Photon beam quality variations of a flattening filter free linear accelerator.

Dietmar Georg1, Gabriele Kragl, Sacha af Wetterstedt, Patrick McCavana, Brendan McClean, Tommy Knöös.   

Abstract

PURPOSE: Recently, there has been an increasing interest in operating conventional linear accelerators without a flattening filter. The aim of this study was to determine beam quality variations as a function of off-axis ray angle for unflattened beams. In addition, a comparison was made with the off-axis energy variation in flattened beams.
METHODS: Two Elekta Precise linear accelerators were modified in order to enable radiation delivery with and without the flattening filter in the beam line. At the Medical University Vienna (Vienna, Austria), half value layer (HVL) measurements were performed for 6 and 10 MV with an in-house developed device that can be easily mounted on the gantry. At St. Luke's Hospital (Dublin, Ireland), measurements were performed at 6 MV in narrow beam geometry with the gantry tilted around 270 degrees with pinhole collimators, an attenuator, and the chamber positioned on the table. All attenuation measurements were performed with ionization chambers and a buildup cap (2 mm brass) or a PMMA mini phantom (diameter 3 cm, measurement depth 2.5 cm).
RESULTS: For flattened 6 and 10 MV photon beams from the Elekta linac the relative HVL(theta) varies by about 11% for an off-axis ray angle theta = 10 degrees. These results agree within +/- 2% with a previously proposed generic off-axis energy correction. For unflattened beams, the variation was less than 5% in the whole range of off-axis ray angles up to 10 degrees. The difference in relative HVL data was less than 1% for unflattened beams at 6 and 10 MV.
CONCLUSIONS: Off-axis energy variation is rather small in unflattened beams and less than half the one for flattened beams. Thus, ignoring the effect of off-axis energy variation for dose calculations in unflattened beams can be clinically justified.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20175465     DOI: 10.1118/1.3264617

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Unflattened photon beams from the standard flattening filter free accelerators for radiotherapy: Advantages, limitations and challenges.

Authors:  Sunil Dutt Sharma
Journal:  J Med Phys       Date:  2011-07

2.  Acceptance criteria for flattening filter-free photon beam from standard medical electron linear accelerator: AERB task group recommendations.

Authors:  G Sahani; S D Sharma; P K Dash Sharma; D D Deshpande; P S Negi; V K Sathianarayanan; G K Rath
Journal:  J Med Phys       Date:  2014-10

3.  Equivalent-quality unflattened photon beam modeling, planning, and delivery.

Authors:  Yunfei Huang; Ryan T Flynn; R Alfredo C Siochi; John E Bayouth
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

4.  Dosimetric properties of equivalent-quality flattening filter-free (FFF) and flattened photon beams of Versa HD linear accelerator.

Authors:  Mukesh N Meshram; Srimanta Pramanik; C P Ranjith; Saravana K Gopal; Rishabh Dobhal
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

5.  Structural Shielding Design of a 6 MV Flattening Filter Free Linear Accelerator: Indian Scenario.

Authors:  Bibekananda Mishra; T Palani Selvam; P K Dash Sharma
Journal:  J Med Phys       Date:  2017 Jan-Mar

6.  Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group.

Authors:  Ying Xiao; Stephen F Kry; Richard Popple; Ellen Yorke; Niko Papanikolaou; Sotirios Stathakis; Ping Xia; Saiful Huq; John Bayouth; James Galvin; Fang-Fang Yin
Journal:  J Appl Clin Med Phys       Date:  2015-05-08       Impact factor: 2.102

7.  Comparison of detector performance in small 6 MV and 6 MV FFF beams using a Versa HD accelerator.

Authors:  Paula Monasor Denia; María Del Carmen Castellet García; Carla Manjón García; Juan David Quirós Higueras; Noelia de Marco Blancas; Jorge Bonaque Alandí; Xavier Jordi Juan Senabre; Agustín Santos Serra; Juan López-Tarjuelo
Journal:  PLoS One       Date:  2019-03-11       Impact factor: 3.240

8.  Commissioning measurements for photon beam data on three TrueBeam linear accelerators, and comparison with Trilogy and Clinac 2100 linear accelerators.

Authors:  Gloria P Beyer
Journal:  J Appl Clin Med Phys       Date:  2013-01-07       Impact factor: 2.102

9.  Surface dose variations in 6 and 10 MV flattened and flattening filter-free (FFF) photon beams.

Authors:  Jason Cashmore
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

10.  Commissioning and quality assurance of HalcyonTM 2.0 linear accelerator.

Authors:  Pushpraj K Pathak; S K Vashisht; S Baby; P K Jithin; Y Jain; R Mahawar; V G G K Sharan
Journal:  Rep Pract Oncol Radiother       Date:  2021-06-09
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.