Literature DB >> 8284417

Metallic artifacts on MR images of the postoperative spine: reduction with fast spin-echo techniques.

L M Tartaglino1, A E Flanders, S Vinitski, D P Friedman.   

Abstract

PURPOSE: To determine whether the relative insensitivity of T2-weighted fast spin-echo (FSE) techniques to magnetic susceptibility can be exploited to reduce metallic artifacts on images of the postoperative spine and, thus, improve the interpretation of the postoperative study.
MATERIALS AND METHODS: Three neuroradiologists retrospectively evaluated sagittal T2-weighted conventional spin-echo and FSE images obtained in 15 patients with metallic artifacts from various sources including drill particles from anterior cervical diskectomy, posterior fixation wires, fixation rods or plates, and an inferior vena cava filter. The amount of artifact present and whether these artifacts affected image interpretation were evaluated.
RESULTS: Among the 45 paired evaluations, the artifact was judged to be less apparent with FSE sequences in 39. In eight of 45 evaluations (18%), the interpretation of the area of interest was possible only on the FSE images.
CONCLUSION: FSE imaging, especially when performed with shorter echo spacing, increases the amount of T2-weighted information in the presence of metallic artifact because it decreases magnetic susceptibility effects.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8284417     DOI: 10.1148/radiology.190.2.8284417

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  17 in total

1.  Fully refocused multi-shot spatiotemporally encoded MRI: robust imaging in the presence of metallic implants.

Authors:  Noam Ben-Eliezer; Eddy Solomon; Elad Harel; Nava Nevo; Lucio Frydman
Journal:  MAGMA       Date:  2012-06-29       Impact factor: 2.310

Review 2.  Medical imaging in the diagnosis and management of cancer pain.

Authors:  Carlos Cuevas; Dean Shibata
Journal:  Curr Pain Headache Rep       Date:  2009-08

3.  PETRA, MSVAT-SPACE and SEMAC sequences for metal artefact reduction in dental MR imaging.

Authors:  Tim Hilgenfeld; Marcel Prager; Alexander Heil; Franz Sebastian Schwindling; Mathias Nittka; David Grodzki; Peter Rammelsberg; Martin Bendszus; Sabine Heiland
Journal:  Eur Radiol       Date:  2017-07-11       Impact factor: 5.315

4.  Influence of dental materials on dental MRI.

Authors:  O Tymofiyeva; S Vaegler; K Rottner; J Boldt; A J Hopfgartner; P C Proff; E J Richter; P M Jakob
Journal:  Dentomaxillofac Radiol       Date:  2013-04-22       Impact factor: 2.419

5.  Contrast on T2-weighted images of the lumbar spine using fast spin-echo and gated conventional spin-echo sequences.

Authors:  P M Chappell; G H Glover; D R Enzmann
Journal:  Neuroradiology       Date:  1995-04       Impact factor: 2.804

6.  Artifacts in magnetic resonance imaging of the spine after surgery with or without implant.

Authors:  H Leclet
Journal:  Eur Spine J       Date:  1994       Impact factor: 3.134

7.  Reduced field of view echo-planar imaging diffusion tensor MRI for pediatric spinal tumors.

Authors:  Lily H Kim; Edward H Lee; Michelle Galvez; Murat Aksoy; Stefan Skare; Rafael O'Halloran; Michael S B Edwards; Samantha J Holdsworth; Kristen W Yeom
Journal:  J Neurosurg Spine       Date:  2019-07-05

8.  Spinal fusion-hardware construct: Basic concepts and imaging review.

Authors:  Mohamed Ragab Nouh
Journal:  World J Radiol       Date:  2012-05-28

Review 9.  Current applications of advanced cross-sectional imaging techniques in evaluating the painful arthroplasty.

Authors:  Carolyn M Sofka
Journal:  Skeletal Radiol       Date:  2006-12-06       Impact factor: 2.199

Review 10.  Metal-related artifacts in instrumented spine. Techniques for reducing artifacts in CT and MRI: state of the art.

Authors:  P Stradiotti; A Curti; G Castellazzi; A Zerbi
Journal:  Eur Spine J       Date:  2009-05-13       Impact factor: 3.134

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.