Kyoung Hae Kang1, Mi Yeon Song1, Kook Young Kim1, Kyu-Yeon Hwang1, Young-A Kwon1, Kyungmin Koh1.
Abstract
PURPOSE: To evaluate the performance of two intraocular lenses (IOLs). The new monofocal IOL using a higher-order aspheric optic (Tecnis Eyhance ICB00) was compared to a standard monofocal IOL (Tecnis monofocal ZCB00) of the identical platform and material.Entities:
Keywords: Cataract; Intraocular lens implantation; Intraocular lenses; Ocular accommodation; Ocular vision
Year: 2021 PMID: 33845556 PMCID: PMC8046615 DOI: 10.3341/kjo.2020.1115
Source DB: PubMed Journal: Korean J Ophthalmol ISSN: 1011-8942
Preoperative characteristics of patients by group
| Parameter | Eyhance ICB00 | ZCB00 | |
|---|---|---|---|
| No. of eyes | 111 | 86 | NA |
| No. of patients | 111 | 86 | NA |
| Female patients | 57 (51) | 51 (59) | 0.201 |
| Right eye dominant | 99 (89) | 79 (92) | 0.677 |
| Age (yr) | 65.23 ± 8.15 (45 to 82) | 64.34 ± 9.15 (41 to 81) | 0.475 |
| UDVA (logMAR) | 0.58 ± 0.29 (0.2 to 1.4) | 0.57 ± 0.29 (0.2 to 1.6) | 0.848 |
| CDVA (logMAR) | 0.33 ± 0.31 (0.05 to 1) | 0.30 ± 0.18 (0.05 to 1) | 0.433 |
| MRSE (D) | −0.67 ± 2.38 (−10.00 to 3.00) | −0.41 ± 2.48 (−11.50 to 3.00) | 0.461 |
Values are presented as number (%) or mean ± standard deviation (range).
NA = not applicable; UDVA = uncorrected distance visual acuity; logMAR = logarithm of the minimum angle of resolution; CDVA = corrected distance visual acuity; MRSE = manifest refraction spherical equivalent; D = diopters.
Postoperative 3 months visual outcomes
| Parameter | Eyhance ICB00 | ZCB00 | |
|---|---|---|---|
| UDVA (logMAR) | 0.11 ± 0.10 (0 to 0.50) | 0.10 ± 0.14 (0 to 0.65) | 0.451 |
| CDVA (logMAR) | 0.04 ± 0.06 (0 to 0.20) | 0.03 ± 0.06 (0 to 0.40) | 0.215 |
| UIVA (logMAR) | 0.24 ± 0.10 (0.10 to 0.50) | 0.34 ± 0.12 (0.20 to 0.70) | <0.001 |
| UNVA (logMAR) | 0.46 ± 0.14 (0.15 to 0.80) | 0.51 ± 0.19 (0.15 to 1.00) | 0.038 |
| Target SE (D) | −0.21 ± 0.23 (−0.81 to 0) | −0.25 ± 0.27 (−0.92 to 0) | 0.098 |
| MRSE (D) | −0.22 ± 0.39 (−1.25 to 0.50) | −0.21 ± 0.37 (−1.35 to 0.50) | 0.322 |
| Mean error (D) | −0.01 ± 0.35 (−0.63 to 1.00) | 0.04 ± 0.36 (−0.62 to 0.95) | 0.134 |
| Mean absolute error (D) | 0.29 ± 0.21 (0 to 1.00) | 0.31 ± 0.20 (0 to 0.95) | 0.855 |
| RE within ± 0.5 D | 103 (93) | 81 (94) | 0.919 |
Values are presented as mean ± standard deviation (range) or number (%).
UDVA = uncorrected distance visual acuity; logMAR = logarithm of the minimum angle of resolution; CDVA = corrected distance visual acuity; UIVA = uncorrected intermediate visual acuity; UNVA = uncorrected near visual acuity; SE = spherical equivalent; D = diopters; MRSE = manifest refraction spherical equivalent; RE = refractory error.
p < 0.001;
p < 0.05.
Fig. 1Mean monocular defocus curves obtained in Tecnis Eyhance ICB00 and Tecnis monofocal ZCB00 groups. logMAR = logarithm of the minimal angle of resolution; D = diopters.
Optical quality parameters evaluated by OQAS at a 4.0-mm pupil diameter
| Parameter | Eyhance ICB00 | ZCB00 | |
|---|---|---|---|
| OSI | 1.28 ± 0.41 (0.76 to 2.20) | 1.22 ± 0.39 (0.69 to 2.45) | 0.241 |
| MTF cutoff (c/deg) | 26.53 ± 5.05 (17.01 to 40.12) | 27.08 ± 4.42 (16.22 to 36.11) | 0.414 |
| Strehl ratio | 0.17 ± 0.03 (0.11 to 0.23) | 0.18 ± 0.02 (0.13 to 0.24) | 0.825 |
Values are presented as mean ± standard deviation (range).
OQAS = optical quality analysis system; OSI = objective scatter index; MTF = modulation transfer function.