| Literature DB >> 36048835 |
Kook Young Kim1, Seongjun Lee1, Young Joon Jeon2, Ji Sang Min2.
Abstract
PURPOSE: This study aimed to evaluate and compare the discriminating ability of corneal elevation maps generated using a swept-source optical coherence tomography (SS-OCT) (SS-OCT ANTERION, Heidelberg Engineering, Heidelberg, Germany), which was estimated with different reference surfaces, to distinguish normal corneas from those with keratoconus and keratoconus suspect.Entities:
Mesh:
Year: 2022 PMID: 36048835 PMCID: PMC9436129 DOI: 10.1371/journal.pone.0274071
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.752
Patient demographics of three groups.
| Parameters | Normal (n = 46) | KC suspect (n = 37) | KC (n = 43) | P value |
|---|---|---|---|---|
|
| 29.76 ± 10.26 | 28.81 ± 10.94 | 30.95 ± 9.11 | 0.153 |
|
| ||||
| | 22:24 | 20:17 | 20:23 | 0.776 |
|
| ||||
| | 21 (45.65%) | 20 (54.05%) | 22 (51.16%) | 0.736 |
| | 25 (54.35%) | 17 (45.95%) | 21 (48.84%) | |
|
| ||||
| | -3.28 ± 2.41 | -3.21 ± 2.77 | -2.35 ± 2.75 | 0.197 |
| | -0.94 ± 0.85 | -2.09 ± 1.86 | -2.55 ± 1.98 | <0.001 |
KC, keratoconus; D, diopter
* Kruskal-Wallis test. P < 0.05 is statistically significant.
† Chi-square teat. P < 0.05 is statistically significant.
Anterior segment parameters of ANTERION in three groups.
| Parameters | Normal (n = 46) | KC suspect (n = 37) | KC (n = 43) | P value | P value | P value | P value |
|---|---|---|---|---|---|---|---|
|
| 43.29 ± 1.73 | 44.64 ± 1.29 | 48.12 ± 6.22 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| 44.13 ± 1.96 | 45.91 ± 1.77 | 50.04 ± 7.27 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| 42.50 ± 1.66 | 43.47 ± 1.26 | 46.40 ± 5.61 | <0.001 | 0.005 | <0.001 | <0.001 |
|
| 1.62 ± 1.05 | 2.44 ± 1.60 | 3.64 ± 2.92 | <0.001 | 0.018 | 0.012 | <0.001 |
|
| 44.71 ± 2.01 | 46.73 ± 1.71 | 54.66 ± 11.05 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| -0.21 ± 0.71 | -0.08 ± 0.70 | 0.04 ± 0.49 | 0.306 | 0.956 | 0.152 | 0.220 |
|
| -0.69 ± 1.86 | -0.75 ± 1.47 | -1.17 ± 0.87 | 0.481 | 0.855 | 0.257 | 0.335 |
|
| -6.18 ± 0.28 | -6.40 ± 0.24 | -7.23 ± 1.32 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| -6.41 ± 0.34 | -6.66 ± 0.30 | -7.64 ± 1.42 | <0.001 | 0.001 | <0.001 | <0.001 |
|
| -5.97 ± 0.25 | -6.16 ± 0.23 | -6.97 ± 1.27 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| -0.44 ± 0.18 | -0.50 ± 0.22 | -0.76 ± 0.48 | <0.001 | 0.341 | 0.002 | <0.001 |
|
| -6.48 ± 0.35 | -6.80 ± 0.33 | -9.09 ± 2.67 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| 0.06 ±0.19 | -0.03 ± 0.25 | -0.02 ± 0.48 | 0.157 | 0.049 | 0.241 | 0.538 |
|
| -0.40 ± 1.68 | -0.42 ± 1.43 | -0.91 ± 1.01 | 0.322 | 0.711 | 0.105 | 0.307 |
|
| 42.74 ± 1.76 | 44.17 ± 1.38 | 48.06 ± 7.08 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| 43.49 ± 1.98 | 45.36 ± 1.89 | 49.84 ± 7.99 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| 42.00 ± 1.68 | 43.00 ± 1.25 | 46.28 ± 6.33 | <0.001 | 0.003 | <0.001 | <0.001 |
|
| 1.49 ± 1.02 | 2.35 ± 1.61 | 3.57 ± 2.71 | <0.001 | 0.013 | 0.011 | <0.001 |
|
| 539.74 ± 32.07 | 535.95 ± 26.65 | 492.97 ± 49.47 | <0.001 | 0.586 | <0.001 | <0.001 |
|
| 537.11 ± 31.56 | 532.43 ± 26.55 | 475.19 ± 52.31 | <0.001 | 0.501 | <0.001 | <0.001 |
|
| -0.01 ± 0.40 | -0.04 ± 0.49 | 0.06 ± 0.63 | 0.536 | 0.748 | 0.288 | 0.421 |
|
| -0.38 ± 0.23 | -0.42 ± 0.31 | -0.71 ± 0.47 | <0.001 | 0.728 | <0.001 | <0.001 |
|
| 2.53 ± 1.44 | 5.04 ± 2.17 | 22.65 ± 18.56 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| 2.57 ± 1.23 | 5.72 ± 2.08 | 24.36 ± 19.68 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| 5.51 ± 4.24 | 10.35 ± 7.72 | 57.01 ± 43.88 | <0.001 | <0.001 | <0.001 | <0.001 |
|
| 5.66 ± 3.44 | 11.60 ± 7.71 | 60.93 ± 47.39 | <0.001 | <0.001 | <0.001 | <0.001 |
KC, keratoconus; D, diopter; BFS, best-fit sphere; BFT, best-fit toric-ellipsoid
* Kruskal-Wallis test. P < 0.05 is statistically significant.
† Mann-Whitney U test with Bonferroni correction between the normal and suspected KC groups. P < 0.05 is statistically significant.
‡ Mann-Whitney U test with Bonferroni correction between the normal and KC groups. P < 0.05 is statistically significant.
§ Mann-Whitney U test with Bonferroni correction between the KC suspect and KC groups. P < 0.05 is statistically significant.
Fig 1Receiver operator characteristic (ROC) curves analysis for keratoconus versus keratoconus suspect or normal cornea.
(A) ROC curves for keratoconus versus normal and keratoconus suspect cornea (B) ROC curves for normal versus keratoconus cornea (C) ROC curves for keratoconus suspect versus normal cornea (D) ROC curves for keratoconus versus keratoconus suspect cornea.
The cut-off point, specificity, and sensitivity values identified by the different reference surface.
| Thinnest point | AUROC | SE | P | Cut-off (μm) | Se (%) | Sp (%) | PLR | NLR | |
|---|---|---|---|---|---|---|---|---|---|
|
|
| 0.967 | 0.013 | 0.000 | 4.97 | 100 | 80 | 4.88 | 0 |
|
| 0.969 | 0.013 | 0.000 | 8.83 | 84 | 96 | 23.16 | 0.17 | |
|
| 0.953 | 0.017 | 0.000 | 14.92 | 84 | 94 | 13.90 | 0.17 | |
|
| 0.961 | 0.017 | 0.000 | 16.44 | 86 | 98 | 35.71 | 0.14 | |
|
|
| 0.836 | 0.047 | 0.000 | 3.73 | 76 | 89 | 6.96 | 0.27 |
|
| 0.925 | 0.031 | 0.000 | 4.03 | 81 | 89 | 7.46 | 0.21 | |
|
| 0.749 | 0.054 | 0.000 | 9.68 | 54 | 88 | 6.22 | 0.50 | |
|
| 0.841 | 0.043 | 0.000 | 10.14 | 60 | 91 | 6.84 | 0.44 | |
|
|
| 0.996 | 0.004 | 0.000 | 4.95 | 100 | 98 | 46.00 | 0 |
|
| 0.999 | 0.001 | 0.000 | 5.06 | 98 | 100 | n.a. | 0.02 | |
|
| 0.981 | 0.010 | 0.000 | 9.88 | 95 | 89 | 8.77 | 0.05 | |
|
| 0.988 | 0.007 | 0.000 | 11.13 | 93 | 96 | 21.40 | 0.07 | |
|
|
| 0.930 | 0.026 | 0.000 | 7.76 | 79 | 92 | 9.75 | 0.23 |
|
| 0.931 | 0.027 | 0.000 | 8.96 | 81 | 95 | 15.06 | 0.20 | |
|
| 0.919 | 0.031 | 0.000 | 15.64 | 81 | 95 | 15.06 | 0.20 | |
|
| 0.926 | 0.030 | 0.000 | 16.44 | 86 | 95 | 15.92 | 0.15 |
AUROC, area under the receiver operating characteristic curve; CI, confidence interval; KC, keratoconus; BFS, best-fit sphere; BFT, best-fit toric-ellipsoid; SE, standard error; Se, sensitivity; Sp, specificity; PLR, positive likelihood ratio; NLR, negative likelihood ratio; n.a, not applicable
Cut-off point, specificity, and sensitivity values identified in previous study.
| Study | Device | Reference surface | Elevation values(μm) | Normal control | FFKC or KC suspect | KC | Cut-off value (μm) | Sensitivity | Specificity |
|---|---|---|---|---|---|---|---|---|---|
| SS-OCT CASIA 1000 | BFS (central 5 mm of the anterior or posterior cornea) | anterior | 5.43 ± 1.38 (n = 88) | 13.85 ± 7.23 (n = 13) | 44.00 ± 22.15 (n = 29) | 8.00 | 85% | 96% | |
| posterior | 10.00 ± 2.82 | 22.69 ± 12.07 | 89.93 ± 41.93 | 13.00 | 85% | 86% | |||
| Orbscan II | BFS | anterior | 0.005 ± 0.002 | 0.009 ± 0.007 | 0.013 ± 0.010 | - | - | - | |
| posterior | 0.021 ± 0.006 | 0.035 ± 0.015 | 0.044 ±0.025 | 40 | - | - | |||
| Pentacam | BFS (central 5 mm of the anterior or posterior cornea) | anterior | 2.7 ± 2.3 | - | 33.3 ± 28.6 | - | - | - | |
| posterior | 5.7 ± 5.5 | 55.8 ± 33.2 | 15.5 | 95.1% | 94.3% | ||||
| Pentacam | BFS, BFTE (fixed 8-mm-diameter) | Posterior elevation BFS 8 mm | 1.38 ± 5.75 | - | 42.68 ± 33.44 | 15.5 | 91% | 98% | |
| Posterior elevation BFTE 8 mm | 5.66 ± 5.02 | 67.57 ± 48.57 | 10.5 | 91% | 95% | ||||
| Pentacam | BFS (8-mm-diameter) | Anterior | 3.59 ± 2.43 | - | 16.28 ± 12.14 | 8.5 | 71.6% | 100% | |
| at thinnest point | Anterior elevation at thinnest point | 2.29 ± 1.80 | 20.53 ± 14.33 | 5.5 | 91.3% | 97.4% | |||
| BFS (8-mm-diameter) | Posterior | 5.64 ± 3.50 | 35.74 ± 25.95 | 12.5 | 87.3% | 96.1% | |||
| at thinnest point | Posterior elevation at thinnest point | 6.41 ± 3.83 | 47.39 ± 28.52 | 13.5 | 93.2% | 94.9% | |||
| Sirius | Elevation values at Thinnest corneal point | Spherical MAE | 5.2 ± 1.9 | - | 21.2 ± 11.3 | 24 | 91% | 92% | |
| Aspherotoric MAE | 5.6 ± 2.4 | 23.9 ± 17.1 | 19 | 98% | 99% | ||||
| Spherical MPE | 6.3 ± 2.0 | 28.9 ± 18.1 | 15 | 92% | 93% | ||||
| Aspherotoric MPE | 6.7 ± 2.3 | 29.8 ± 17 | 12 | 98% | 99% | ||||
| GALILEI | BFS and BFTA (8-mm-diameter) | BFS MAE | 5.4 ± 3.15 | 7.2 ± 4.3 | 25.2 ± 12.5 | 11 | 92% | 91% | |
| BFTA MAE | 4.8 ± 1.7 | 8.4 ± 3.9 | 31.9 ± 15.8 | 9 | 98% | 99% | |||
| BFS MPE | 13.1 ± 5.2 | 15.4 ± 6.5 | 46.0 ± 20.6 | 21 | 93% | 95% | |||
| BFTA MPE | 8.6 ± 2.8 | 16.9 ± 6.9 | 57.8 ± 28.4 | 16 | 99% | 99% |
FFKC, Forme-Fruste Keratoconus; KC, keratoconus; BFS, best-fit sphere; BFTA, best-fit toric and aspherical surface; BFTE, best-fit toric-ellipsoid; MAE, mean anterior elevation; MPE, mean posterior elevation