| Literature DB >> 26921246 |
Zhongsu Feng1, Haizhen Yue2, Yibao Zhang3, Hao Wu4, Jinsheng Cheng5, Xu Su6.
Abstract
PURPOSE: Through the Monte Carlo (MC) simulation of 6 and 10 MV flattening-filter-free (FFF) beams from Varian TrueBeam accelerator, this study aims to find the best incident electron distribution for further studying the small field characteristics of these beams.Entities:
Mesh:
Year: 2016 PMID: 26921246 PMCID: PMC4769502 DOI: 10.1186/s13014-016-0601-2
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
The beam qualities of 6 and 10 MV FFF beams
| Beam | TPR20, 10 | ||
|---|---|---|---|
| Simulated | Measured | %Diff | |
| 6 FFF | 0.628 ± 0.001 | 0.630 | −0.29 ± 0.15 |
| 10 FFF | 0.705 ± 0.001 | 0.707 | −0.28 ± 0.14 |
The simulated and measured profiles penumbras (mm) of various field sizes at 10 cm depth
| Field size (cm2) | 6 FFF | 10 FFF | ||||
|---|---|---|---|---|---|---|
| Measured | Simulated | Diff. | Measured | Simulated | Diff. | |
| 1 × 1 | 2.4 | 2.0 | 0.4 | 3.2 | 2.7 | 0.5 |
| 2 × 2 | 2.6 | 2.2 | 0.4 | 3.6 | 3.3 | 0.3 |
| 3 × 3 | 2.7 | 2.4 | 0.3 | 3.8 | 3.4 | 0.4 |
| 4 × 4 | 2.9 | 2.8 | 0.1 | 3.9 | 3.7 | 0.2 |
| 6 × 6 | 3.2 | 3.0 | 0.2 | 4.1 | 4.0 | 0.1 |
| 10 × 10 | 3.6 | 3.6 | 0.0 | 4.5 | 4.5 | 0.0 |
| 20 × 20 | 5.1 | 5.1 | 0.0 | 5.4 | 5.4 | 0.0 |
| 40 × 40 | 8.3 | 8.6 | −0.3 | 6.6 | 7.0 | −0.4 |
Fig. 1Measured and MC calculated PDDs for various field sizes (cm2) of 6 and 10 MV FFF beams. Statistical uncertainties of all dose values are between 0.1 and 0.5 % and are not shown
Fig. 2The gamma analysis results of PDDs for various field sizes (cm2) of 6 and 10 MV FFF beams. (DTA: Distance-To-Agreement, DD: Dose Difference)
The simulated and measured relative output factors of various field sizes
| Field size (cm2) | 6 FFF | 10 FFF | ||||
|---|---|---|---|---|---|---|
| Measured | Simulated | %Diff. | Measured | Simulated | %Diff. | |
| 1 × 1 | 0.684 | 0.717 | 4.82 | 0.697 | 0.721 | 3.40 |
| 2 × 2 | 0.801 | 0.807 | 0.80 | 0.845 | 0.848 | 0.34 |
| 3 × 3 | 0.842 | 0.849 | 0.82 | 0.894 | 0.895 | 0.10 |
| 4 × 4 | 0.875 | 0.873 | −0.17 | 0.921 | 0.926 | 0.52 |
| 6 × 6 | 0.929 | 0.932 | 0.25 | 0.956 | 0.962 | 0.66 |
| 10 × 10 | 1.000 | 1.000 | — | 1.000 | 1.000 | — |
| 20 × 20 | 1.080 | 1.073 | −0.66 | 1.049 | 1.045 | −0.36 |
| 40 × 40 | 1.122 | 1.117 | −0.47 | 1.069 | 1.063 | −0.58 |
Fig. 3Measured and MC calculated off-axis dose profiles for various field sizes (cm2) at 10 cm depth of 6 and 10 MV FFF beams. Statistical uncertainties of all dose values are between 0.1 and 0.5 % and are not shown
Fig. 4The gamma analysis results of OARs for various field sizes (cm2) at 10 cm depth of 6 and 10 MV FFF beams. (DTA: Distance-To-Agreement, DD: Dose Difference)
Fig. 5The comparison of measured and MC calculated off-axis dose profiles for 4 × 4 and 10 × 10 cm2 field sizes (cm2) at different depth (dmax, 5, 10, 20 and 30 cm) of 6 and 10 MV FFF beams. Statistical uncertainties of all dose values are between 0.1 and 0.5 % and are not shown