A complete distortion correction for MR images: I. Gradient warp correction.

Simon J Doran1, Liz Charles-Edwards, Stefan A Reinsberg, Martin O Leach.   

Abstract

MR images are known to be distorted because of both gradient nonlinearity and imperfections in the B0 field, the latter caused either by an imperfect shim or sample-induced distortions. This paper describes in detail a method for correcting the gradient warp distortion, based on a direct field mapping using a custom-built phantom with three orthogonal grids of fluid-filled rods. The key advance of the current work over previous contributions is the large volume of the mapping phantom and the large distortions (>25 mm) corrected, making the method suitable for use with large field of view, extra-cranial images. Experimental measurements on the Siemens AS25 gradient set, as installed on a Siemens Vision scanner, are compared with a theoretical description of the gradient set, based on the manufacturer's spherical harmonic coefficients. It was found that over a volume of 320x200x340 mm3 distortions can be successfully mapped to within the voxel resolution of the raw imaging data, whilst outside this volume, correction is still good but some systematic errors are present. The phenomenon of through-plane distortion (also known as 'slice warp') is examined in detail, and the perturbation it causes to the measurements is quantified and corrected. At the very edges of the region of support provided by the phantom, through-plane distortion is extreme and only partially corrected by the present method. Solutions to this problem are discussed. Both phantom and patient data demonstrate the efficacy of the gradient warp correction.

Entities:  

Mesh:

Year:  2005        PMID: 15798328     DOI: 10.1088/0031-9155/50/7/001

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  63 in total

1.  NonCartesian MR image reconstruction with integrated gradient nonlinearity correction.

Authors:  Shengzhen Tao; Joshua D Trzasko; Yunhong Shu; John Huston; Kevin M Johnson; Paul T Weavers; Erin M Gray; Matt A Bernstein
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

2.  Novel measurement technique of the tibial slope on conventional MRI.

Authors:  Robert Hudek; Silvia Schmutz; Felix Regenfelder; Bruno Fuchs; Peter P Koch
Journal:  Clin Orthop Relat Res       Date:  2009-02-04       Impact factor: 4.176

3.  Partial fourier and parallel MR image reconstruction with integrated gradient nonlinearity correction.

Authors:  Shengzhen Tao; Joshua D Trzasko; Yunhong Shu; Paul T Weavers; John Huston; Erin M Gray; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2015-07-17       Impact factor: 4.668

4.  MRI with intraoral orthodontic appliance-a comparative in vitro and in vivo study of image artefacts at 1.5 T.

Authors:  C Zachriat; P Asbach; K I Blankenstein; I Peroz; F H Blankenstein
Journal:  Dentomaxillofac Radiol       Date:  2015-03-03       Impact factor: 2.419

Review 5.  MRI in medical practice and its future use in radiation oncology. Resume of XXV GOCO Congress (Montpellier) 2017.

Authors:  Xavier Druet; Estrella Acosta Sanchez; Ken Soleakhena; Anne Laprie; Jordi Sáez; Stéphanie Nougaret; Olivier Riou; Elodie Rigal; Laura Kibranian; Miguel Palacios; Ismael Membrive
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-05

6.  Characterization of hardware-related spatial distortions for IR-PETRA pulse sequence using a brain specific phantom.

Authors:  Sima Ahmadian; Iraj Jabbari; Seyed Mehdi Bagherimofidi; Hamidreza Saligheh Rad
Journal:  MAGMA       Date:  2020-07-06       Impact factor: 2.310

Review 7.  Measuring and Establishing the Accuracy and Reproducibility of 3D Printed Medical Models.

Authors:  Elizabeth George; Peter Liacouras; Frank J Rybicki; Dimitrios Mitsouras
Journal:  Radiographics       Date:  2017-08-11       Impact factor: 5.333

8.  Large field of view distortion assessment in a low-field MR-linac.

Authors:  Siamak P Nejad-Davarani; Joshua P Kim; Dongsu Du; Carri Glide-Hurst
Journal:  Med Phys       Date:  2019-03-23       Impact factor: 4.071

Review 9.  MRI-only treatment planning: benefits and challenges.

Authors:  Amir M Owrangi; Peter B Greer; Carri K Glide-Hurst
Journal:  Phys Med Biol       Date:  2018-02-26       Impact factor: 3.609

10.  Distortion inherent to magnetic resonance imaging can lead to geometric miss in radiosurgery planning.

Authors:  Tyler M Seibert; Nathan S White; Gwe-Ya Kim; Vitali Moiseenko; Carrie R McDonald; Nikdokht Farid; Hauke Bartsch; Joshua Kuperman; Roshan Karunamuni; Deborah Marshall; Dominic Holland; Parag Sanghvi; Daniel R Simpson; Arno J Mundt; Anders M Dale; Jona A Hattangadi-Gluth
Journal:  Pract Radiat Oncol       Date:  2016-06-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.