Tsuneari Takahashi1,2, Siddharth Thaker3, Giovanni Lettieri1,3, Anthony Redmond1,3,4, Michael R Backhouse1,5, Martin Stone3,4, Hemant Pandit1,3,4, Philip O'Connor3,4.
Abstract
OBJECTIVES: To validate reliability of slice-encoding for metal artefact correction (SEMAC)-MRI findings in prosthesis loosening detection by comparing them to surgical outcomes (gold standard) in symptomatic patients following hip arthroplasties. To evaluate periprosthetic anatomical structures in symptomatic patients to identify an alternative cause of hip symptoms.Entities:
Mesh:
Substances:
Year: 2022 PMID: 35148205 PMCID: PMC9153704 DOI: 10.1259/bjr.20210940
Source DB: PubMed Journal: Br J Radiol ISSN: 0007-1285 Impact factor: 3.629
MRI parameters
| Sequence | T1W axial | T2W axial | T2W sagittal | T1W coronal | PDW coronal | STIR coronal | T1W-SEMAC coronal | PDW-SEMAC coronal | STIR-SEMAC coronal |
|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
| - | - | - | - | - |
| - | - |
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
(TR – Repetition time; TE – Echo time; TI – Time to invert; ETL – echo train length; FOV – field of view; NSA – Number of signal average; ms – miliseconds; mm – milimeters; and Hz/Px – Hertz/Pixel).
Figure 1.68-year-old male with left hip pain following THA. (a). T1 SEMAC sequence - absent cancellous markings in Gruen zone 1 and 2 suggesting osteolysis (yellow arrows), (b). STIR-SEMAC sequence – Gruen zone 1 and 2 indicates the fluid infiltration around the femoral component (yellow arrows) and increased signal at zone four around the femoral stem tip not following fluid signal is consistent with the BME (curved amber arrows), (c). PDW-SEMAC sequence. Fluid signal corresponding to area of osteolysis (yellow arrows) – please note the sensitivity of STIR-SEMAC as compared to PDW-SEMAC.
Figure 2.76-year-old female with right hip pain following THA. T1W-SEMAC image demonstrating ‘Delee and Charnley zones’ for acetabular component and Gruen’s zone for femoral components.
Figure 3.80-year-old male with right hip pain following THA. (a). T1-weighted image with SEMAC and (b). T1W without SEMAC acquisition parameters demonstrating effective metal artefact suppression on T1w sequences.
Visualisation of interface between bone and cement/prosthesis (coronal images)
| STIR CONV | STIR SEMAC | T1 CONV | T1 SEMAC | PD CONV | PD SEMAC | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Acetabulum | Zone.1 | 0.64 (1.3) | 2.0 (1.1) | 0.0047 | 0.73 (1.1) | 2.4 (1.2) | 0.0053 | 0.75 (1.5) | 2.3 (0.5) | NS |
| Zone.2 | 0.73 (1.3) | 2.2 (1.1) | 0.0048 | 0.73 (1.1) | 2.6 (1.2) | 0.0057 | 0.75 (1.5) | 2.8 (0.5) | NS | |
| Zone.3 | 0.73 (1.3) | 2.3 (1.1) | 0.0048 | 0.73 (1.1) | 2.6 (1.2) | 0.0057 | 0.75 (1.5) | 2.8 (0.5) | NS | |
| Femur | Zone.1 | 1.6 (1.5) | 3.2 (0.8) | 0.013 | 3.3 (0.9) | 3.7 (0.9) | 0.018 | 3.8 (0.5) | 4.0 (0.0) | NS |
| Zone.2 | 2.5 (1.1) | 3.4 (0.5) | 0.011 | 3.5 (0.9) | 3.8 (0.4) | NS | 3.8 (0.5) | 3.8 (0.5) | NS | |
| Zone.3 | 2.5 (1.4) | 3.5 (0.5) | 0.0044 | 3.5 (0.9) | 3.8 (0.4) | NS | 3.8 (0.5) | 3.8 (0.5) | NS | |
| Zone.4 | 2.6 (1.4) | 3.5 (0.5) | 0.050 | 3.7 (0.5) | 3.9 (0.3) | NS | 3.7 (0.5) | 3.9 (0.3) | NS | |
| Zone.5 | 2.7 (1.4) | 3.5 (0.5) | NS | 3.8 (0.4) | 3.9 (0.3) | NS | 3.8 (0.5) | 3.8 (0.5) | NS | |
| Zone.6 | 2.8 (1.2) | 3.5 (0.5) | 0.033 | 3.8 (0.4) | 3.9 (0.3) | NS | 3.5 (1.0) | 4.0 (0.0) | NS | |
| Zone.7 | 1.8 (1.4) | 3.2 (0.6) | 0.0077 | 3.1 (0.7) | 3.8 (0.5) | 0.019 | 2.8 (1.9) | 4.0 (0.0) | NS |
CONV, conventional; NS, not significant.
Paired T test was used, and data are expressed as mean (standard deviation).
Abnormal findings in cases with and without prosthesis loosening (coronal images)
| STIR SEMAC | T1 SEMAC | PD SEMAC | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Stable | Loosening | Stable | Loosening | Stable | Loosening | |||||
| Acetabulum | Zone.1 | 0.071 (0.26) | 0.09 (0.30) | NS | 0.071 (0.26) | 0.09 (0.30) | NS | 0.0 (0.0) | 0.0 (0.0) | NS |
| Zone.2 | 0.11 (0.31) | 0.09 (0.30) | NS | 0.071 (0.26) | 0.09 (0.30) | NS | 0.0 (0.0) | 0.0 (0.0) | NS | |
| Zone.3 | 0.18 (0.39) | 0.27 (0.47) | NS | 0.071 (0.26) | 0.18 (0.40) | NS | 0.071 (0.26) | 0.09 (0.30) | NS | |
| Femur | Zone.1 | 0.25 (0.44) | 0.73 (0.47) | 0.0049 | 0.18 (0.39) | 0.18 (0.40) | NS | 0.0 (0.0) | 0.0 (0.0) | NS |
| Zone.2 | 0.14 (0.36) | 0.09 (0.30) | NS | 0.071 (0.26) | 0.09 (0.30) | NS | 0.0 (0.0) | 0.0 (0.0) | NS | |
| Zone.3 | 0.071 (0.26) | 0.09 (0.30) | NS | 0.0 (0.0) | 0.09 (0.30) | NS | 0.071 (0.26) | 0.09 (0.30) | NS | |
| Zone.4 | 0.18 (0.39) | 0.18 (0.40) | NS | 0.0 (0.0) | 0.09 (0.30) | NS | 0.0 (0.0) | 0.0 (0.0) | NS | |
| Zone.5 | 0.25 (0.97) | 0.0 (0.0) | NS | 0.0 (0.0) | 0.09 (0.30) | NS | 0.0 (0.0) | 0.0 (0.0) | NS | |
| Zone.6 | 0.11 (0.31) | 0.09 (0.30) | NS | 0.036 (0.19) | 0.0 (0.0) | NS | 0.071 (0.26) | 0.09 (0.30) | NS | |
| Zone.7 | 0.11 (0.31) | 0.55 (0.52) | 0.0027 | 0.11 (0.31) | 0.36 (0.50) | NS | 0.0 (0.0) | 0.0 (0.0) | NS | |
NS, not significant.
Student’s T test was used, and data are expressed as median (standard deviation).
Visualisation of anatomic structures
| STIR | STIR | T1 | T1 | PD CONV | PD SEMAC | ||||
|---|---|---|---|---|---|---|---|---|---|
| Psoas tendon insertion | 2.3 (1.6) | 3.4 (0.5) | NS | 3.8 (0.4) | 3.9 (0.3) | NS | 3.8 (0.5) | 4.0 (0.0) | NS |
| Gluteus medius | 3.1 (0.8) | 3.9 (0.3) | 0.011 | 3.7 (0.6) | 3.9 (0.3) | NS | 3.5 (0.6) | 3.8 (0.5) | NS |
| Gluteus minimus | 1.5 (1.4) | 3.1 (1.2) | 0.0077 | 3.0 (1.5) | 3.3 (1.2) | NS | 1.8 (1.7) | 3.0 (0.8) | NS |
| Obturator intenus | 1.8 (1.3) | 3.6 (0.7) | 0.0048 | 3.6 (0.5) | 3.8 (0.4) | NS | 2.3 (1.7) | 3.8 (0.5) | NS |
| Obturator externus | 2.1 (1.0) | 3.3 (0.7) | 0.012 | 3.7 (0.5) | 3.8 (0.4) | NS | 3.0 (0.8) | 3.3 (1.0) | NS |
CONV, conventional; NS, not significant.
Paired T test was used, and data are expressed as mean (standard deviation).
Figure 4.Patient flow during the study
Sensitivity, Specificity, PPV and NPV of SEMAC-MRI to detect loosening of the prosthesis
| Sensitivity | Specificity | PPV | NPV | |
|---|---|---|---|---|
| T1-SEMAC | 72.7 | 64.3 | 44.4 | 85.7 |
| STIR-SEMAC | 90.9 | 46.4 | 40.0 | 92.9 |
| PD-SEMAC | 36.3 | 78.5 | 40.0 | 75.8 |
NPV: Negative predictive value;PPV: Positive predictive value.