Literature DB >> 31995177

Magnetic resonance imaging in non-conditional pacemakers and implantable cardioverter-defibrillators: a systematic review and meta-analysis.

Dian A Munawar1,2, Joel E Z Chan1, Mehrdad Emami1, Kadhim Kadhim1, Kashif Khokhar1, Catherine O'Shea1, Shinsuke Iwai1, Bradley Pitman1, Dominik Linz1, Muhammad Munawar2, Kurt Roberts-Thomson1, Glenn D Young1, Rajiv Mahajan1,3, Prashanthan Sanders1, Dennis H Lau1.   

Abstract

AIMS: There is growing evidence that magnetic resonance imaging (MRI) scanning in patients with non-conditional cardiac implantable electronic devices (CIEDs) can be performed safely. Here, we aim to assess the safety of MRI in patients with non-conditional CIEDs. METHODS AND
RESULTS: English scientific literature was searched using PubMed/Embase/CINAHL with keywords of 'magnetic resonance imaging', 'pacemaker', 'implantable defibrillator', and 'cardiac resynchronization therapy'. Studies assessing outcomes of adverse events or significant changes in CIED parameters after MRI scanning in patients with non-conditional CIEDs were included. References were excluded if the MRI conditionality of the CIEDs was undisclosed; number of patients enrolled was <10; or studies were case reports/series. About 35 cohort studies with a total of 5625 patients and 7196 MRI scans (0.5-3 T) in non-conditional CIEDs were included. The overall incidence of lead failure, electrical reset, arrhythmia, inappropriate pacing and symptoms related to pocket heating, or torque ranged between 0% and 1.43%. Increase in pacing lead threshold >0.5 V and impedance >50Ω was seen in 1.1% [95% confidence interval (CI) 0.7-1.8%] and 4.8% (95% CI 3.3-6.4%) respectively. The incidence of reduction in P- and R-wave sensing by >50% was 1.5% (95% CI 0.6-2.9%) and 0.4% (95% CI 0.06-1.1%), respectively. Battery voltage reduction of >0.04 V was reported in 2.2% (95% CI 0.2-6.1%).
CONCLUSION: This meta-analysis affirms the safety of MR imaging in non-conditional CIEDs with no death or implantable cardioverter-defibrillator shocks and extremely low incidence of lead or device-related complications. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author(s) 2020. For permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  Cardiac implantable electronic devices; Cardiac resynchronization therapy; Implantable cardioverter-defibrillators; Magnetic resonance imaging; Meta-analysis; Pacemakers

Mesh:

Year:  2020        PMID: 31995177     DOI: 10.1093/europace/euz343

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  3 in total

1.  Need for MRI scans in a real-world CIED population over long-term follow-up: Data from a large single-centre experience.

Authors:  Giosuè Mascioli; Elena Lucca; Federica Michelotti; Luca Tarantino; Fabrizio Giofré; Ilaria Finamora
Journal:  PLoS One       Date:  2020-12-30       Impact factor: 3.240

2.  Is diversity harmful?-Mixed-brand cardiac implantable electronic devices undergoing magnetic resonance imaging.

Authors:  Christoph Alexander König; Florian Tinhofer; Thomas Puntus; Achim Leo Burger; Nikolaus Neubauer; Herbert Langenberger; Kurt Huber; Michael Nürnberg; David Zweiker
Journal:  Wien Klin Wochenschr       Date:  2021-08-17       Impact factor: 1.704

3.  2021 PACES expert consensus statement on the indications and management of cardiovascular implantable electronic devices in pediatric patients.

Authors:  Maully J Shah; Michael J Silka; Jennifer N Avari Silva; Seshadri Balaji; Cheyenne M Beach; Monica N Benjamin; Charles I Berul; Bryan Cannon; Frank Cecchin; Mitchell I Cohen; Aarti S Dalal; Brynn E Dechert; Anne Foster; Roman Gebauer; M Cecilia Gonzalez Corcia; Prince J Kannankeril; Peter P Karpawich; Jeffery J Kim; Mani Ram Krishna; Peter Kubuš; Martin J LaPage; Douglas Y Mah; Lindsey Malloy-Walton; Aya Miyazaki; Kara S Motonaga; Mary C Niu; Melissa Olen; Thomas Paul; Eric Rosenthal; Elizabeth V Saarel; Massimo Stefano Silvetti; Elizabeth A Stephenson; Reina B Tan; John Triedman; Nicholas H Von Bergen; Philip L Wackel
Journal:  Indian Pacing Electrophysiol J       Date:  2021-07-29
  3 in total