Multicentre study: reliability and repeatability of Scheimpflug system measurement in keratoconus.

Beatriz de Luis Eguileor1, Pedro Arriola-Villalobos2, Jose Ignacio Pijoan Zubizarreta3, Raquel Feijoo Lera3, Alaitz Santamaria Carro3, David Diaz-Valle2, Jaime Etxebarria3.   

Abstract

PURPOSE: To assess the repeatability and reliability of the most important tomographic parameters for characterising keratoconus measured with a Pentacam HR (high resolution).
METHODS: Overall, 230 eyes in 158 patients with keratoconus were analysed. We performed five consecutive corneal tomography examinations for each eye with a Pentacam HR in patients with keratoconus. Study eyes were classified into three groups depending on the maximum posterior elevation (max_BFS_post): grade 1 for cases of keratoconus with a max_BFS_post of 40 µm; grade 2 for those with a max_BFS_post of between 41 and 75 µm and grade 3 for those with a max_BFS_post of over 75 µm. We calculated the intraclass correlation coefficients (ICCs) and repeatability limits of parameters from tomography and aberrometry.
RESULTS: All the parameters were found to have excellent ICCs (0.9). The repeatability limits for the key parameters were higher than 0.5D for the power parameters, 20° for the axis of corneal astigmatism and 10 µm for the thinnest corneal thickness. Further, we obtained repeatability limits of above 0.1 µm for the aberrometry values and overall greater than 15° for the coma axis. All the values increase with the severity of keratoconus, except for that of the coma axis which falls with keratoconus grade.
CONCLUSIONS: The reliability indicated by ICCs supports the view that the Pentacam HR is useful for the diagnosis of keratoconus. The repeatability limits suggest that new criteria should be established for monitoring progression taking into account the real measurements that can be made using this system. © Author(s) (or their employer(s)) 2021. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Keywords:  cornea; ocular surface

Year:  2020        PMID: 32217537     DOI: 10.1136/bjophthalmol-2019-314954

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  5 in total

1.  A novel analysis of Scheimpflug total corneal refractive power following corneal cross-linking in mild to moderate keratoconus.

Authors:  Chao Pan; Wei-Na Tan; Dan Chen; Yu Liu; Hao-Yu Wang; Deng-Feng Liang; Yan-Jun Hua; Xiao-Hua Lei; Qing-Yan Zeng; Shao-Zhen Zhao
Journal:  Int J Ophthalmol       Date:  2022-05-18       Impact factor: 1.779

Review 2.  [Treatment indications for corneal crosslinking and clinical results of new corneal crosslinking techniques].

Authors:  Klara Borgardts; Johannes Menzel-Severing; Gerd Geerling; Theo G Seiler
Journal:  Ophthalmologe       Date:  2022-02-11       Impact factor: 1.059

3.  Fourier analysis of corneal Scheimpflug imaging: clinical use in keratoconus.

Authors:  Dorukcan Akincioglu; Gokhan Ozge; Onder Ayyildiz; Gokcen Gokce; Umut Karaca; Fatih Mehmet Mutlu
Journal:  Int J Ophthalmol       Date:  2021-12-18       Impact factor: 1.779

4.  Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis.

Authors:  Robert Herber; Janine Lenk; Lutz E Pillunat; Frederik Raiskup
Journal:  Eye Vis (Lond)       Date:  2022-05-23

Review 5.  Spotlight on the Corneal Back Surface Astigmatism: A Review.

Authors:  Sudi Patel; Larysa Tutchenko
Journal:  Clin Ophthalmol       Date:  2021-07-26
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.