A clustering approach to 4D MRI retrospective sorting for the investigation of different surrogates.

Giorgia Meschini1, Chiara Paganelli2, Chiara Gianoli3, Paul Summers4, Massimo Bellomi5, Guido Baroni6, Marco Riboldi3.   

Abstract

PURPOSE: In retrospective 4-Dimensional Magnetic Resonance Imaging (4D MRI) sorting, respiratory surrogate selection affects the image quality of reconstructed volumes. We propose a method for retrospective 4D MRI sorting based on clustering, which allowed us to compare the performance of single or multiple internal surrogates vs. a conventional external signal.
METHODS: A k-medoids clustering algorithm was exploited for sorting 2D MRI into 4D MRI, relying on (A) multiple or (B) single automatically tracked internal landmarks or (C) respiratory belt signal. 4D MRI reconstructions for seven liver cancer patients were compared to those of the state-of-the-art mutual information (MI) approach. Sorting artifacts were measured by the root mean square error (RMSE) between the diaphragm profile and a fitted second order curve. Diaphragm and tumor motions were evaluated.
RESULTS: The median RMSEs ranged 0.97-1.66 mm, 1.24-1.89 mm, 1.43-2.27 mm, 1.74-3.72 mm for the MI, (A), (B) and (C) methods, respectively. Significant differences (Friedman, α = 5%) were found between (C) and all other methods, and between (B) and MI approaches. The discrepancies between (A) and MI approaches ranged 1.1-6.2 mm and 0.7-5.3 mm respectively in diaphragm and tumor motions. Methods (A) and (B) showed similar ranges of motion.
CONCLUSION: With multiple internal points, our method yielded the description of a higher range of motion and similar image quality with respect to the MI approach. The single point method led to more artifacts, suggesting the superior suitability of multiple internal surrogates for retrospective 4D MRI sorting. Considering internal rather than external information favored superior performance.
Copyright © 2019 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  4D MRI; Liver; Retrospective sorting

Mesh:

Year:  2019        PMID: 30824141     DOI: 10.1016/j.ejmp.2019.02.003

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Time-resolved MRI for off-line treatment robustness evaluation in carbon-ion radiotherapy of pancreatic cancer.

Authors:  Giorgia Meschini; Alessandro Vai; Amelia Barcellini; Giulia Fontana; Silvia Molinelli; Edoardo Mastella; Andrea Pella; Viviana Vitolo; Sara Imparato; Ester Orlandi; Mario Ciocca; Guido Baroni; Chiara Paganelli
Journal:  Med Phys       Date:  2022-02-17       Impact factor: 4.506

2.  Stability and Reliability of Enhanced External-Internal Motion Correlation via Dynamic Phase-Shift Corrections Over 30-min Timeframe for Respiratory-Gated Radiotherapy.

Authors:  Andrew Milewski; Guang Li
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.