Repeatability of measurements with a double-pass system.

Alain Saad1, Marc Saab, Damien Gatinel.   

Abstract

PURPOSE: To evaluate the repeatability of measurements with a double-pass system.
SETTING: Rothschild Foundation, Paris, France.
METHODS: Eyes were separated into 2 control groups (<30 years old and >40 years), a post-refractive surgery group, and a cataract group. Measurements were performed using the Optical Quality Analysis System. The main outcome measures were the objective scattering index (OSI), the cutoff frequency of the modulation transfer function (MTF), and the Strehl ratio. The repeatability limit was obtained from the individual standard deviations.
RESULTS: Forty-two eyes were evaluated. The mean OSI value was 0.47 +/- 0.11 (SD) in the younger control group, 1.73 +/- 0.26 in the older control group, 1.34 +/- 0.16 in the post-refractive surgery group, and 6.15 +/- 0.50 in the cataract group. The mean cutoff MTF value was 39.44 +/- 3.93 cycles per degree (cpd), 26.07 +/- 3.89 cpd, 28.34 +/- 2.84 cpd, and 13.3 +/- 1.69 cpd, respectively, and the mean Strehl ratio, 0.234 +/- 0.023, 0.146 +/- 0.021, 0.169 +/- 0.023, and 0.098 +/- 0.010, respectively. The repeatability limit for the whole population was 0.841 (33.5%) for the OSI, 8.499 (31.1%) for the cutoff MTF, and 0.051 (31%) for the Strehl ratio.
CONCLUSIONS: The repeatability limit was good and equivalent for the OSI, the MTF, and the Strehl ratio values. There was a wide interval between the normal and pathologic threshold for OSI measurements, indicating that the reliability of the double-pass device complies with the requirements for quantitative assessment of scattering. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned. Copyright 2010 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20117702     DOI: 10.1016/j.jcrs.2009.07.033

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  40 in total

1.  Reproducibility of optical quality parameters measured at objective and subjective best focuses in a double-pass system.

Authors:  Ai-Lian Hu; Li-Ya Qiao; Ye Zhang; Xiao-Gu Cai; Lei Li; Xiu-Hua Wan
Journal:  Int J Ophthalmol       Date:  2015-10-18       Impact factor: 1.779

2.  Non-cycloplegic spherical equivalent refraction in adults: comparison of the double-pass system, retinoscopy, subjective refraction and a table-mounted autorefractor.

Authors:  Meritxell Vilaseca; Montserrat Arjona; Jaume Pujol; Elvira Peris; Vanessa Martínez
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

3.  Visual quality assessment after presbyopic laser in-situ keratomileusis.

Authors:  Dong Hui Lim; Eui-Sang Chung; Myoung Joon Kim; Tae-Young Chung
Journal:  Int J Ophthalmol       Date:  2018-03-18       Impact factor: 1.779

4.  Three-dimensional visualization of opacifications in the murine crystalline lens by in vivo optical coherence tomography.

Authors:  Pablo Eugui; Danielle J Harper; Stefan Kummer; Antonia Lichtenegger; Johanna Gesperger; Tanja Himmel; Marco Augustin; Conrad W Merkle; Martin Glösmann; Bernhard Baumann
Journal:  Biomed Opt Express       Date:  2020-03-19       Impact factor: 3.732

5.  Optical quality of toric intraocular lens implantation in cataract surgery.

Authors:  Xian-Wen Xiao; Jing Hao; Hong Zhang; Fang Tian
Journal:  Int J Ophthalmol       Date:  2015-02-18       Impact factor: 1.779

6.  Binocular summation and visual function with induced anisocoria and monovision.

Authors:  José J Castro; Margarita Soler; Carolina Ortiz; José R Jiménez; Rosario G Anera
Journal:  Biomed Opt Express       Date:  2016-09-23       Impact factor: 3.732

7.  Comparative study of objective visual quality between FS-LASIK and SMART in myopia.

Authors:  Yuan Wu; Yue Huang; Shu-Han Wang; Gui-Qin Wang; Ao-Miao Yu; Shao-Zhen Zhao; Rui-Hua Wei; Rui-Bo Yang; Chen Zhang
Journal:  Int J Ophthalmol       Date:  2022-03-18       Impact factor: 1.779

8.  Comparative analysis of a presbyopia-correcting intraocular lens that combines extended depth-of-focus and bifocal profiles with a standard monofocal intraocular lens.

Authors:  Da Eun Shin; Hun Lee; Kyungmin Koh
Journal:  BMC Ophthalmol       Date:  2022-07-14       Impact factor: 2.086

9.  Intra- and Intersession Repeatability of an Optical Quality and Intraocular Scattering Measurement System in Children.

Authors:  Mi Tian; Huamao Miao; Yang Shen; Jian Gao; Xiaofen Mo; Xingtao Zhou
Journal:  PLoS One       Date:  2015-11-04       Impact factor: 3.240

10.  Effect of pupil size on optical quality parameters in astigmatic eyes using a double-pass instrument.

Authors:  Hidenaga Kobashi; Kazutaka Kamiya; Kyohei Yanome; Akihito Igarashi; Kimiya Shimizu
Journal:  Biomed Res Int       Date:  2013-06-23       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.