Samantha A M Lloyd1, Tze Yee Lim1, Xenia Fave1, Everardo Flores-Martinez1, Todd F Atwood1, Vitali Moiseenko1.
Abstract
Halcyon™ is a single-energy (6 MV-FFF), bore-enclosed linear accelerator. Patient setup is performed by first aligning to external lasers mounted to the front of the bore, and then loading to isocenter through pre-defined couch shifts. There is no light field, optical distance indicator or front pointer mechanism, so positioning is verified through MV imaging with kV imaging scheduled to become available in the future. TG-51 reference dosimetry was successfully performed for Halcyon™ in this imaging-based setup paradigm. The beam quality conversion factor, kQ , was determined by measuring %dd(10)x three ways: (a) using a Farmer chamber with lead filtering, (b) using a Farmer chamber without lead filtering, and (c) using a PinPoint chamber without lead filtering. Values of kQ were determined to be 0.995, 0.996, and 0.996 by each measurement technique, respectively. Halcyon™'s 6 MV-FFF beam was found to be broader than other FFF beams produced by Varian accelerators, and profile measurements at dmax showed the beam to vary less than 0.5% over the dimensions of our Farmer chamber's active volume. Reference dosimetry can be performed for the Halcyon™ accelerator simply, without specialized equipment or lead filtering with minimal dosimetric impact. This simplicity will prove advantageous in clinics with limited resources or physics support.Entities:
Keywords: Halcyon; reference dosimetry
Mesh:
Substances:
Year: 2018 PMID: 29785729 PMCID: PMC6036354 DOI: 10.1002/acm2.12349
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1TG‐51 setup on Halcyon™.
Figure 2Crossplane and inplane profiles for the Halcyon™ 6 MV‐FFF beam as measured by an IC Profiler at isocenter with 0.9 cm buildup. The active volume dimensions (width and length) of the Farmer chamber used to perform calibration measurements is indicated by the green and blue bars.
Figure 3Percent depth dose measurements performed with Farmer and PinPoint chambers. Farmer data is presented for lead filtered and unfiltered beams.
P ion and P pol for PTW 30013 Farmer chamber in Halcyon™, TrueBeam® and Clinac® beams
| Halcyon™ 6 MV‐FFF | TrueBeam® 6 MV‐FFF | TrueBeam®/Clinac® 6 MV | |
|---|---|---|---|
|
| 1.0066 | 1.0066 ± 0.0003 | 1.0031 ± 0.0004 |
|
| 0.9994 | 0.9994 ± 0.0001 | 0.9994 ± 0.0002 |
Uncertainties are standard deviations.
TrueBeam® 6 MV‐FFF data averaged over four annual calibrations.
6 MV data averaged over four matched machines (TrueBeam® and Clinac®) and four annual calibrations.
Figure 4MV images acquired at gantry angles of 90° (a) and 0° (b, c) for alignment of water tank and Farmer chamber at isocenter. Green crosshairs indicate isocenter while yellow crosshairs indicate the center of the imager.