Clinical evaluation of a new monofocal IOL with enhanced intermediate function in patients with cataract.

Gerd U Auffarth1, Matthias Gerl, Linda Tsai, D Priya Janakiraman, Beth Jackson, Aixa Alarcon, H Burkhard Dick.   

Abstract

PURPOSE: To evaluate the effectiveness and safety of 2 enhanced monofocal intraocular lenses (IOLs). The TECNIS Eyhance IOL (Model ICB00) was compared with a standard monofocal IOL (TECNIS Monofocal, Model ZCB00).
SETTING: European multicenter study.
DESIGN: Prospective, bilateral, randomized, comparative/evaluator-masked, controlled study.
METHODS: Adult subjects scheduled to undergo bilateral, primary phacoemulsification cataract extraction and posterior IOL implantation were randomized to receive the enhanced monofocal ICB00 IOL or the monofocal ZCB00 IOL in both eyes. Monocular endpoints at 6 months included distance-corrected intermediate visual acuity (DCIVA), photopic corrected distance visual acuity, and uncorrected intermediate visual acuity (UIVA). Binocular visual acuities, monocular corrected distance contrast sensitivity (first eyes), patient-reported outcomes, and safety were assessed at 6 months.
RESULTS: Overall, 139 patients were bilaterally implanted with the enhanced monofocal IOL (n = 67) or standard monofocal IOL (n = 72) and available for the 6-month visit. The enhanced monofocal IOL significantly improved mean monocular and binocular DCIVA and UIVA by at least 1-line logarithm of the minimum angle of resolution vs the standard monofocal IOL (all P ≤ .0001). Distance vision for the enhanced monofocal IOL was 20/20 or better and comparable with that of the standard monofocal lens at 6 months. Contrast sensitivity, photic phenomena outcomes, and rates of adverse events were similar between the 2 groups.
CONCLUSIONS: In patients undergoing cataract surgery, TECNIS Eyhance IOL Model ICB00 provided enhanced intermediate vision and similar distance performance and photic phenomena compared with a standard monofocal IOL, along with improved functional performance in daily life.
Copyright © 2021 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of ASCRS and ESCRS.

Entities:  

Year:  2021        PMID: 32932369     DOI: 10.1097/j.jcrs.0000000000000399

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

1.  Visual and optical quality of enhanced intermediate monofocal versus standard monofocal intraocular lens.

Authors:  Nuria Garzón; Francisco Poyales; César Albarrán-Diego; Laura Rico-Del-Viejo; Lidia Pérez-Sanz; María García-Montero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-05-27       Impact factor: 3.535

2.  Exploring Vision-Related Quality of Life: A Qualitative Study Comparing Patients' Experience of Cataract Surgery with a Standard Monofocal IOL and an Enhanced Monofocal IOL.

Authors:  Sergi Blancafort Alias; Zoraida Del Campo Carrasco; Ignacio Salvador-Miras; Sabina Luna Mariné; María José Gómez Prieto; Francesca Liñán Martín; Antoni Salvà Casanovas
Journal:  Clin Ophthalmol       Date:  2022-05-27

3.  Visual Performance and Optical Quality after Implantation of a New Generation Monofocal Intraocular Lens.

Authors:  Kyoung Hae Kang; Mi Yeon Song; Kook Young Kim; Kyu-Yeon Hwang; Young-A Kwon; Kyungmin Koh
Journal:  Korean J Ophthalmol       Date:  2021-04-05

4.  Monofocal intraocular lens with enhanced intermediate function as substitute for multifocal intraocular lens in positive dysphotopsia.

Authors:  Tadas Naujokaitis; Ling Zhao; Debora Scharf; Ramin Khoramnia; Gerd U Auffarth
Journal:  Am J Ophthalmol Case Rep       Date:  2022-04-10

5.  Analysis of a novel hydrophobic acrylic enhanced monofocal intraocular lens compared to its standard monofocal type on the optical bench.

Authors:  A F Borkenstein; E M Borkenstein; R Schmid
Journal:  BMC Ophthalmol       Date:  2022-09-03       Impact factor: 2.086

6.  Evaluation of macular visibility through a high-order aspheric intraocular lens using a simulated model eye.

Authors:  Yuji Yoshikawa; Junji Kanno; Kei Shinoda; Jun Makita
Journal:  Medicine (Baltimore)       Date:  2022-10-14       Impact factor: 1.817

7.  Optical Bench Analysis of 2 Depth of Focus Intraocular Lenses.

Authors:  Andreas F Borkenstein; Eva-Maria Borkenstein; Holger Luedtke; Ruediger Schmid
Journal:  Biomed Hub       Date:  2021-09-27

8.  Comparison Between an Intraocular Lens With Extended Depth of Focus (Tecnis Symfony ZXR00) and a New Monofocal Intraocular Lens With Enhanced Intermediate Vision (Tecnis Eyhance ICB00).

Authors:  Young Joon Jeon; Yisang Yoon; Tae-Im Kim; Kyungmin Koh
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2021-10-01
  8 in total