Novel spirometry based on optical surface imaging.

Guang Li1, Hailiang Huang1, Jie Wei2, Diana G Li1, Qing Chen1, Carl P Gaebler1, James Sullivan3, Joan Zatcky4, Andreas Rimner4, James Mechalakos1.   

Abstract

PURPOSE: To evaluate the feasibility of using optical surface imaging (OSI) to measure the dynamic tidal volume (TV) of the human torso during free breathing.
METHODS: We performed experiments to measure volume or volume change in geometric and deformable phantoms as well as human subjects using OSI. To assess the accuracy of OSI in volume determination, we performed experiments using five geometric phantoms and two deformable body phantoms and compared the values with those derived from geometric calculations and computed tomography (CT) measurements, respectively. To apply this technique to human subjects, an institutional review board protocol was established and three healthy volunteers were studied. In the human experiment, a high-speed image capture mode of OSI was applied to acquire torso images at 4-5 frames per second, which was synchronized with conventional spirometric measurements at 5 Hz. An in-house matlab program was developed to interactively define the volume of interest (VOI), separate the thorax and abdomen, and automatically calculate the thoracic and abdominal volumes within the VOIs. The torso volume change (TV C = ΔVtorso = ΔVthorax + ΔVabdomen) was automatically calculated using full-exhalation phase as the reference. The volumetric breathing pattern (BPv = ΔVthorax/ΔVtorso) quantifying thoracic and abdominal volume variations was also calculated. Under quiet breathing, TVC should equal the tidal volume measured concurrently by a spirometer with a conversion factor (1.08) accounting for internal and external differences of temperature and moisture. Another matlab program was implemented to control the conventional spirometer that was used as the standard.
RESULTS: The volumes measured from the OSI imaging of geometric phantoms agreed with the calculated volumes with a discrepancy of 0.0% ± 1.6% (range -1.9% to 2.5%). In measurements from the deformable torso/thorax phantoms, the volume differences measured using OSI imaging and CT imaging were 1.2% ± 2.1% (range -0.5% to 3.6%), with a linear regression fitting (slope = 1.02 and R(2) = 0.999). In volunteers, the relative error in OSI tidal volume measurement was -2.2% ± 4.9% (range -9.2% to 4.8%) and a correlation of r = 0.98 was found with spirometric measurement. The breathing pattern values of the three volunteers were substantially different from each other (BPv = 0.15, 0.45, and 0.32).
CONCLUSIONS: This study demonstrates the feasibility of using OSI to measure breathing tidal volumes and breathing patterns with adequate accuracy. This is the first time that dynamic breathing tidal volume as well as breathing patterns is measured using optical surface imaging. The OSI-observed movement of the entire torso could serve as a new respiratory surrogate in the treatment room during radiation therapy.

Entities:  

Mesh:

Year:  2015        PMID: 25832058      PMCID: PMC4368594          DOI: 10.1118/1.4914391

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  17 in total

1.  Comparison of spirometry and abdominal height as four-dimensional computed tomography metrics in lung.

Authors:  Wei Lu; Daniel A Low; Parag J Parikh; Michelle M Nystrom; Issam M El Naqa; Sasha H Wahab; Maureen Handoko; David Fooshee; Jeffrey D Bradley
Journal:  Med Phys       Date:  2005-07       Impact factor: 4.071

2.  On the sources of drift in a turbine-based spirometer.

Authors:  Jonathan K Ha; Daniel B Perlow; Byong Yong Yi; Cedric X Yu
Journal:  Phys Med Biol       Date:  2008-07-24       Impact factor: 3.609

3.  Coupling surface cameras with on-board fluoroscopy: a feasibility study.

Authors:  Carri K Glide-Hurst; Dan Ionascu; Ross Berbeco; Di Yan
Journal:  Med Phys       Date:  2011-06       Impact factor: 4.071

4.  Tumor motion control in the treatment of non small cell lung cancer.

Authors:  Kenneth E Rosenzweig; Ellen Yorke; Howard Amols; Gikas S Mageras; Phillippe Giraud; Matthew S Katz; Steven A Leibel
Journal:  Cancer Invest       Date:  2005       Impact factor: 2.176

5.  Rapid estimation of 4DCT motion-artifact severity based on 1D breathing-surrogate periodicity.

Authors:  Guang Li; Marshall Caraveo; Jie Wei; Andreas Rimner; Abraham J Wu; Karyn A Goodman; Ellen Yorke
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

6.  Quantitation of the reconstruction quality of a four-dimensional computed tomography process for lung cancer patients.

Authors:  Wei Lu; Parag J Parikh; Issam M El Naqa; Michelle M Nystrom; James P Hubenschmidt; Sasha H Wahab; Sasa Mutic; Anurag K Singh; Gary E Christensen; Jeffrey D Bradley; Daniel A Low
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

7.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

8.  Assessment of two novel ventilatory surrogates for use in the delivery of gated/tracked radiotherapy for non-small cell lung cancer.

Authors:  Simon Hughes; James McClelland; Segolene Tarte; David Lawrence; Shahreen Ahmad; David Hawkes; David Landau
Journal:  Radiother Oncol       Date:  2009-04-22       Impact factor: 6.280

9.  Quantitative prediction of respiratory tidal volume based on the external torso volume change: a potential volumetric surrogate.

Authors:  Guang Li; Naveen C Arora; Huchen Xie; Holly Ning; Wei Lu; Daniel Low; Deborah Citrin; Aradhana Kaushal; Leor Zach; Kevin Camphausen; Robert W Miller
Journal:  Phys Med Biol       Date:  2009-03-05       Impact factor: 3.609

10.  A novel analytical approach to the prediction of respiratory diaphragm motion based on external torso volume change.

Authors:  Guang Li; Huchen Xie; Holly Ning; Wei Lu; Daniel Low; Deborah Citrin; Aradhana Kaushal; Leor Zach; Kevin Camphausen; Robert W Miller
Journal:  Phys Med Biol       Date:  2009-06-12       Impact factor: 3.609

View more
  9 in total

1.  A Novel Respiratory Motion Perturbation Model Adaptable to Patient Breathing Irregularities.

Authors:  Amy Yuan; Jie Wei; Carl P Gaebler; Hailiang Huang; Devin Olek; Guang Li
Journal:  Int J Radiat Oncol Biol Phys       Date:  2016-09-03       Impact factor: 7.038

2.  Direct Comparison of Respiration-Correlated Four-Dimensional Magnetic Resonance Imaging Reconstructed Using Concurrent Internal Navigator and External Bellows.

Authors:  Guang Li; Jie Wei; Devin Olek; Mo Kadbi; Neelam Tyagi; Kristen Zakian; James Mechalakos; Joseph O Deasy; Margie Hunt
Journal:  Int J Radiat Oncol Biol Phys       Date:  2016-11-09       Impact factor: 7.038

3.  Characterization of optical-surface-imaging-based spirometry for respiratory surrogating in radiotherapy.

Authors:  Guang Li; Jie Wei; Hailiang Huang; Qing Chen; Carl P Gaebler; Tiffany Lin; Amy Yuan; Andreas Rimner; James Mechalakos
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

4.  Automatic assessment of average diaphragm motion trajectory from 4DCT images through machine learning.

Authors:  Guang Li; Jie Wei; Hailiang Huang; Carl Philipp Gaebler; Amy Yuan; Joseph O Deasy
Journal:  Biomed Phys Eng Express       Date:  2015-12-29

5.  Accuracy of surface-guided patient setup for conventional radiotherapy of brain and nasopharynx cancer.

Authors:  Sang Kyu Lee; Sheng Huang; Lei Zhang; Ase M Ballangrud; Michalis Aristophanous; Laura I Cervino Arriba; Guang Li
Journal:  J Appl Clin Med Phys       Date:  2021-03-31       Impact factor: 2.102

6.  Enhancement of Long-Term External-Internal Correlation by Phase-Shift Detection and Correction Based on Concurrent External Bellows and Internal Navigator Signals.

Authors:  Andrew R Milewski; Devin Olek; Joseph O Deasy; Andreas Rimner; Guang Li
Journal:  Adv Radiat Oncol       Date:  2019-03-01

7.  Stability and Reliability of Enhanced External-Internal Motion Correlation via Dynamic Phase-Shift Corrections Over 30-min Timeframe for Respiratory-Gated Radiotherapy.

Authors:  Andrew Milewski; Guang Li
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec

Review 8.  Recent advanced in Surface Guided Radiation Therapy.

Authors:  P Freislederer; M Kügele; M Öllers; A Swinnen; T-O Sauer; C Bert; D Giantsoudi; S Corradini; V Batista
Journal:  Radiat Oncol       Date:  2020-07-31       Impact factor: 3.481

9.  Use of surface-guided radiation therapy in combination with IGRT for setup and intrafraction motion monitoring during stereotactic body radiation therapy treatments of the lung and abdomen.

Authors:  John H Heinzerling; Carnell J Hampton; Myra Robinson; Megan Bright; Benjamin J Moeller; Justin Ruiz; Roshan Prabhu; Stuart H Burri; Ryan D Foster
Journal:  J Appl Clin Med Phys       Date:  2020-03-20       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.