BACKGROUND: Many patients with cardiac implantable electronic devices (CIED) undergo magnetic resonance imaging (MRI); however, a relevant proportion have a CIED system that has not been classified as MRI-conditional because of generators and leads from different brands (mixed-brand group). The available data concerning the outcome of these mixed patients undergoing MRI is limited.Entities:
Keywords: Defibrillator; MRI-conditional; Magnetic resonance imaging; Pacemaker; Safety
Mesh:
Year: 2021 PMID: 34402991 PMCID: PMC9023390 DOI: 10.1007/s00508-021-01924-w
Source DB: PubMed Journal: Wien Klin Wochenschr ISSN: 0043-5325 Impact factor: 1.704
Baseline characteristics of all included patients, mixed-brand and MRI-conditional groups
| Parameter | All included patients | Mixed-brand group | MRI-conditional group | |
|---|---|---|---|---|
| Number of patients | 127 | 38 | 89 | N/A |
| Number of MRI procedures | 188 | 54 | 134 | N/A |
| Age (years) (interquartile range) | 73 (67–80) | 77 (73–81) | 72 (64–79) | 0.003* |
| Female gender | 34.6% | 39.5% | 32.6% | 0.542 |
| Height (cm)a (± standard deviation) | 173 ± 8 | 173 ± 6 | 173 ± 9 | 0.935 |
| Weight (kg)a (interquartile range) | 80 (73–84) | 78 (71–80) | 82 (75–93) | 0.126 |
| 0.338 | ||||
| Pacemaker | 92.1% | 97.4% | 89.9% | – |
| Transvenous ICD | 6.3% | 2.6% | 7.9% | – |
| Subcutaneous ICD | 1.6% | 0% | 2.2% | – |
| < 0.001 | ||||
| Sick sinus syndrome | 30.8% | 17.2% | 35.9% | – |
| Atrioventricular block | 27.1% | 20.7% | 29.5% | – |
| Bradycardia-tachycardia syndrome | 23.4% | 37.9% | 17.9% | – |
| Other block | 4.7% | 13.8% | 1.3% | – |
| Heart failure | 3.7% | 3.4% | 3.8% | – |
| Secondary prophylaxis | 3.7% | 3.4% | 3.8% | – |
| Other Indication | 6.5% | 3.4% | 7.7% | – |
| < 0.001 | ||||
| Biotronikb | 32.3% | 5.3% | 43.8% | – |
| Boston Scientificc | 12.6% | 10.5% | 13.5% | – |
| Medtronicd | 24.4% | 36.8% | 19.1% | – |
| Sorin/LivaNovae | 17.3% | 42.1% | 6.7% | – |
| St. Jude Medical/Abbottf | 13.4% | 5.3% | 16.9% | – |
| Months since prior box change before MRI | 26.8 (12.1–48.1) | 33.0 (11.4–48.9) | 25.4 (12.3–47.9) | 0.953 |
| Prior box change < 3 months | 2.4% | 2.6% | 2.2% | 1.000 |
| 0.002 | ||||
| Biotronikb | 54.3% | 78.9% | 43.8% | – |
| Boston Scientificc | 10.2% | 2.6% | 13.5% | – |
| Medtronicd | 18.1% | 15.8% | 19.1% | – |
| Sorin/LivaNovae | 4.7% | 0.0% | 6.7% | – |
| St. Jude Medical/Abbottf | 12.6% | 2.6% | 16.9% | – |
| Months since last lead implantation before MRI | 36.6 (15.1–66.1) | 39.2 (18.0–80.4) | 34.3 (14.2–61.6) | 0.662 |
| Last lead implantation < 3 months | 0.8% | 2.6% | 0.0% | 0.299 |
| 0.322 | ||||
| Head or extremities | 40.4% | 48.1% | 37.3% | – |
| Extrathoracic torso | 55.3% | 50.0% | 57.5% | – |
| Thorax | 4.3% | 1.9% | 5.2% | – |
| 0.681 | ||||
| 000 | 7.4% | 10.0% | 6.8% | – |
| A00/V00 | 15.7% | 10.0% | 17.0% | – |
| D00 | 76.9% | 80.0% | 76.1% | – |
| Estimated ERI before MRI (years) | 7.4 (6.0–9.3) | 7.0 (5.0–10.3) | 7.5 (6.0–9.1) | 0.831 |
CIED cardiac implantable electronic device, ICD implantable cardioverter/defibrillator, MRI magnetic resonance imaging, ERI elective replacement indicator
* p < 0.05
aincomplete data for some patients
b(Berlin, Germany)
c(Marborough, MA, USA)
d(Dublin, Ireland)
e(London, UK)
f(St. Paul, MN, USA)
Outcome of included patients, mixed-brands and MRI-conditional groups
| Parameter | All included patients (%) | Mixed-brands group (%) | MRI-conditional group (%) | |
|---|---|---|---|---|
| 1.6 | 0.0 | 2.2 | 1.000 | |
| 2.4 | 0.0 | 3.4 | 0.554 | |
| Patient discomfort | 1.6 | 0.0 | 2.2 | 1.000 |
| Increase of ventricular threshold | 0.8 | 0.0 | 1.1 | 1.000 |
| Decrease of amplitude > 50% | 4.7 | 7.9 | 3.4 | 0.363 |
| Increase of pacing threshold > 1 V | 3.1 | 0.0 | 4.5 | 0.316 |
| 1‑year rehospitalization | 37.8 | 36.8 | 38.2 | 1.000 |
CIED cardiac implantable electronic device, ICD implantable cardioverter/defibrillator, MRI magnetic resonance imaging
* p < 0.05
Fig. 1Study flowchart. CIED cardiac implantable electric device, MRI magnetic resonance imaging