| Literature DB >> 30016578 |
Yuting Li1,2, Tucker Netherton1,3, Paige L Nitsch3, Song Gao3, Ann H Klopp3, Peter A Balter3, Laurence E Court3.
Abstract
PURPOSE: To evaluate the ability of the machine performance check (MPC) on the Halcyon to detect errors, with comparison with the TrueBeam.Entities:
Keywords: zzm321990MPCzzm321990; Halcyon; TrueBeam
Mesh:
Year: 2018 PMID: 30016578 PMCID: PMC6123154 DOI: 10.1002/acm2.12391
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Summary of measurements for MPC evaluation
| MPC test | Measurement method | |
|---|---|---|
| TrueBeam | Halcyon | |
| Beam constancy — output change | Attenuate the entire beam by solid water slabs | |
| Beam constancy — uniformity change | Attenuate half the beam by solid water slabs | |
| MLC position | NA | Miscalibrate one MLC position |
| Absolute gantry offset | NA | Miscalibrate gantry position |
| Relative gantry offset | Rotate the MPC phantom during MPC test | |
| Absolute couch position | Shift the MPC phantom | |
| Relative couch shift | Shift the phantom during the MPC test | |
| Virtual to isocenter shift | NA | Shift the MPC phantom during the MPC test |
Figure 1Illustration of measurement setup for (a) output (a) and (b) uniformity tests.
Figure 2(a–c) Measurement setup for MPC couch translation tests of the Halcyon system. These pictures show the setup for shifting the phantom (a) longitudinally, (b) laterally and (c) vertically. (d) Measurement setup for MPC relative gantry rotation test of Halcyon.
Results of comparison between known offset and MPC check results
| Test | Machine | Difference between known value and MPC | ||||
|---|---|---|---|---|---|---|
| Average difference | Standard deviation | Minimum difference | Maximum difference | |||
| % Output | Halcyon | 1.31 | 0.48 | 0.70 | 1.95 | |
| TrueBeam | 0.51 | 0.28 | 0.16 | 0.84 | ||
| Absolute gantry angle (degree) | Halcyon | 0.01 | 0.01 | 0.00 | 0.03 | |
| Couch translation (mm) | Halcyon | Longitudinal | 0.02 | 0.03 | 0.00 | 0.06 |
| Lateral | 0.09 | 0.06 | 0.04 | 0.15 | ||
| Vertical | 0.02 | 0.02 | 0.01 | 0.05 | ||
| TrueBeam | Longitudinal | 0.02 | 0.02 | 0.01 | 0.05 | |
| Lateral | 0.10 | 0.09 | 0.01 | 0.22 | ||
| Virtual to actual isocenter couch translation (mm) | Halcyon | Longitudinal | 0.18 | 0.06 | 0.13 | 0.27 |
| Lateral | 0.28 | 0.15 | 0.06 | 0.43 | ||
| Vertical | 0.33 | 0.02 | 0.31 | 0.36 | ||
| MLC position (mm) | Halcyon | 0.05 | 0.10 | −0.01 | 0.23 | |
Output data is reported for intentional reductions in output by up to 5%. The differences between MPC and Profiler reported output increased with larger changes in output.
Figure 3(a) Halcyon data: Uniformity measured using the MPC compared with symmetry measured, using the IC profiler. (b) TrueBeam data: Uniformity measured using the MPC compared with symmetry and flatness measured, using the IC Profiler.
Detailed results of MPC relative gantry rotation tests
| Machine | Angle offset (degree) | Number of gantry positions where phantom was rotated | ||||||
|---|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | ||
| Halcyon | MPC results (degree) | |||||||
| 0.1 | 0.05 | 0.02 | 0.04 | 0.05 | 0.04 | −0.04 | ||
| 0.3 | 0.19 | 0.16 | 0.16 | 0.13 | −0.14 | −0.16 | ||
| 0.5 | 0.37 | 0.31 | 0.28 | 0.23 | −0.26 | −0.31 | ||
| 1 | 0.8175 | 0.69 | 0.58 | 0.47 | −0.57 | −0.69 | ||
| 1.5 | 1.21 | 1.04 | 0.89 | 0.71 | −1.19 | −1.09 | ||
| 2 | 1.64 | 1.42 | 1.19 | 0.96 | −1.19 | −1.42 | ||
| 5 | 4.29 | 3.68 | 3.09 | 2.47 | −3.07 | −3.65 | ||
| TrueBeam | 0.5 | 0.51 | 0.48 | 0.41 | 0.35 | −0.4 | −0.46 | −0.48 |
| 1 | 0.99 | 0.88 | 0.8 | 0.66 | −0.79 | −0.88 | −0.94 | |
| 2 | 2.16 | 1.89 | 1.58 | 1.28 | −1.57 | −1.88 | −2.12 | |