| Literature DB >> 31553525 |
Song Gao1, Tucker Netherton1, Mikhail A Chetvertkov1, Yuting Li1, Laurence E Court1, William E Simon2, Jie Shi2, Peter A Balter1.
Abstract
We tested whether an ionization chamber array (ICA) and a one-dimensional water scanner (1DS) could be used instead of a three-dimensional water scanning system (3DWS) for acceptance testing and commissioning verification of the Varian Halcyon-Eclipse Treatment Planning System (TPS). The Halcyon linear accelerator has a single 6-MV flattening-filter-free beam and a nonadjustable beam model for the TPS. Beam data were measured with a 1DS, ICA, ionization chambers, and electrometer. Acceptance testing and commissioning were done simultaneously by comparing the measured data with TPS-calculated percent-depth-dose (PDD) and profiles. The ICA was used to measure profiles of various field sizes (10-, 20-, and 28 cm2 ) at depths of dmax (1.3 cm), 5-, 10-, and 20 cm. The 1DS was used for output factors (OFs) and PDDs. OFs were measured with 1DS for various fields (2-28 cm2 ) at a source-to-surface distance of 90 cm. All measured data were compared with TPS-calculations. Profiles, off-axis ratios (OAR), PDDs and OFs were also measured with a 3DWS as a secondary check. Profiles between the ICA and TPS (ICA and 3DWS) at various depths across the fields indicated that the maximum discrepancies in high-dose and low-dose tail were within 2% and 3%, respectively, and the maximum distance-to-agreement in the penumbra region was <3 mm. The largest OAR differences between ICA and TPS (ICA and 3DWS) values were 0.23% (-0.25%) for a 28 × 28 cm2 field, and the largest point-by-point PDD differences between 1DS and TPS (1DS and 3DWS) were -0.41% ± 0.12% (-0.32% ± 0.17%) across the fields. Both OAR and PDD showed the beam energy is well matched to the TPS model. The average ratios of 1DS-measured OFs to the TPS (1DS to 3DWS) values were 1.000 ± 0.002 (0.999 ± 0.003). The Halcyon-Eclipse system can be accepted and commissioned without the need for a 3DWS.Entities:
Keywords: Halcyon linear accelerator; acceptance and commissioning; ionization chamber array
Mesh:
Substances:
Year: 2019 PMID: 31553525 PMCID: PMC6806699 DOI: 10.1002/acm2.12719
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1The IC array measurement setup position with solid water buildup.
Figure 21DS water tank setup. 1: water tank is aligned at external setup position. 2: water tank is filled, and position calibrated. 3: Image guidance is performed to set SSD and visualize the chamber (3a without brass cap; 3b with brass cap). 1DS, one‐dimensional water scanner.
Figure 3Profile measured with an ionization chamber profiler (ICA) with an SSD of 90 cm, for 10 × 10, 20 × 20, and 28 × 28 cm2 fields compare to TPS calculated and 3DWS scanned profiles of the same setup and field. Depths: (a) dmax, (b) 5 cm, (c) 10 cm, and (d) 20 cm. 3DWS, 3D water scanning system; TPS, treatment planning system; SSD, source to surface distance.
High‐dose region profile comparison between the ICA and TPS (ICA and 3DWS) for three fields and four depths. Shaded cells are mean (average) ± SD (σ) of point‐by‐point differences (%) from profiles. Clear cells are the maximum (max) and minimum (min) difference values.
| Depth, cm | 10 × 10 cm2 | 20 × 20 cm2 | 28 × 28 cm2 | ||
|---|---|---|---|---|---|
| ICA‐TPS | dmax | mean±σ | −0.03 ± 0.24 | 0.21 ± 0.32 | 0.19 ± 0.39 |
| max/min | 0.4/–0.8 | 1.1/−0.5 | 1.2/–1.0 | ||
| 5 | mean±σ | −0.24 ± 0.37 | −0.08 ± 0.41 | −0.06 ± 0.45 | |
| max/min | 0.3/–1.0 | 0.7/–1.0 | 0.8/–1.4 | ||
| 10 | mean±σ | −0.1 ± 0.36 | −0.04 ± 0.29 | 0.06 ± 0.24 | |
| max/min | 0.6/–1.1 | 0.6/–0.8 | 0.7/–0.8 | ||
| 20 | mean±σ | 0.05 ± 0.46 | 0.01 ± 0.25 | −0.08 ± 0.20 | |
| max/min | 1.0/–1.2 | 0.7/−0.6 | 0.5/‐0.8 | ||
| ICA‐3DWS | dmax | mean±σ | −0.06 ± 0.38 | −0.09 ± 0.27 | −0.09 ± 0.33 |
| max/min | 0.9/–1.0 | 0.4/−0.8 | 0.6/–0.7 | ||
| 5 | mean±σ | −0.19 ± 0.37 | –0.14 ± 0.37 | −0.21 ± 0.45 | |
| max/min | 0.5/–1.0 | 1.1/−1.1 | 0.6/–1.2 | ||
| 10 | mean±σ | 0.09 ± 0.53 | –0.19 ± 0.47 | −0.05 ± 0.31 | |
| max/min | 1.5/–0.7 | 1.4/–1.1 | 0.8/–0.8 | ||
| 20 | mean±σ | 0.07 ± 0.67 | −0.15 ± 0.45 | −0.17 ± 0.40 | |
| max/min | 1.8/–0.9 | 1.6/−1.1 | 1.8/–0.8 |
Abbreviations: 3DWS, 3D water tanks scans; ICA, ionization chamber array; TPS, treatment planning system.
Low‐dose region profiles and penumbra region comparison between the ICA and TPS (ICA and 3DWS) for three fields and four depths. Shaded cells are maximum point‐by‐point differences (%) from profiles. Clear cells are the maximum DTA in the penumbra region.
| Depth, cm | 10 × 10 cm2 | 20 × 20 cm2 | 28 × 28 cm2 | ||
|---|---|---|---|---|---|
| ICA‐TPS | dmax | Low dose, % | 1.0 | 1.8 | 1.7 |
| DTA, mm | –1.25 | 2.44 | 2.82 | ||
| 5 | Low dose, % | 2.5 | −0.6 | −2.8 | |
| DTA, mm | –2.18 | −2.91 | −2.73 | ||
| 10 | Low dose, % | 1.3 | −0.65 | 1.0 | |
| DTA, mm | −1.68 | 1.75 | −1.06 | ||
| 20 | Low dose, % | −1.11 | 1.1 | 2.1 | |
| DTA, mm | −2.36 | 1.71 | 2.46 | ||
| ICA‐3DWS | dmax | Low dose, % | −1.6 | 2.2 | 1.2 |
| DTA, mm | 1.9 | −2.46 | 1.56 | ||
| 5 | Low dose, % | −2.8 | −1.6 | −2.7 | |
| DTA, mm | 2.94 | 2.2 | −2.72 | ||
| 10 | Low dose. % | −2.1 | −2.21 | 1.7 | |
| DTA, mm | −2.94 | 1.54 | −2.2 | ||
| 20 | Low dose, % | −2.3 | −2.6 | −2.8 | |
| DTA, mm | 2.75 | −1.74 | 1.34 |
Abbreviations: 3DWS, 3D water tanks scans; DTA, Distance to agreement; ICA, ionization chamber array; TPS, Treatment Planning System.
Differences in off‐axis ratio between measurements between ICA and TPS (ICA and 3DWS).
| Field size | 20 × 20 cm2 | 28 × 28 cm2 | ||
|---|---|---|---|---|
| Depth, cm | dmax | 10 | dmax | 10 |
| ICA‐TPS | −0.02% | −0.38% | 0.23% | 0.04% |
| ICA‐3DWS | 0.25% | 0.25% | −0.25% | −0.22% |
Abbreviations: 3DWS, 3D water tanks scans; ICA, ionization chamber array; TPS, treatment planning system.
Figure 4PDDs scanned using the 1DS water tank with an SSD of 90 cm for field sizes from 2 × 2 cm2 to 28 × 28 cm2 compared with PDDs of the same setup calculated by the treatment planning system (TPS) and 3D water tanks scans (3DWS). 1DS, one‐dimensional water scanner; PDDs, percent‐depth‐dose; SSD, source to surface distance.
PDD comparison between the 1DS and TPS (1DS and 3DWS) from 1.0 cm to 30.0 cm for seven fields. Shaded cells are mean (average) ± SD (σ) of point‐by‐point differences (%) between PDDs. Clear cells are the maximum (max) and minimum (min) difference values.
| Field | 2 × 2 cm2 | 4 × 4 cm2 | 6 × 6 cm2 | 8 × 8 cm2 | 10 × 10 cm2 | 20 × 20 cm2 | 28 × 28 cm2 | |
|---|---|---|---|---|---|---|---|---|
| 1DS‐TPS | mean ± σ | −0.41 ± 0.12 | −0.30 ± 0.11 | −0.29 ± 0.17 | −0.25 ± 0.27 | −0.16 ± 0.22 | 0.02 ± 0.24 | −0.25 ± 0.23 |
| max/min | 0.13/−0.59 | 0.15/−0.49 | 0.14/−0.47 | 0.29/−0.56 | 0.23/−0.57 | 0.80/−0.32 | 0.77/−0.50 | |
| 1DS‐3DWS | mean ± σ | −0.25 ± 0.10 | −0.05 ± 0.13 | −0.05 ± 0.09 | −0.05 ± 0.10 | 0.04 ± 0.09 | −0.03 ± 0.08 | −0.32 ± 0.17 |
| max/min | 0.13/−0.43 | 0.43/−0.25 | 0.30/−0.19 | 0.30/−0.23 | 0.35/−0.18 | 0.25/−0.18 | 0.14/−0.55 | |
| dmax (cm) | 1DS | 1.19 | 1.29 | 1.26 | 1.29 | 1.25 | 1.2 | 1.06 |
| TPS | 1.15 | 1.25 | 1.25 | 1.20 | 1.20 | 1.18 | 1.12 | |
| 3DWS | 1.16 | 1.26 | 1.30 | 1.25 | 1.26 | 1.26 | 1.06 |
Abbreviations: 3DWS, 3D water tanks scans; 1DS, one‐dimensional water scanner; PDDs, percent‐depth‐dose; TPS, treatment planning system.
The output factors measured using the 1DS with an SSD of 90 cm at a depth of 10 cm for various field sizes and compared with data from the TPS and measurements obtained with a 3DWS.
| Field Size, cm2 | Output factor | Difference, ratio | |||
|---|---|---|---|---|---|
| 1DS | TPS | 3DWS | 1DS/TPS | 1DS/3DWS | |
| 2 × 2 | 0.801 | 0.801 | 0.803 | 1.000 | 0.998 |
| 4 × 4 | 0.881 | 0.878 | 0.880 | 1.003 | 1.001 |
| 6 × 6 | 0.931 | 0.927 | 0.931 | 1.004 | 1.000 |
| 8 × 8 | 0.970 | 0.970 | 0.970 | 1.000 | 1.000 |
| 10 × 10 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 |
| 14 × 14 | 1.044 | 1.045 | 1.045 | 0.999 | 0.999 |
| 20 × 20 | 1.085 | 1.088 | 1.088 | 0.997 | 0.997 |
| 24 × 24 | 1.102 | 1.106 | 1.107 | 0.996 | 0.995 |
| 28 × 28 | 1.111 | 1.115 | 1.117 | 0.998 | 0.995 |
| 4 × 20 | 0.940 | 0.941 | 0.938 | 0.999 | 1.002 |
| 20 × 4 | 0.940 | 0.940 | 0.940 | 1.000 | 1.000 |
Abbreviations: 3DWS, 3D water tanks scans; 1DS, one‐dimensional water scanner; SSD, source to surface distance; TPS, treatment planning system.