, the bellows-triggered
gating is inefficient and remains an area of active investigation.The Motion Margin, the Median, and 95% Confidence Interval (CI) for the
Navigator's Position at the Instant the Beam Turns On, and the
Percentage of Harm (Percentage of Beam-On Time That Radiation Is
Delivered Outside The Motion Margin) Before and After Correcting the MPD
and TCC Phase Shifts
.Among 20 cases of 10 volunteers (vol), only 4 cases produce
<10%harm before phase-shift correction, but the number increases
to 9 cases after the MPD correction and to 17 cases after the TCC
correction. A negative value in the lower bound of the 95% CI
indicates a baseline drift in the navigator from its position during
the initial 10 s of the 4DMRI scan. Because the baseline drift is
independent of the phase shift, neither method for correcting the
phase shift can rectify the drift. Five cases contain drifts that
exceed 5.0 mm and cannot be addressed by a 5.0 mm setup margin.
Discussion
Subject-Specific Phase Shifts Estimated by the Phase- and Time-Domain
Methods
By combining three individual methods with each weighted by the reciprocal of its
estimated RE (see Eq. 1), the MPD method succeeds in detecting the phase shift
and enhancing the internal–external correlation in more than 99% of the analyzed
time windows. The 10 s windows were advanced in time by 5 s
increments—dynamically updating the window by adding 5 s of new data and
removing 5 s of old data—to capture approximately one new breathing cycle in
each window. The moving time window allows dynamic phase-shift detection,
produces a robust phase-shift estimation, and minimizes the impact of sudden
breathing irregularities. This may suggest the origin of the stability of the
subject-specific phase shifts, even with large breathing irregularities, as
shown in Figures 2 and
3.The results of phase-shift detection and motion correlation enhancement are
confirmed by the TCC method. If a low correlation arises from a phase shift
between two waveforms, then the phase shift should be observed in both the phase
and time domains. The close agreement in the correlation enhancement achieved by
the phase-domain and time-domain methods suggests an accurate estimation of the
phase shift and accords with the proposition that the phase shift constitutes
the primary cause of the originally weak correlations between the bellows and
navigator waveforms. The consistency of the results from two independent
approaches indicates that phase-shift correction is an effective means to
enhance the external–internal correlation.Based on clinical observations of patients’ breathing behaviors,[16,29,35,36] it was
reasonable to hypothesize that some patient-specific respiratory features,
including the phase shift, may remain relatively invariant over a short period,
such as 30 min. Physiologically, thoracoabdominal movements during respiration
are initiated by a joint effort of the diaphragm and intercostal muscles, which
can be quantified as the ratio of thoracic to abdominal involvement during respiration.
Because little volitional control is exerted over respirations during
free-breathing, human subjects exhibit their natural breathing behavior—governed
by involuntary muscle actions—in this state. A phase shift should therefore
represent a fundamental feature of an individual's free-breathing pattern, and
differences in anatomy and physiology will engender variation among subjects in
the magnitudes of their respiratory phase shifts. Although many individuals may
exhibit irregularities in the amplitude and frequency of their breathing
patterns, switching between classes of breathing patterns—such as from chest
breathing to belly breathing—usually requires voluntary actions. Whereas
inhalation is an active process triggered by contraction of the diaphragm and
intercostal muscles, exhalation is a passive process of muscle relaxation and is
consequently more reproducible, particularly the state of a full exhalation. The
relationship between external surface motion and internal muscle action should,
consequently, remain steady with mild fluctuations.
The detected phase shift does, in fact, fluctuate (Figures 3A and 3B and Table 2), but the enhanced
correlation, nevertheless, is fairly stable over time (Figures 3C and 3B and Table 3) owing to its tolerance to
small, residual phase shifts estimated by either the MPD or TCC method (further
discussion is provided in the following sub-section).Numerous physiological and technical factors contribute to the phase shifts that
are detected between the motions of internal and external structures. Unlike the
real-time position management (RPM) that senses the motion of the body's
anterior surface, the bellows senses the pressure differences around the
circumference (or perimeter) of the body during respiration. In fact, a phase
shift between the RPM and bellows was previously reported.
When it is placed just inferior to the xiphoid process of the sternum,
the bellows encircles the lateral and posterior ribs and therefore senses the
movements of both the diaphragm and the intercostal muscles. By contrast, the
internal navigator senses only the motion of the diaphragm. The external and
internal surrogates therefore represent somewhat different motions. These
differences arise, in part, from the motion delays engendered by the elasticity
of the soft tissues that separate the internal and external structures. If the
bellows is instead placed around the umbilicus, the rib cage and the motion of
the intercostal muscles contained therein are excluded. In this scenario, the
bellows may primarily sense the motion caused by the diaphragm and a phase shift
that differs from that for placing the bellows at the xiphoid may be
obtained.[4,38] In summary, surrogate types and placements,
muscle-engagement profiles, tissue elasticities, and variations in breathing
patterns may all affect measured phase shifts.[4,35,39,40] Notwithstanding the
complexity of the phenomenon, the phase shift may be stable during involuntary
free-breathing, possibly until a voluntary action occurs.
The Stability and Reproducibility of the Subject-Specific Phase Shift and
Correlation Enhancement
In this study, the average ratio of the phase shifts estimated by the MPD between
the two scans is 1.07 ± 0.45 (0.95 ± 0.19, excluding 2 outliers: patients 5 and
7), illustrating a rather reproducible phase shift over 30 min. On average, the
correlation is enhanced from 0.56 ± 0.22 to 0.85 ± 0.11 in scan 1 and from
0.47 ± 0.30 to 0.84 ± 0.08 in scan 2 by correcting the phase shifts detected by
the MPD method. For the two subjects with the largest phase shifts (1.2 and 1.6
radians for subjects 4 and 8, respectively) and no apparent initial correlation
(r = 0.33 and r = 0.10), correcting the
phase shift enhances the correlation to 0.86 and 0.70 in scan 1 and to 0.85 and
0.69 in scan 2, respectively. Correcting the phase shift reliably enhances the
correlation to above 0.8 over timeframes lasting up to—and perhaps exceeding—30
min in 8 out of 10 subjects for the MPD method and in 9 out of 10 subjects for
the TCC method. Phase-shift correction, therefore, exhibits the fidelity and
stability that are requisite criteria for clinical applications.There is close agreement (%Diff ≤ 0.05) between the correlation enhancements
attained by the MPD and TCC methods for 8 subjects in scan 1 and for 7 subjects
in scan 2. Although the performances of the MPD and TCC methods are comparable,
focusing on the phase shift, rather than on the more variable time shift, may
allow algorithms that correct the shift between respiratory waveforms to update
less frequently while still achieving a steady, enhanced correlation.In accord with previously reported findings for external–internal motion models,
the degree of correlation enhancement achieved in this study indicates that the
phase shift is the primary cause for the initially low correlations.[41,42] Aside
from the phase shift, baseline drifts in the respiratory waveforms may also
degrade the correlation between external and internal motions.
A baseline drift would certainly affect the targeting during radiotherapy
and should consequently be corrected in correlation-based models for predicting
tumor motion. Because a baseline drift unfolds over low frequencies, it could be
represented as a separate, dedicated term in a prediction model, rather than
entangling this process with the respiratory correlation term. In a simplified
model, a baseline drift could even be regarded as linear over the timeframe of a
few breathing cycles, but exploring this phenomenon is beyond the scope of the
current investigation.
Correlation Enhancement by Dynamic Phase-Shift Correction Using the MPD
Method
The temporal resolution of the phase-shift calculation by the MPD method is
approximately equal to the length of the time window, which should cover at
least one breathing cycle. Time windows encompassing 12.5 s and 7.5 s of data
were previously examined and produced results that were similar and
statistically indistinguishable from the results obtained with the10 s windows,
suggesting that any of these window lengths are reasonable choices. Owing to its
limited temporal resolution, the MPD method tracks slow changes in the phase
shift well but may lag behind abrupt changes by the length of a time window. In
this study, a small number of sharp changes are observed in the TCC-estimated
phase shift, suggesting that rapid changes in the phase shift occur only
occasionally and that the 10 s resolution of the MPD method is adequate to track
changes in rather stable phase shifts. As discussed above, even if the
phase-shift detection fails to adapt rapidly to a fast-changing event, the
impact on the correlation enhancement should be minimal owing to a moderate
tolerance for uncertainties in the phase shift. The fact that the phase shift
remains relatively steady over 30 min for most subjects suggests a simple
solution to improve respiratory-motion management during gated radiotherapy. For
clinical applications, however, a combination of the TCC and MPD methods may be
applied to balance the shortcomings of each method and enable even more robust
predictions.In this study, the MPD method fails to produce a phase-shift estimate in only
∼0.1% of the analyzed time windows, meaning that all three individual methods
failed simultaneously. Abrupt breathing irregularities may occur in these rare
events, producing—for example—a phase-domain trajectory that resembles two
disparately oriented ovals that, together, cannot be accurately represented by a
single ellipse, and throws off the MPD calculation. After subdividing such a
time window into two parts that each contained only one of these ovals, the MPD
method succeeded in estimating phase shifts of approximately 1.45 radians,
values that are similar to phase shifts found for subsequent time windows. The
sudden reorientation of the phase-space trajectory may reflect an abrupt change
in the subject's breathing pattern. When these rare events are detected, the
treatment beam can be temporarily held, like gating, until the MPD method is
again able to estimate the phase shift.
The Potential Usefulness of Phase-Shift Correction in Respiratory-Gated
Radiotherapy
Correcting the phase shift improves the validity of the amplitude-triggered
gating technique (see Figure
5). To account for the large uncertainties—even after phase-shift
correction—in predicting the motion of an internal organ, we chose a
conservative threshold for triggering the radiation beam. Setting the triggering
threshold at 30% of the bellows’ amplitude from full exhalation ensures that the
majority of radiation would be delivered within an acceptable motion margin,
defined to be 50% of the navigator's amplitude. Before correcting the phase
shift, only 4 out of 20 cases achieve acceptable gating, defined as <10%harm.
After correcting the phase shift, the number of cases that achieve acceptable
gating rises to 9 cases using the MPD correction and to 17 cases using the TCC
correction. The significant improvement in the number of cases that are suitable
for RGRT suggests a potential clinical role for phase-shift correction.Baseline drifts in the navigator waveforms were observed in some of the scans.
The lower bound in the 95% CI of the navigator's position when the radiation
beam is turned on should ideally be close to zero: an upward baseline shift
would tend to increase the value of this lower bound and a downward shift would
tend to decrease the value. The drift cannot be rectified by correcting the
phase shift, but the 5.0 mm setup margin can cover most minor baseline drifts.
There are, however, 5 cases (25%) wherein the baseline drift exceeds 5.0 mm.
Monitoring the baseline over a longer scan period—which should better represent
the averaged respiratory baseline—may help identify a more appropriate setup
margin for the individual patient. For a long-term baseline drift at treatment,
it may become necessary to shift the couch to reset the baseline to zero or even
redo the setup.We examined how phase-shift correction affects amplitude-triggered respiratory
gating. Phase gating is also employed as an alternative strategy to account for
respiratory motion. Regardless of the gating scheme, the gating window is
usually determined at simulation by a physician based on 4DCT, such as 30–70% as
the gating window (50% is the full exhalation). For amplitude-gated treatment,
it can be converted to an amplitude-gating window.
In general, phase gating depends on online phase determination and is
more sensitive to breathing irregularities.[43,44] The MPD method can offer
online phase determination with high accuracy and resolution (compared with the
10-bin 4DCT at simulation), showing the potential to improve the robustness of
phase gating. Additionally, a stable phase shift between external and internal
motions can also help the phase gating to treat an internal target. Therefore,
by evaluating the variability of a patient's phase shift during simulation, the
MPD method could identify suitable candidates with stable phase shifts for
phase-gated radiotherapy.
Clinical Considerations for Potential Phase-Shift-Corrected External
Surrogates
Because this study focuses on healthy volunteers, the movement of the
diaphragm—the organ with the greatest superior-to-inferior motion during
free-breathing—was monitored. In patients, the same methodology used for
tracking the diaphragm's motion can be applied instead of monitoring the motion
of lung or liver tumors. Specifically, it is possible to place a small navigator
window on the tumor–lung interface to measure the motion of a sizable lung
tumor, i.e., a tumor whose diameter exceeds 2 cm. Alternatively, the
time-resolved 4DMRI with a 2 Hz frame rate[36,45-47] can be applied to
investigate the correlation between the motions of an external structure and an
internal tumor. Depending on the clinical needs, the dynamic MR imaging data can
be used to build either a simple phase-shift-corrected respiratory model or a
sophisticated physics-based perturbation model using surface-imaging-based
spirometry as a surrogate.[48,49]Diaphragm motion, which has been investigated as an important surrogate for
predicting the motion of lung and liver cancers,[50-52] is readily detected using
fluoroscopic imaging in the clinic.
A study has illustrated only a slight superiority of correlating an
external motion with multiple internal surrogates compared to a single internal surrogate.
Therefore, a method that enhances the correlation between the bellows and
diaphragm motions could be extended to predict tumor motion with only minor
modifications.Based on the results from this study, patients could be categorized in terms of
their external–internal motion correlation into three groups: (1) naturally high
correlation, (2) enhanced high correlation, and (3) enhanced correlation but not
high enough. This study has demonstrated that phase-shift correction can be
applied to make most patients have sufficient correlation. Whereas only 2 out of
10 subjects in both scans originally have correlation coefficients above 0.8,
correcting the phase shift raises that number to 8 out of 10 subjects with
enhanced correlations exceeding 0.8. Therefore, 80% of the subjects in this
study possess respiratory patterns that are appropriate for respiratory gating.
During simulation, the MPD and TCC methods can identify suitable subjects for
respiratory-gated radiotherapy planning.A clinical workflow for applying the phase-shift correction in an RGRT trial
could follow the following seven steps: (1) simulate patient FB motion during
the simulation 4DMRI scan and acquire concurrent external and internal
waveforms, (2) assess the patient-specific phase shift and the degree of
enhancement of the correlation between external and internal waveforms by
correcting the phase shift, (3) select for RGRT planning only those patients
whose enhanced correlation exceeds 0.9, (4) simulate bellows-triggered gating to
select a beam-on threshold that minimizes the %harm by delivering radiation only
within an acceptable motion margin (i.e., set a gating threshold at around 30%
of the external surrogate's amplitude to achieve a motion margin of 50% or less
of the internal target's amplitude), (5) use the simulated motion margin to
design the patient's RGRT plan and devise a backup plan that accounts for the
full-motion margin, (6) assess the correlation on the day of treatment by
acquiring 20–30 s of fluoroscopic imaging to confirm the accuracy of the planned
gating before initiating treatment, and (7) during treatment, radiographic
imaging may be acquired periodically to verify the phase shift and check
baseline to ensure accurate respiratory gating. Initially, it may take more
effort to handle various clinical events and scenarios as any implementation of
new procedures will. This motion assessment and management strategy could be
useful for future MR-only RGRT simulation and planning and delivering the
treatment using a conventional LINAC system.Due to the differences between the shapes of the bellows and navigator waveforms,
the efficiency of amplitude-triggered respiratory gating is relatively low (low
duty cycle). Because the radiation beam may turn off too early, a large portion
of the time that the navigator transits its motion margin goes unused. Further
investigation—such as a deep-learning approach—may establish a patient-specific
model that optimizes the accuracy and efficiency of respiratory gating.In the future, strategies for enhancing the correlation between the motions of
the diaphragm and a tumor and between the motions of the bellows and a tumor can
also be investigated. It may then be possible to use the diaphragm as the
internal surrogate for the motion of lung or liver tumors. With 20–30 s of
fluoroscopic data and concurrent data acquired from an external surrogate, the
patient's phase shift can be determined immediately before treatment and
corrected—thereby ensuring enhanced correlation between the motions of an
external surrogate and the tumor—throughout radiotherapy sessions lasting up to
30 min. The automatic motion-sampling tool can be readily applied to clinical
applications. The correlation between the diaphragm and tumor can also be
obtained during a simulation using TR-4DMRI. The strategy described herein for
correcting the phase shift can readily be used to identify patients with severe
breathing irregularities
who are unsuitable candidates for respiratory-gated radiotherapy.
Conclusions
This study has demonstrated that the external–internal motion correlation can be
significantly and stably enhanced to a high value (eg, r > 0.8) for 30 min of
free-breathing by dynamically correcting a subject-specific phase shift. This result
is supported by both the MPD and TCC methods. The subject-specific phase shifts tend
to be stable over 30 min with small fluctuations, which can be well tolerated by the
correlation-enhancement methods. Correcting the phase shift between an external
surrogate and the internal target significantly improves the accuracy of
amplitude-triggered respiratory gating, rendering correlation-based strategies safer
and better equipped to guide radiotherapy.
Authors: Ruijiang Li; John H Lewis; Xun Jia; Tianyu Zhao; Weifeng Liu; Sara Wuenschel; James Lamb; Deshan Yang; Daniel A Low; Steve B Jiang Journal: Phys Med Biol Date: 2011-08-24 Impact factor: 3.609
Authors: Guang Li; Naveen C Arora; Huchen Xie; Holly Ning; Wei Lu; Daniel Low; Deborah Citrin; Aradhana Kaushal; Leor Zach; Kevin Camphausen; Robert W Miller Journal: Phys Med Biol Date: 2009-03-05 Impact factor: 3.609