| Literature DB >> 15809957 |
Thomas A Gruen1, Robert A Poggie, David G Lewallen, Arlen D Hanssen, Randall J Lewis, Thomas J O'Keefe, S David Stulberg, Charles J Sutherland.
Abstract
Serial radiographs of a porous tantalum monoblock acetabular cup design were evaluated for cup stability and signs of successful osteointegration. Of 574 primary consecutive total hip replacements in 542 patients performed by 9 surgeons at 7 hospitals, 414 cases were available for minimum 2-year follow-up. Follow-up averaged 33 months and ranged from 24 to 58 months. Postoperative radiographs revealed acetabular gaps in 100 zones in 80 (19%) hips: 29 in zone I, 67 in zone II, and 4 in zone III. At last follow-up, 84 (84%) of the zones with gaps completely filled in, and all 4- and 5-mm gaps filled in. There was no progression of any postoperative gap, no evidence of continuous periacetabular interface radiolucencies, no evidence of lysis, and no revisions for loosening. Although these short-term results are encouraging, further follow-up will be required to assess whether the monoblock design and the low modulus of elasticity of porous tantalum will reduce the incidence of periacetabular stress shielding and occurrence of osteolysis.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15809957 DOI: 10.1016/j.arth.2004.12.049
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757