Jurgen Peerlings1,2, Inge Compter1, Fiere Janssen1, Christopher J Wiggins3, Alida A Postma2, Felix M Mottaghy2,4, Philippe Lambin1, Aswin L Hoffmann1,5,6,7.
Abstract
BACKGROUND ANDEntities:
Keywords: Anthropomorphic phantom; Diametric spherical volume; Geometrical distortion; Neuro-oncology; Radiation treatment planning; Ultra-high field MRI
Year: 2019 PMID: 33458423 PMCID: PMC7807620 DOI: 10.1016/j.phro.2018.12.001
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Overview of 3D data acquisition and analyses. XYZ-coordinates are determined on CT and MR images (1a.), and based on product characteristics of the CIRS’ phantom model 603A (1b.). Global MAD (2a.) is based on distances with 2 variable grid-intersection points (*example given for 1 intersection but applies for all points). Local MAD (2b.) is based on distances between magnetic field isocentre and 1 variable grid-intersection points. Displacements vectors (3a.) are determined between the measured and reference coordinates of each individual data-point and indicate the relative 3D geometrical distortion.
Global geometrical deformation as measured on 3T and 7T MRI (*** = P < 0.001).
| MADGlobal (in mm) | MADGlobal (in mm) | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| B0 | Sequence | Range | Mean ± sd | |CI95| | B0 | Sequence | Range | Mean ± sd | |CI95| | |
| 7T | MP2RAGE | 0.3–1.9 | 1.0 ± 0.3 | 1.7 | 3T | 3D TFE | 0.22–0.51 | 0.27 ± 0.23 | 0.7 | *** |
| SPACE | 0.3–2.0 | 1.0 ± 0.4 | 1.7 | T2-VISTA | 0.20–0.70 | 0.26 ± 0.20 | 0.7 | *** | ||
| SPACEFLAIR | 0.3–2.2 | 0.9 ± 0.3 | 1.6 | T2-VISTA FLAIR | 0.19–0.62 | 0.29 ± 0.23 | 0.7 | *** | ||
| GRE | 0.3–1.9 | 1.0 ± 0.4 | 1.8 | T1-FFE | 0.3–0.7 | 0.4 ± 0.3 | 1.0 | *** | ||
Fig. 2Absolute differences in Euclidian distance between the measured and reference dataset (grey dot) relative to the unique distances found in the reference dataset, observed within CT (a), MP2RAGE (b), T2-SPACE (c), T2-SPACE FLAIR (d), T1-GRE (e), 3D TFE (f), T2-VISTA (g), T2-VISTA FLAIR (h), T1-FFE (i). The overall geometric distortion was quantified by MADGlobal (±SD) (blue square). The 95% confidence interval (CI95) is shown as the dotted horizontal line. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Fig. 3MADLocal values (±SD) relative to the distance from the magnetic isocentre at 7T (a) and 3T (b) MRI, both relative to CT (black square). Presented 7T MR sequence include MP2RAGE (green circle), T2-SPACE (purple diamond), T2-SPACE FLAIR (orange downward triangle), and T1-GRE (blue upward triangle). The same colour- and shape-code was used for the equivalent 3T sequences, 3D TFE, T2-VISTA, T2-VISTA FLAIR, and T1-FFE, respectively. The dotted horizontal line represents the 1 mm-acceptability level required for spatially reliable RTP. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)
Fig. 4Distortion maps of MP2RAGE on 7T MRI (a) and 3D TFE on 3T MRI (b) measured in the axial plane nearest the magnetic isocentre (Y = 0), in the coronal plane (Z = 0), and in the sagittal plane (X = 0).
Local geometrical deformation as measured on 3T and 7T MRI (*** = P < 0.001, § = DSV at which statistical significance between sequences occurred).
| MADLocal (in mm) | MADLocal (in mm) | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B0 | Sequence | Range | Mean ± sd | |CI95| | <1 mm at DSV | B0 | Sequence | Range | Mean ± sd | |CI95| | <1 mm at DSV | Sign. | |
| 7T | MP2RAGE | 0.2–1.8 | 0.8 ± 0.3 | 1.4 | 69 | 3T | 3D TFE | 0.1–0.6 | 0.3 ± 0.2 | 0.6 | all | *** | 42 |
| SPACE | 0.2–1.18 | 0.7 ± 0.3 | 1.2 | 102 | T2-VISTA | 0.2–0.4 | 0.4 ± 0.3 | 0.9 | all | *** | 69 | ||
| SPACEFLAIR | 0.3–1.2 | 0.6 ± 0.3 | 1.2 | 93 | T2- VISTA FLAIR | 0.1–0.4 | 0.2 ± 0.2 | 0.5 | all | *** | 47 | ||
| GRE | 0.2–1.1 | 0.7 ± 0.3 | 1.3 | 87–89 | T1-FFE | 0.2–0.7 | 0.3 ± 0.3 | 0.9 | all | *** | 45–56 | ||