Literature DB >> 18793960

Commissioning and quality assurance of RapidArc radiotherapy delivery system.

C Clifton Ling1, Pengpeng Zhang, Yves Archambault, Jiri Bocanek, Grace Tang, Thomas Losasso.   

Abstract

PURPOSE: The Varian RapidArc is a system for intensity-modulated radiotherapy (IMRT) treatment planning and delivery. RapidArc incorporates capabilities such as variable dose-rate, variable gantry speed, and accurate and fast dynamic multileaf collimators (DMLC), to optimize dose conformality, delivery efficiency, accuracy and reliability. We developed RapidArc system commissioning and quality assurance (QA) procedures. METHODS AND MATERIALS: Tests have been designed that evaluate RapidArc performance in a stepwise manner. First, the accuracy of DMLC position during gantry rotation is examined. Second, the ability to vary and control the dose-rate and gantry speed is evaluated. Third, the combined use of variable DMLC speed and dose-rate is studied.
RESULTS: Adapting the picket fence test for RapidArc, we compared the patterns obtained with stationary gantry and in RapidArc mode, and showed that the effect of gantry rotation on leaf accuracy was minimal (< or =0.2 mm). We then combine different dose-rates (111-600 MU/min), gantry speeds (5.5-4.3 degrees /s), and gantry range (Deltatheta = 90-12.9 degrees ) to give the same dose to seven parts of a film. When normalized to a corresponding open field (to account for flatness and asymmetry), the dose of the seven portions show good agreement, with a mean deviation of 0.7%. In assessing DMLC speed (0.46, 0.92, 1.84, and 2.76 cm/s) during RapidArc, the analysis of designed radiation pattern indicates good agreement, with a mean deviation of 0.4%.
CONCLUSIONS: The results of these tests provide strong evidence that DMLC movement, variable dose-rates and gantry speeds can be precisely controlled during RapidArc.

Entities:  

Mesh:

Year:  2008        PMID: 18793960     DOI: 10.1016/j.ijrobp.2008.05.060

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  74 in total

1.  Quality assurance of dynamic parameters in volumetric modulated arc therapy.

Authors:  A Manikandan; B Sarkar; R Holla; T R Vivek; N Sujatha
Journal:  Br J Radiol       Date:  2012-07       Impact factor: 3.039

Review 2.  PET-guided delineation of radiation therapy treatment volumes: a survey of image segmentation techniques.

Authors:  Habib Zaidi; Issam El Naqa
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-03-25       Impact factor: 9.236

3.  Dose verification for respiratory-gated volumetric modulated arc therapy.

Authors:  Jianguo Qian; Lei Xing; Wu Liu; Gary Luxton
Journal:  Phys Med Biol       Date:  2011-07-13       Impact factor: 3.609

4.  A study on dosimetric properties of electronic portal imaging device and its use as a quality assurance tool in Volumetric Modulated Arc Therapy.

Authors:  Prabakar Sukumar; Sriram Padmanaban; Prakash Jeevanandam; S A Syam Kumar; Vivekanandan Nagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2011-08-26

5.  Minimum requirements for commissioning and long-term quality assurance of Elekta multi-leaf collimator for volumetric modulated arc therapy.

Authors:  Daisaku Tatsumi; Ryosei Nakada; Akane Yomoda; Kentaro Ishii; Shinichi Tsutsumi; Makoto Inoue; Takao Ichida; Masako N Hosono; Yukio Miki
Journal:  Radiol Phys Technol       Date:  2012-08-14

6.  Target tracking using DMLC for volumetric modulated arc therapy: a simulation study.

Authors:  Baozhou Sun; Dharanipathy Rangaraj; Lech Papiez; Swetha Oddiraju; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

7.  Failure mode and effect analysis-based quality assurance for dynamic MLC tracking systems.

Authors:  Amit Sawant; Sonja Dieterich; Michelle Svatos; Paul Keall
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

8.  Independent calculation of monitor units for VMAT and SPORT.

Authors:  Xin Chen; Karl Bush; Aiping Ding; Lei Xing
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

9.  Application of volumetric modulated arc therapy (VMAT) in a dual-vendor environment.

Authors:  Barbara Dobler; Karin Weidner; Oliver Koelbl
Journal:  Radiat Oncol       Date:  2010-10-25       Impact factor: 3.481

10.  Plan comparison of volumetric-modulated arc therapy (RapidArc) and conventional intensity-modulated radiation therapy (IMRT) in anal canal cancer.

Authors:  Sabine Vieillot; David Azria; Claire Lemanski; Carmen Llacer Moscardo; Sophie Gourgou; Jean-Bernard Dubois; Norbert Aillères; Pascal Fenoglietto
Journal:  Radiat Oncol       Date:  2010-10-13       Impact factor: 3.481