| Literature DB >> 34608065 |
Young Joon Jeon1, Yisang Yoon1, Tae-Im Kim2, Kyungmin Koh1.
Abstract
PURPOSE: This study compared the extended depth of focus (EDOF) intraocular lens (IOL) (ZXR00; Tecnis Symfony, Johnson & Johnson Vision, Santa Ana, CA, US) to a novel, higher-order aspheric monofocal IOL (ICB00; Tecnis Eyhance, Johnson & Johnson Vision, Santa Ana, CA, US) which uses the same platform and material.Entities:
Mesh:
Year: 2021 PMID: 34608065 PMCID: PMC8673848 DOI: 10.1097/APO.0000000000000439
Source DB: PubMed Journal: Asia Pac J Ophthalmol (Phila) ISSN: 2162-0989
Preoperative Characteristics of Patients Per Group
| Parameter | Symfony ZXR00 | Eyhance ICB00 | |
| Eyes, n | 72 | 102 | N/A |
| Female, n | 38 (53%) | 57 (56%) | 0.361 |
| Right eye, n | 37 (51%) | 53 (52%) | 0.398 |
| Age, y | 59.6 ± 10.6 (49 to 70) | 65.2 ± 8.2 (45 to 82) | <0.001∗ |
| UDVA (LogMAR) | |||
| Mean ± SD | 0.53 ± 0.37 | 0.55 ± 0.35 | 0.771 |
| Range | 0.16 to 0.9 | 0.2 to 0.9 | |
| CDVA (LogMAR) | |||
| Mean ± SD | 0.30 ± 0.30 | 0.33 ± 0.32 | 0.382 |
| Range | 0.00 to 0.60 | 0.01 to 0.65 | |
| MRSE (D) | |||
| Mean ± SD | −1.38 ± 2.98 | −0.74 ± 2.45 | 0.134 |
| Range | −4.36 to 1.60 | −3.19 to 1.71 | |
CDVA indicates corrected distance visual acuity; D, diopter; LogMAR, logarithm of the minimum angle of resolution; MRSE, manifest refraction spherical equivalent; N/A, not applicable; SD, standard deviation; UDVA, uncorrected distance visual acuity.
P < 0.001.
Postoperative 3 Months Visual Outcomes
| Parameter | Symfony ZXR00 | Eyhance ICB00 | |
| UDVA (LogMAR) | |||
| Mean ± SD | 0.13 ± 0.13 | 0.11 ± 0.10 | 0.163 |
| Range | 0 to 0.26 | 0.01 to 0.21 | |
| CDVA (LogMAR) | |||
| Mean ± SD | 0.05 ± 0.08 | 0.03 ± 0.05 | 0.253 |
| Range | 0 to 0.13 | 0 to 0.08 | |
| UIVA (LogMAR) | |||
| Mean ± SD | 0.22 ± 0.12 | 0.26 ± 0.09 | 0.105 |
| Range | 0.10 to 0.34 | 0.16 to 0.36 | |
| UNVA (LogMAR) | |||
| Mean ± SD | 0.32 ± 0.11 | 0.42 ± 0.11 | <0.001∗ |
| Range | 0.21 to 0.43 | 0.31 to 0.53 | |
| Target SE (D) | |||
| Mean ± SD | −0.35 ± 0.19 | −0.32 ± 0.20 | 0.395 |
| Range | −0.54 to −0.16 | −0.52 to −0.12 | |
| MRSE (D) | |||
| Mean ± SD | −0.22 ± 0.50 | −0.22 ± 0.40 | 0.993 |
| Range | −0.72 to 0.28 | −0.62 to 0.18 | |
| Mean prediction error (D) | |||
| Mean ± SD | 0.12 ± 0.53 | 0.10 ± 0.39 | 0.720 |
| Range | −0.41 to 0.65 | −0.29 to 0.49 | |
| Mean absolute error (D) | |||
| Mean ± SD | 0.36 ± 0.29 | 0.31 ± 0.21 | 0.121 |
| Range | 0.09 to 0.77 | 0.10 to 1.52 | |
| RE within ± 0.5 D, n | 65 (90%) | 94 (92%) | 0.211 |
CDVA indicates corrected distance visual acuity; D, diopter; LogMAR, logarithm of the minimum angle of resolution; MRSE, manifest refraction spherical equivalent; n, number; RE, refractory error; SD, standard deviation; SE, spherical equivalent; UIVA, uncorrected intermediate visual acuity; UNVA, uncorrected near visual acuity; UDVA, uncorrected distance visual acuity.
P < 0.001
FIGURE 1Mean monocular defocus curves obtained from the Tecnis Symfony (ZXR00) and Tecnis Eyhance (ICB00) groups. D indicates diopters; LogMAR interval = logarithm of the minimal angle of resolution.
Optical Quality Parameters Assessed by OQAS With a Pupil Diameter of 4.0 mm After 3 Months of Surgery
| Parameter | Symfony ZXR00 | Eyhance ICB00 | |
| OSI | 2.08 ± 0.92 (1.00 to 4.30) | 1.22 ± 0.37 (0.76 to 2.20) | <0.001 |
| MTF cutoff (c/deg) | 24.29 ± 9.08 (11.47 to 46.73) | 27.29 ± 4.82 (17.22 to 40.12) | <0.001 |
| Strehl ratio | 0.14 ± 0.05 (0.08 to 0.26) | 0.18 ± 0.03 (0.11 to 0.23) | <0.001 |
D indicates diopters; MTF, modulation transfer function; OQAS, optical quality analysis system; OSI, objective scatter index; SD, standard deviation.