Robert Herber1, Janine Lenk2, Lutz E Pillunat2, Frederik Raiskup2.
Abstract
BACKGROUND: To determine the repeatability and agreement using corneal tomography of a swept-source optical coherence tomographer (SS-OCT) compared to a rotating Scheimpflug camera (RSC) in normal eyes and keratoconus (KC) eyes.Entities:
Keywords: ANTERION; Cornea; Keratoconus; OCT; Pentacam; Scheimpflug; Swept-source; Tomography; Topography
Year: 2022 PMID: 35606839 PMCID: PMC9128256 DOI: 10.1186/s40662-022-00290-6
Source DB: PubMed Journal: Eye Vis (Lond) ISSN: 2326-0254
Fig. 1Comparison of topography maps of RSC (left) and SS-OCT (right) of a keratoconus with corresponding Scheimpflug image (left) and OCT scan (right). Color scaling is set to 0.5 D
Demographics for patients with normal eyes, mild, and moderate keratoconus (KC)
| Parameter | Normal eyes group | Mild KC group | Moderate KC group | |
|---|---|---|---|---|
| N | 43 | 16 | 41 | |
| Age (years) (mean ± SD) (min–max) | 29.2 ± 7.5 (18.0–45.0) | 35.6 ± 6.7 (23.0–45.0) | 30.9 ± 7.4 (18.0–43.0) | |
| Gender (m/f) (%) | 27 (63%)/16 (37%) | 10 (63%)/6 (37%) | 37 (90%)/4 (10%) | |
| Eyes (r/l) (%) | 22 (51%)/21 (49%) | 9 (56%)/7 (44%) | 15 (37%)/26 (63%) | 0.272** |
| Kmax (D) (mean ± SD) (min–max) | 43.7 ± 1.5 (40.7–46.6) | 51.5 ± 3.3 (48.1–59.7) | 59.7 ± 6.8 (47.3–73.5) | |
| I-S value (D) (mean ± SD) (min–max) | − 0.01 ± 0.66 (− 1.80–1.00) | 5.46 ± 1.74 (2.80–8.58) | 8.66 ± 2.87 (3.23–13.92) | |
| KISA (mean ± SD) (min–max) | 6.2 ± 5.9 (0.3–27.2) | 440.6 ± 331.9 (101.0–1449.0) | 4369.7 ± 5392.3 (104.0–25,295.0) | |
| BAD-D (mean ± SD) (min–max) | 0.57 ± 0.59 ( − 0.96–1.52) | 5.59 ± 0.82 (4.09–6.78) | 11.23 ± 3.22 (7.14–19.86) |
BAD-D = Belin/Ambrosio deviation value; f = female; I-S = inferior-superior; Kmax = maximum keratometry; KC = keratoconus; l = left; m = male; max = maximum; min = minimum; r = right; SD = standard deviation. *One-way ANOVA; **χ2 test. Significant differences between groups marked in bold
Repeatability of anterior corneal curvature parameters of all used devices
| Parameter | Device | Normal eyes | Mild KC | Moderate KC | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean ± SD | Sw | Repeatability | CV (%) | Mean ± SD | Sw | Repeatability | CV (%) | Mean ± SD | Sw | Repeatability | CV (%) | ||
| Flat SimK (D) | RSC | 41.9 ± 1.1 | 0.071 | 0.196 | 0.129 | 43.2 ± 1.5 | 0.182 | 0.504 | 0.309 | 48.2 ± 4.8 | 0.283 | 0.784 | 0.414 |
| SS-OCT | 42.1 ± 1.1 | 0.095 | 0.263 | 0.184 | 43.4 ± 1.4 | 0.134 | 0.372 | 0.262 | 47.6 ± 4.6 | 0.173 | 0.480 | 0.305 | |
| Steep SimK (D) | RSC | 43.2 ± 1.4 | 0.071 | 0.196 | 0.130 | 46.3 ± 2 | 0.187 | 0.519 | 0.271 | 52.0 ± 4.9 | 0.332 | 0.919 | 0.464 |
| SS-OCT | 43.4 ± 1.3 | 0.100 | 0.277 | 0.188 | 46.8 ± 2.0 | 0.145 | 0.402 | 0.271 | 51.9 ± 4.6 | 0.200 | 0.554 | 0.309 | |
| Average SimK (D) | RSC | 42.5 ± 1.1 | 0.063 | 0.175 | 0.128 | 44.7 ± 1.6 | 0.145 | 0.402 | 0.248 | 50.0 ± 4.7 | 0.265 | 0.733 | 0.351 |
| SS-OCT | 42.7 ± 1.1 | 0.089 | 0.248 | 0.163 | 45.0 ± 1.5 | 0.114 | 0.316 | 0.217 | 49.6 ± 4.4 | 0.173 | 0.480 | 0.281 | |
| Astigmatism (D) | RSC | 1.2 ± 1.0 | 0.069 | 0.190 | 7.456 | 3.2 ± 1.6 | 0.247 | 0.685 | 5.603 | 3.9 ± 2.1 | 0.313 | 0.868 | 10.038 |
| SS-OCT | 1.3 ± 1.0 | 0.082 | 0.229 | 8.618 | 3.4 ± 1.7 | 0.161 | 0.447 | 4.413 | 4.3 ± 2.2 | 0.192 | 0.533 | 4.600 | |
| Kmax (D) | RSC | 43.7 ± 1.5 | 0.134 | 0.372 | 0.233 | 51.5 ± 3.3 | 0.224 | 0.620 | 0.369 | 59.7 ± 6.8 | 0.480 | 1.329 | 0.620 |
| SS-OCT | 43.8 ± 1.4 | 0.130 | 0.361 | 0.248 | 50.9 ± 3.1 | 0.179 | 0.496 | 0.308 | 58.1 ± 6.1 | 0.283 | 0.784 | 0.377 | |
| BFS (mm) | RSC | 8.0 ± 0.2 | 0.009 | 0.025 | 0.071 | 7.7 ± 0.2 | 0.009 | 0.026 | 0.092 | 7.3 ± 0.4 | 0.014 | 0.039 | 0.144 |
| SS-OCT | 8.0 ± 0.2 | 0.010 | 0.028 | 0.099 | 7.6 ± 0.2 | 0.011 | 0.030 | 0.116 | 7.3 ± 0.4 | 0.010 | 0.028 | 0.120 | |
BFS = best-fit sphere; CV = coefficient of variation; Kmax = maximum keratometry; KC = keratoconus; repeatability = coefficient of repeatability; RSC = rotating Scheimpflug camera; SimK = simulated keratometry value; SS-OCT = swept-source optical coherence tomography; Sw = within-subject standard deviation
Fig. 2Bland-Altman plots of steep SimK (anterior), average SimK (anterior) of each subgroup
Fig. 3Bland-Altman plots of Kmax (anterior) of each subgroup
Repeatability of posterior corneal curvature parameters of all used devices
| Parameter | Device | Normal eyes | Mild KC | Moderate KC | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean ± SD | Sw | Repeatability | CV (%) | Mean ± SD | Sw | Repeatability | CV (%) | Mean ± SD | Sw | Repeatability | CV (%) | ||
| Flat SimK (D) | RSC | − 5.9 ± 0.3 | 0.035 | 0.098 | − 0.362 | − 6.2 ± 0.4 | 0.105 | 0.291 | − 0.976 | − 7.2 ± 0.9 | 0.091 | 0.251 | − 0.926 |
| SS-OCT | − 5.9 ± 0.2 | 0.005 | 0.015 | − 0.078 | − 6.1 ± 0.4 | 0.017 | 0.048 | − 0.236 | − 7.3 ± 1.0 | 0.041 | 0.114 | − 0.408 | |
| Steep SimK (D) | RSC | − 6.3 ± 0.2 | 0.028 | 0.077 | − 0.211 | − 6.8 ± 0.4 | 0.068 | 0.188 | − 0.722 | − 8.0 ± 1.0 | 0.074 | 0.206 | − 0.734 |
| SS-OCT | − 6.2 ± 0.2 | 0.008 | 0.021 | − 0.100 | − 6.9 ± 0.5 | 0.020 | 0.055 | − 0.258 | − 8.1 ± 1.0 | 0.033 | 0.092 | − 0.312 | |
| Average SimK (D) | RSC | − 6.1 ± 0.2 | 0.028 | 0.077 | − 0.220 | − 6.5 ± 0.4 | 0.065 | 0.180 | − 0.726 | − 7.6 ± 0.9 | 0.064 | 0.177 | − 0.681 |
| SS-OCT | − 6.0 ± 0.2 | 0.005 | 0.015 | − 0.073 | − 6.5 ± 0.4 | 0.014 | 0.039 | − 0.190 | − 7.7 ± 1.0 | 0.033 | 0.092 | − 0.340 | |
| Astigmatism (D) | RSC | − 0.4 ± 0.2 | 0.036 | 0.101 | − 7.107 | − 0.6 ± 0.3 | 0.120 | 0.333 | − 23.142 | − 0.8 ± 0.5 | 0.099 | 0.276 | − 14.586 |
| SS-OCT | − 0.4 ± 0.2 | 0.009 | 0.025 | − 2.720 | − 0.8 ± 0.4 | 0.022 | 0.062 | − 2.281 | − 0.8 ± 0.4 | 0.039 | 0.107 | − 4.616 | |
| Kmax (D) | RSC | − 6.5 ± 0.3 | 0.034 | 0.095 | − 0.460 | − 8.2 ± 0.7 | 0.160 | 0.443 | − 1.570 | − 9.9 ± 1.5 | 0.176 | 0.488 | − 1.459 |
| SS-OCT | − 6.3 ± 0.3 | 0.014 | 0.039 | − 0.188 | − 8.1 ± 0.8 | 0.045 | 0.124 | − 0.407 | − 9.8 ± 1.5 | 0.071 | 0.196 | − 0.519 | |
| BFS (mm) | RSC | 6.6 ± 0.2 | 0.013 | 0.036 | 0.150 | 6.4 ± 0.2 | 0.016 | 0.045 | 0.213 | 6.1 ± 0.4 | 0.020 | 0.055 | 0.237 |
| SS-OCT | 6.7 ± 0.2 | 0.010 | 0.029 | 0.100 | 6.5 ± 0.2 | 0.008 | 0.023 | 0.099 | 6.1 ± 0.4 | 0.010 | 0.028 | 0.161 | |
BFS = best-fit sphere; CV = coefficient of variation; Kmax = maximum keratometry; KC = keratoconus; repeatability = coefficient of repeatability; RSC = rotating Scheimpflug camera; SimK = simulated keratometry value; SS-OCT = swept-source optical coherence tomography; Sw = within-subject standard deviation
Repeatability of corneal thickness of all used devices
| Parameter | Device | Normal eyes | Mild KC | Moderate KC | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean ± SD | Sw | Repeatability | CV (%) | Mean ± SD | Sw | Repeatability | CV (%) | Mean ± SD | Sw | Repeatability | CV (%) | ||
| CCT (µm) | RSC | 556 ± 25 | 1.449 | 4.017 | 0.223 | 515 ± 31 | 2.366 | 6.559 | 0.307 | 466 ± 32 | 3.033 | 8.407 | 0.561 |
| SS-OCT | 549 ± 26 | 0.775 | 2.147 | 0.097 | 513 ± 30 | 0.775 | 2.147 | 0.134 | 462 ± 34 | 1.732 | 4.801 | 0.271 | |
| MCT (µm) | RSC | 553 ± 25 | 1.761 | 4.880 | 0.272 | 501 ± 30 | 2.387 | 6.618 | 0.402 | 451 ± 31 | 3.033 | 8.407 | 0.566 |
| SS-OCT | 546 ± 27 | 0.775 | 2.147 | 0.095 | 493 ± 27 | 0.894 | 2.479 | 0.125 | 444 ± 30 | 1.225 | 3.395 | 0.196 | |
CCT = central corneal thickness; CV = coefficient of variation; KC = keratoconus; MCT = minimal corneal thickness; repeatability = coefficient of repeatability; RSC = rotating Scheimpflug camera; SS-OCT = swept-source optical coherence tomography; Sw = within-subject standard deviation
Fig. 4Bland-Altman plots of central corneal thickness (CCT), and minimal corneal thickness (MCT) of each subgroup