Radiation exposure to patients from image guidance procedures and techniques to reduce the imaging dose.

George X Ding1, Peter Munro.   

Abstract

PURPOSE: To compare imaging doses from MV images, kV radiographs, and kV-CBCT and describe methods to reduce the dose to patient's organs using existing on-board imaging devices. METHOD AND MATERIALS: Monte Carlo techniques were used to simulate kV X-ray sources. The kV image doses to a variety of patient anatomies were calculated by using the simulated realistic sources to deposit dose in patient CT images. For MV imaging, the doses for the same patients were calculated using a commercial treatment planning system.
RESULTS: Portal imaging results in the largest dose to anatomic structures, followed by Varian OBI CBCT, Varian TrueBeam CBCT and then kV radiographs. The imaging doses for the 50% volume from the DVHs, D50, to the eyes for representative head images are 4.3-4.8cGy; 0.05-0.06cGy; 0.04-0.05cGy; and, 0.12cGy; D50 to the bladder for representative pelvis images are 3.3cGy; 1.6cGy; 1.0cGy; and, 0.07cGy; while D50 to the heart for representative thorax images are 3.5cGy; 0.42cGy; 0.2cGy; and, 0.07cGy; when using portal imaging, OBI kV-CBCT scans, TrueBeam kV-CBCT scans and kV radiographs, respectively. The orientation of the kV beam can affect organ dose. For example, D50 to the eyes can be reduced from 0.12cGy using AP and right lateral radiographs to 0.008-0.017cGy when using PA and right lateral radiographs. In addition, organ exposures can be further reduced to 15-70% of their original values with the use of a full-fan, bow-tie filter for kV radiographs. In contrast, organ doses increase by a factor of ∼2-4 if bow-tie filters are not used during kV-CBCT acquisitions.
CONCLUSION: Current on-board kV imaging devices result in much lower imaging doses compared to MV imagers even taking into account of higher bone dose from kV X-rays. And a variety of approaches are available to significantly reduce the image doses.
Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Imaging dose reduction techniques; MV imaging dose; kV-CBCT dose; kV-CBCT in TrueBeam; kV-radiograph dose

Mesh:

Year:  2013        PMID: 23830468     DOI: 10.1016/j.radonc.2013.05.034

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  30 in total

1.  Technical Note: Imaging dose resulting from optimized procedures with limited-angle intrafractional verification system during stereotactic body radiation therapy lung treatment.

Authors:  George X Ding; Yawei Zhang; Lei Ren
Journal:  Med Phys       Date:  2019-04-29       Impact factor: 4.071

Review 2.  "Advances in the surgical management of bone tumors".

Authors:  Justin E Bird
Journal:  Curr Oncol Rep       Date:  2014-07       Impact factor: 5.075

3.  Use of a laser-guided collimation system to perform direct kilovoltage x-ray spectra measurements on a linear accelerator onboard imager.

Authors:  Zachary Grelewicz; Andrew H Belcher; Rodney D Wiersma
Journal:  Med Phys       Date:  2018-10-09       Impact factor: 4.071

4.  Spatial and rotational quality assurance of 6DOF patient tracking systems.

Authors:  Andrew H Belcher; Xinmin Liu; Zachary Grelewicz; Rodney D Wiersma
Journal:  Med Phys       Date:  2016-06       Impact factor: 4.071

5.  Radiation protection in radiological imaging: a survey of imaging modalities used in Japanese institutions for verifying applicator placements in high-dose-rate brachytherapy.

Authors:  Hiroyuki Okamoto; Satoshi Kito; Naoki Tohyama; Shunsuke Yonai; Ryu Kawamorita; Masaru Nakamura; Takahiro Fujimoto; Syoji Tani; Akihiro Yomoda; Toru Isobe; Hiroshi Furukawa; Kikuo Kotaka; Jun Itami; Hitoshi Ikushima; Takushi Dokiya; Yoshiyuki Shioyama
Journal:  J Radiat Res       Date:  2021-01-01       Impact factor: 2.724

6.  Evaluation of administered dose using portal images in craniospinal irradiation of pediatric patients.

Authors:  Carina Marques Coelho; Raquel Calçada; Sofia Rodrigues; Juan Antonio Barragán; Ana Cravo Sá; Ana Paula Macedo; Maria de Fátima Monsanto
Journal:  Radiol Phys Technol       Date:  2017-03-21

7.  Accounting for patient size in the optimization of dose and image quality of pelvis cone beam CT protocols on the Varian OBI system.

Authors:  Tim J Wood; Craig S Moore; Carl J Horsfield; John R Saunderson; Andrew W Beavis
Journal:  Br J Radiol       Date:  2015-09-30       Impact factor: 3.039

8.  Streamlining the image-guided radiotherapy process for proton beam therapy.

Authors:  Lucy Siew Chen Davies; Louise McHugh; Marianne Aznar; Josh Lindsay; Cynthia Eccles
Journal:  Br J Radiol       Date:  2021-09-14       Impact factor: 3.039

9.  Evaluation of patient organ doses from kilovoltage cone-beam CT imaging in radiation therapy.

Authors:  Alper Özseven; Bahar Dirican
Journal:  Rep Pract Oncol Radiother       Date:  2021-04-14

10.  Commissioning and quality assurance of HalcyonTM 2.0 linear accelerator.

Authors:  Pushpraj K Pathak; S K Vashisht; S Baby; P K Jithin; Y Jain; R Mahawar; V G G K Sharan
Journal:  Rep Pract Oncol Radiother       Date:  2021-06-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.