| Literature DB >> 34277097 |
Pushpraj K Pathak1, S K Vashisht1, S Baby1, P K Jithin1, Y Jain2, R Mahawar2, V G G K Sharan2.
Abstract
BACKGROUND: Varian Medical Systems has introduced a new medical linear accelerator called HalcyonTM 2.0, which is based on the ring delivery system (RDS). It is a true IGRT machine having 6MV FFF photon energy. In addition to the planar and MV-CBCT imaging techniques it also has an option of ultra-fast kV-iCBCT which enhances the image reconstruction and improves the visualization of soft tissue. The field portals are shaped by a unique dual layer MLC with special stacked and staggered design which enables high modulation with low radiation leakage. Recently, we have commissioned our first Halcyon 2.0 machine. The aim of this work was to systematically investigate various parameters of a newly installed HalcyonTM 2.0 linear accelerator.Entities:
Keywords: commissioning; dose; halcyon; radiotherapy; tissue
Year: 2021 PMID: 34277097 PMCID: PMC8281907 DOI: 10.5603/RPOR.a2021.0065
Source DB: PubMed Journal: Rep Pract Oncol Radiother ISSN: 1507-1367
Mechanical tests parameters of couch
| Sr. no. | Tests | Tolerance | Observations |
|---|---|---|---|
| 1. | Minimum level of table top above the floor | ≤ 80 cm | 62.5 cm |
| 2. | Longitudinal motion of the couch | ≥ 70 cm | 165.5 cm |
| 3. | Vertical motion of couch from the isocenter | > 40 cm below the isocenter | 47.4 cm below the isocenter |
| 4. | Minimum linear speed of table top | ≤ 1 cm/sec | 0.43 cm/sec |
| 5. | Accuracy of longitudinal motion of table top | 0.3 cm for speed ≤ 2.5 cm/sec | < 0.1 cm |
| 6. | Accuracy of lateral motion of the table top | 0.3 cm for speed ≤ 2.5 cm/sec | < 0.1 cm |
| 7. | Accuracy of vertical motion of table top | 0.3 cm for speed ≤ 2.5 cm/sec | < 0.1 cm |
| 8. | Table top sag at isocenter when loaded with 135 kg distributed over 2 cm through isocenter | ≤ 2 mm | ≤ 2 mm |
| 9. | Couch transmission | 0.9798 | 0.9796 |
Mechanical tests parameters of multi leaves collimator (MLC)
| Sr. no. | Tests | Tolerance | Observations |
|---|---|---|---|
| 1. | Accuracy of angular scale of collimator | 0.5° at ≤ 1°/sec | 0.1° |
| 2. | Gap between isocenter and end of accessory mount | ≥ 30 cm | 50 cm |
| 3. | Shift in isocenter due to collimator rotation | ≤ 2 mm in dia | 0.6 mm |
| 4. | Accuracy of leaf position | ± 1 mm | 0.11 mm |
| 5. | Reproducibility of leaf position | ± 1 mm | 0.58 mm |
Figure 1Output factor of Halcyon machine
Consistency of TPR20/10 at different time in a day
| Energy | Time of measurements | TPR 20/10 | Difference (%) | Average | Standard deviation | Coefficient of variation (%) |
|---|---|---|---|---|---|---|
| 6MV FFF | T1 (10 AM) | 0.6267 | Ref | 0.6247 | 0.001609348 | 0.257619288 |
| T2 (12 Noon) | 0.6245 | −0.35 | ||||
| T3 (3 PM) | 0.6232 | −0.56 | ||||
| T4 (6 PM) | 0.6264 | −0.05 |
Output constancy at different time in a day for 300 MU and 10 × 10 cm2 field size
| Energy | Time of measurements | MR (nC) | Output [cGy/MU] | Difference (%) | Average | Standard deviation | Coefficient of variation (%) |
|---|---|---|---|---|---|---|---|
| 6MV FFF | T1 (10 AM) | 34.803 | 1.0007 | Ref | 1.0032 | 0.000876 | 0.08728 |
| T2 (12 Noon) | 34.87 | 1.0026 | 0.19 | ||||
| T3 (3 PM) | 34.833 | 1.0015 | 0.14 | ||||
| T4 (6 PM) | 34.81 | 1.0008 | 0.01 |
Figure 2Output constancy check of Halcyon machine
Consistency of PDD at 10 cm depth for 5 × 5 and 10 × 10 cm2 field sizes
| Field size | Energy | Sr. no. | PDD at 10cm depth | Average | Standard deviation | Coefficient of variation (%) |
|---|---|---|---|---|---|---|
| 5 × 5 | 6 FFF | 1 | 58.71 | 58.7325 | 0.035939764 | 0.061192295 |
| 2 | 58.74 | |||||
| 3 | 58.78 | |||||
| 4 | 58.7 | |||||
| 10 × 10 | 6 FFF | 1 | 62.79 | 62.8 | 0.060553007 | 0.096421986 |
| 2 | 62.86 | |||||
| 3 | 62.83 | |||||
| 4 | 62.72 |
MU linearity tests
| A. For > 4 MU | |||
|---|---|---|---|
| Sr. no. | MU (U) | Mtr.Reading (nC) (L) | Factor S (= L/U) |
| 1 | 5 | 0.597 | 0.1194 |
| 2 | 6 | 0.714 | 0.1190 |
| 3 | 7 | 0.833 | 0.1190 |
| 4 | 8 | 0.948 | 0.1185 |
| 5 | 9 | 1.071 | 0.1190 |
| 6 | 10 | 1.187 | 0.1187 |
| 7 | 15 | 1.777 | 0.1184 |
| 8 | 20 | 2.369 | 0.1184 |
| 9 | 30 | 3.55 | 0.1183 |
| 10 | 50 | 5.917 | 0.1183 |
| 11 | 70 | 8.287 | 0.1184 |
| 12 | 100 | 11.83 | 0.1183 |
| 13 | 150 | 17.74 | 0.1183 |
| 14 | 200 | 23.66 | 0.1183 |
| S Max | 0.1194 | ||
| S min | 0.1183 | ||
| Coefficient of linearity (%) | 0.4627 | ||
Figure 3PDD curve at various field sizes of Halcyon 6 MV-FFF beam
Figure 4Measured versus TPS comparison of PDD curve (Lt) and profile (Rt) for 10 × 10 cm2 field size
Figure 5In-line and Cross-line profiles at various field sizes
Consistency of symmetry for 3 × 3 cm at 10 cm depth
| Measurement plane | Energy | Sr. no. | Symmetry (%) | Difference (%) | Tolerance (%) |
|---|---|---|---|---|---|
| In-plane | 6FFF | 1 (baseline) | 100.7 | 1% | |
| 2 | 100.97 | 0.27 | |||
| 3 | 101.17 | 0.47 | |||
| 4 | 100.31 | 0.39 | |||
| Cross-plane | 6FFF | 1 (baseline) | 100.4 | 1% | |
| 2 | 100.49 | 0.09 | |||
| 3 | 100.41 | 0.01 | |||
| 4 | 101.26 | 0.86 |
Symmetry and Penumbra analysis of various field sizes
| No. | Photon beam energy | Field Size | Symmetry | Penumbra |
|---|---|---|---|---|
| 1 | 6 MV FFF | 5 × 5 cm | Inline/Crossline: 0.64%/0.36% | Inline: −4.5 mm, +4.6 mm |
| 2 | 6 MV FFF | 10 × 10 cm | Inline/Crossline: 0.26%/0.22% | Inline: −4.3 mm, +4.1 mm |
| 3 | 6 MV FFF | 20 × 20 cm | Inline/Crossline: 0.34%/0.35% | Inline: −4.9 mm, +4.8 mm |
| 4 | 6 MV FFF | 28 × 28 cm | Inline/Crossline: 0.64%/0.52% | Inline: −5.5 mm, +5.2 mm |
DMLC output at various gantry angle
| Gantry angle (degrees) | Relative output | Deviation (%) | Tolerance (%) |
|---|---|---|---|
| 0 (Ref) | 0.2903 | 0.00 | |
| 90 | 0.2927 | 0.8267 | ±3% |
| 180 | 0.2922 | 0.6545 | ±3% |
| 270 | 0.2928 | 0.8612 | ±3% |
Figure 6Picket fence test images at: A. 0°; B. 90°; C. 270°; D. 180° gantry angles; E. During rapidArc delivery (179° to 187° gantry); F. With intentional error of 0.5 and 0.2 mm at the same gantry rotation
Static picket-fence tests results
| Gantry angle (degree) | Maximum deviation [mm] | Tolerance [mm] |
|---|---|---|
| 0 (Ref) | 0.17 | 1 |
| 90 | 0.47 | 1 |
| 180 | 0.17 | 1 |
| 270 | 0.25 | 1 |
Rotational picket-fence results
| Gantry angle (degree) | Maximum deviation [mm] | Tolerance [mm] |
|---|---|---|
| 179 to 187 | 0.62 | 1 |