SEMAC: Slice Encoding for Metal Artifact Correction in MRI.

Wenmiao Lu1, Kim Butts Pauly, Garry E Gold, John M Pauly, Brian A Hargreaves.   

Abstract

Magnetic resonance imaging (MRI) near metallic implants remains an unmet need because of severe artifacts, which mainly stem from large metal-induced field inhomogeneities. This work addresses MRI near metallic implants with an innovative imaging technique called "Slice Encoding for Metal Artifact Correction" (SEMAC). The SEMAC technique corrects metal artifacts via robust encoding of each excited slice against metal-induced field inhomogeneities. The robust slice encoding is achieved by extending a view-angle-tilting (VAT) spin-echo sequence with additional z-phase encoding. Although the VAT compensation gradient suppresses most in-plane distortions, the z-phase encoding fully resolves distorted excitation profiles that cause through-plane distortions. By positioning all spins in a region-of-interest to their actual spatial locations, the through-plane distortions can be corrected by summing up the resolved spins in each voxel. The SEMAC technique does not require additional hardware and can be deployed to the large installed base of whole-body MRI systems. The efficacy of the SEMAC technique in eliminating metal-induced distortions with feasible scan times is validated in phantom and in vivo spine and knee studies. (c) 2009 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19267347      PMCID: PMC2837371          DOI: 10.1002/mrm.21967

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  21 in total

1.  Composite image formation in z-shimmed functional MR imaging.

Authors:  R T Constable; D D Spencer
Journal:  Magn Reson Med       Date:  1999-07       Impact factor: 4.668

2.  Generalized autocalibrating partially parallel acquisitions (GRAPPA).

Authors:  Mark A Griswold; Peter M Jakob; Robin M Heidemann; Mathias Nittka; Vladimir Jellus; Jianmin Wang; Berthold Kiefer; Axel Haase
Journal:  Magn Reson Med       Date:  2002-06       Impact factor: 4.668

3.  MRI of hip prostheses using single-point methods: in vitro studies towards the artifact-free imaging of individuals with metal implants.

Authors:  P Ramos-Cabrer; J P M van Duynhoven; A Van der Toorn; K Nicolay
Journal:  Magn Reson Imaging       Date:  2004-10       Impact factor: 2.546

4.  Correction of MR image distortions induced by metallic objects using a 3D cubic B-spline basis set: application to stereotactic surgical planning.

Authors:  S Skare; J L R Andersson
Journal:  Magn Reson Med       Date:  2005-07       Impact factor: 4.668

5.  Fast spin echo sequences with very long echo trains: design of variable refocusing flip angle schedules and generation of clinical T2 contrast.

Authors:  Reed F Busse; Hari Hariharan; Anthony Vu; Jean H Brittain
Journal:  Magn Reson Med       Date:  2006-05       Impact factor: 4.668

6.  Sparse MRI: The application of compressed sensing for rapid MR imaging.

Authors:  Michael Lustig; David Donoho; John M Pauly
Journal:  Magn Reson Med       Date:  2007-12       Impact factor: 4.668

7.  Nonsusceptibility artifacts due to metallic objects in MR imaging.

Authors:  C R Camacho; D B Plewes; R M Henkelman
Journal:  J Magn Reson Imaging       Date:  1995 Jan-Feb       Impact factor: 4.813

8.  Total inhomogeneity correction including chemical shifts and susceptibility by view angle tilting.

Authors:  Z H Cho; D J Kim; Y K Kim
Journal:  Med Phys       Date:  1988 Jan-Feb       Impact factor: 4.071

9.  Prevalence of primary and revision total hip and knee arthroplasty in the United States from 1990 through 2002.

Authors:  Steven Kurtz; Fionna Mowat; Kevin Ong; Nathan Chan; Edmund Lau; Michael Halpern
Journal:  J Bone Joint Surg Am       Date:  2005-07       Impact factor: 5.284

10.  Susceptibility artefacts in NMR imaging.

Authors:  K M Lüdeke; P Röschmann; R Tischler
Journal:  Magn Reson Imaging       Date:  1985       Impact factor: 2.546

View more
  123 in total

1.  Modeling of active shimming of metallic needles for interventional MRI.

Authors:  Saikat Sengupta
Journal:  Magn Reson Med       Date:  2020-06-29       Impact factor: 4.668

2.  Advanced MRI of articular cartilage.

Authors:  Hillary J Braun; Garry E Gold
Journal:  Imaging Med       Date:  2011-10

3.  Review. The Agfa Mayneord lecture: MRI of short and ultrashort T₂ and T₂* components of tissues, fluids and materials using clinical systems.

Authors:  G M Bydder
Journal:  Br J Radiol       Date:  2011-12       Impact factor: 3.039

4.  Usefulness of IDEAL T2-weighted FSE and SPGR imaging in reducing metallic artifacts in the postoperative ankles with metallic hardware.

Authors:  Jung Bin Lee; Jang Gyu Cha; Min Hee Lee; Young Koo Lee; Eun Hye Lee; Chan Hong Jeon
Journal:  Skeletal Radiol       Date:  2012-06-06       Impact factor: 2.199

5.  Real-time MR navigation and localization of an intravascular catheter with ferromagnetic components.

Authors:  Ke Zhang; Axel Joachim Krafft; Reiner Umathum; Florian Maier; Wolfhard Semmler; Michael Bock
Journal:  MAGMA       Date:  2010-05-21       Impact factor: 2.310

6.  Metal artifact reduction MRI of total ankle arthroplasty implants.

Authors:  Cesar de Cesar Netto; Lucas F Fonseca; Benjamin Fritz; Steven E Stern; Esther Raithel; Mathias Nittka; Lew C Schon; Jan Fritz
Journal:  Eur Radiol       Date:  2017-12-07       Impact factor: 5.315

Review 7.  MRI-only treatment planning: benefits and challenges.

Authors:  Amir M Owrangi; Peter B Greer; Carri K Glide-Hurst
Journal:  Phys Med Biol       Date:  2018-02-26       Impact factor: 3.609

Review 8.  Advanced MRI strategies for assessing spinal cord injury.

Authors:  Seth A Smith; James J Pekar; Peter C M van Zijl
Journal:  Handb Clin Neurol       Date:  2012

9.  Accelerating sequences in the presence of metal by exploiting the spatial distribution of off-resonance.

Authors:  Matthew R Smith; Nathan S Artz; Kevin M Koch; Alexey Samsonov; Scott B Reeder
Journal:  Magn Reson Med       Date:  2014-01-15       Impact factor: 4.668

10.  Characterizing the limits of MRI near metallic prostheses.

Authors:  Matthew R Smith; Nathan S Artz; Curtis Wiens; Diego Hernando; Scott B Reeder
Journal:  Magn Reson Med       Date:  2014-12-05       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.