| Literature DB >> 35836144 |
Da Eun Shin1, Hun Lee2, Kyungmin Koh3.
Abstract
BACKGROUND: Recently, a new presbyopia-correcting intraocular lens (IOL) that combines extended depth-of-focus and bifocal profiles (ZFR00: Tecnis® Synergy®, Johnson & Johnson Vision, Santa Ana, CA, USA) has been established and several studies have been reported. We attempted to compare the performance with a standard IOL (ZCB00: Tecnis® monofocal, Johnson & Johnson Vision, Santa Ana, CA, USA) manufactured using the same material from the same company, which has been extensively used worldwide.Entities:
Keywords: Continuous range of vision; Extended depth of focus; Intermediate vision; Intraocular lens; Near vision
Mesh:
Year: 2022 PMID: 35836144 PMCID: PMC9284875 DOI: 10.1186/s12886-022-02516-6
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.086
Preoperative characteristics in each group
| Parameter | ZCB00 | ZFR00 | |
|---|---|---|---|
| Patients/Eyes, n | 21/32 | 25/32 | N/A |
| Female, n | 14 (67%) | 11 (44%) | .616* |
| Right eye, n | 15 (47%) | 16 (50%) | .803* |
| Age (years) | 66.0 ± 10.1 (40 to 82) | 65.1 ± 4.7 (59 to 77) | .141† |
| UDVA (LogMAR) | 0.34 ± 0.15 (0.1 to 0.7) | 0.29 ± 0.13 (0.1 to 0.6) | .077† |
| CDVA (LogMAR) | 0.21 ± 0.10 (0.05 to 0.4) | 0.18 ± 0.12 (0 to 0.4) | .311† |
| MRSE (D) | 0.43 ± 1.43 (−3.0 to 3.5) | 0.69 ± 1.48 (−2.0 to 3.0) | .594‡ |
N/A not applicable, UDVA uncorrected distance visual acuity, CDVA corrected distance visual acuity, LogMAR logarithm of the minimum angle of resolution, SD standard deviation, MRSE manifest refraction spherical equivalent, D diopter
*Chi-square test
†Mann-Whitney U test
‡Unpaired student t-test
Postoperative 3 months visual outcomes
| Parameter | ZCB00 | ZFR00 | |
|---|---|---|---|
| UDVA (LogMAR) | 0.08 ± 0.06 (0 to 0.2) | 0.06 ± 0.07 (0 to 0.2) | .204* |
| CDVA (LogMAR) | 0.03 ± 0.05 (0 to 0.2) | 0.03 ± 0.05 (0 to 0.2) | .145* |
| UIVA (LogMAR) 66 cm | 0.33 ± 0.12 (0.2 to 0.6) | 0.02 ± 0.05 (0 to 0.2) | <.001* |
| UNVA (LogMAR) 40 cm | 0.48 ± 0.15 (0.2 to 0.8) | 0.03 ± 0.06 (0 to 0.2) | <.001* |
| UNVA (LogMAR) 33 cm | 0.62 ± 0.09 (0.5 to 0.9) | 0.09 ± 0.10 (0 to 0.3) | <.001* |
| Target SE (D) | 0.06 ± 0.12 (− 0.20 to 0.20) | 0.09 ± 0.14 (− 0.15 to 0.38) | .262* |
| MRSE (D) | − 0.21 ± 0.34 (− 1.00 to 0.50) | −0.09 ± 0.31 (− 0.75 to 0.38) | .271* |
| Refractive error (D) | −0.29 ± 0.34 (− 1.20 to 0.60) | 0.17 ± 0.30 (− 0.71 to 0.29) | .401† |
| Mean absolute error (D) | 0.35 ± 0.28 (0.03 to 1.20) | 0.28 ± 0.20 (0 to 0.71) | .317* |
UDVA uncorrected distance visual acuity, CDVA corrected distance visual acuity, UIVA uncorrected intermediate visual acuity, UNVA uncorrected near visual acuity, LogMAR logarithm of the minimum angle of resolution, SD standard deviation, SE spherical equivalent, MRSE manifest refraction spherical equivalent, D diopter, RE refractory error, n number
*Mann-Whitney U test
†Unpaired student t-test
Fig. 1Comparison of monocular visual acuity between Tecnis® monofocal (ZCB00) intraocular lens (IOL) and Tecnis® Synergy® (ZFR00) IOL groups. LogMAR = logarithm of the minimal angle of resolution; UDVA = uncorrected distance visual acuity; CDVA = corrected distance visual acuity; UIVA = uncorrected intermediate visual acuity; UNVA40 = uncorrected near visual acuity at 40 cm; UNVA33 = uncorrected near visual acuity at 33 cm; *** = p < .001
Fig. 2Mean monocular defocus curves obtained from the Tecnis® monofocal (ZCB00) intraocular lens (IOL) and Tecnis® Synergy® (ZFR00) IOL groups. logMAR = logarithm of the minimal angle of resolution; D = diopters
Objective optical quality parameters assessed by a HD Analyzer with a pupil diameter of 4.0 mm after 3 months of surgery
| Parameter | ZCB00 | ZFR00 | |
|---|---|---|---|
| OSI | 1.22 ± 0.42 (0.76 to 2.20) | 5.19 ± 2.18 (2.00 to 10.20) | <.001* |
| MTF cutoff (c/deg) | 27.37 ± 4.73 (17.22 to 33.15) | 10.40 ± 4.71 (4.00 to 19.81) | <.001* |
| Strehl ratio | 0.18 ± 0.03 (0.13 to 0.24) | 0.08 ± 0.02 (0.05 to 0.12) | <.001* |
OSI objective scatter index, MTF modulation transfer function, SD standard deviation
*Mann-Whitney U test